Clinical predictors of prosthesis-patient mismatch after aortic valve replacement for aortic stenosis.

Luis M Astudillo1, Orlando Santana, Pablo A Urbandt, Alexandre M Benjo, Lior U Elkayam, Francisco O Nascimento, Gervasio A Lamas, Joseph Lamelas.   

Abstract

OBJECTIVE: We sought to ascertain predictors of Patient Prosthesis Mismatch, an independent predictor of mortality, in patients with aortic stenosis using bioprosthetic valves.
METHOD: We analyzed 2,107 sequential surgeries. Patient Prosthesis Mismatch was calculated using the effective orifice area of the prosthesis divided by the patient's body surface area. We defined nonsignificant, moderate, and severe Patient Prosthesis Mismatch as effective orifice area indexes of .0.85 cm(2)/m, 0.85-0.66 cm(2)/m(2), and <0.65 cm(2)/m(2), respectively.
RESULTS: A total of 311 bioprosthetic patients were identified. The incidence of nonsignificant, moderate, and severe Patient Prosthesis Mismatch was 41%, 42, and 16%, respectively. Severe Patient Prosthesis Mismatch was significantly more prevalent in females (82%). In severe Patient Prosthesis Mismatch, the perfusion and the crossclamp times were considerably lower when compared with nonsignificant Patient Prosthesis Mismatch and moderate Patient Prosthesis Mismatch. Patients with severe Patient Prosthesis Mismatch had a significantly higher likelihood of spending time in the intensive care unit and a significantly longer length of stay in the hospital. Body surface area was not different in severe Patient Prosthesis Mismatch when compared with nonsignificant Patient Prosthesis Mismatch. In-hospital mortality in patients with nonsignificant, moderate, and severe Patient Prosthesis Mismatch was 2.3%, 6.1%, and 8%, respectively. Minimally invasive surgery was significantly associated with moderate Patient Prosthesis Mismatch in 49% of the patients, but not with severe Patient Prosthesis Mismatch.
CONCLUSION: Severe Patient Prosthesis Mismatch is more common in females, but not in those with minimal available body surface area. Though operative times were shorter in these patients, intensive care unit and hospital lengths of stay were longer. Surgeons and cardiologists should be cognizant of these clinical predictors and complications prior to valve surgery.

Entities:  

Mesh:

Year:  2012        PMID: 22249481      PMCID: PMC3248602          DOI: 10.6061/clinics/2012(01)09

Source DB:  PubMed          Journal:  Clinics (Sao Paulo)        ISSN: 1807-5932            Impact factor:   2.365


INTRODUCTION

Aortic valve stenosis affects 2% to 4% of older adults and is the most common reason for valve surgery. An estimated 50,000 aortic valve procedures are performed each year in the United States (1). However, once the aortic valve is replaced, issues related to prosthesis-patient mismatch (PPM) may emerge. PPM was first described by Rahimtoola in 1978 (2). PPM is considered to be present when the prosthetic valve used in the patient is smaller than the normal native valve. As a consequence of the smaller prosthetic valve, the left ventricle has to produce higher pressures to overcome the mechanical resistance produced by the new device. This is transmitted in higher transvalvular pressure gradients as measured by Doppler (2). It has been demonstrated that the high-pressure gradients through the prosthesis are associated with higher morbidity and impact both short- and long-term mortality, when compared with prostheses of adequate size, based on the patient's body surface area (4-5). It has been shown that PPM is a common problem in patients undergoing AVR, occurring in up to 70% of aortic valve replacements (3). At the same time, this problem can be prevented through the use of a systematic approach prior to surgery (3). This may result in enhanced recovery with lower morbidity and mortality. The aim of this study was to determine the prevalence, clinical predictors, and short-term impact of severe PPM.

METHODS

Patient population: After obtaining approval from the independent review board, we retrospectively reviewed 2,107 consecutive heart surgeries performed at our institution between January 2005 and June 2009. Of these, 964 were coronary artery bypass surgeries, 683 involved single-valve surgeries, 411 were concomitant valve and bypass surgeries, and 144 were multivalve surgeries. From this group, patients with a diagnosis of severe aortic stenosis (AVA≤1 cm2) were included in the study, along with those in whom a bioprosthetic valve was used for the aortic valve replacement. Patients with concomitant valve or coronary artery bypass surgery were also included in the study, but patients with a diagnosis of aortic insufficiency as the reason for AVR were excluded. Baseline preoperative and operative variables were used based on the information provided by the Society of Thoracic Surgeons database in Tables 1 and 2.
Table 1

Preoperative Data.

PatientsNon PPMn = 129 (41%)Moderate PPMn = 132 (42%)SeverePPM n = 50(16%)p-value
Age (years)75±1173±977±70.6
Female45 (34.9%)42 (32.1%)41 (82%)<0.001
Body surface area (m2)1.80±0.21.97±0.21.75±0.250.155*)
Body mass index (Kg/m2)26.6±4.2729.4±5.427.8±6.20.120*)
NYHA class ≥ III49 (38%)49 (37%)28 (56%)0.05
Diabetes mellitus34 (26%)59 (44%)17 (34%)0.008
Renal failure3 (2.3%)4 (3%)6 (12%)0.01
Hypertension111 (86%)125 (95%)47 (94%)0.04
Cerebrovascular disease12 (9.3%)10 (7.6%)6 (12%)0.64
Smoker33 (25%)37 (28%)17 (34%)0.53
Chronic lung disease79 (61%)86 (65%)34 (68%)0.65
Peripheral vascular disease14 (11%)7 (5.3%)4 (8%)0.25
Prior coronary artery bypass grafting11 (8.5%)16 (12%)6 (12%)0.60
Myocardial infarction11 (8.5%)16 (12%)10 (20%)0.10
Left ventricular ejection fraction ≤55%45 (35%)47 (36%)14 (28%)0.60
Mean ejection fraction52±1252±1354±120.62
Preoperative aortic valve gradient (mmHg)53±1953±2251±180.38
Severe mitral insufficiency13 (10%)18 (14%)14 (28%)0.009
Preoperative Medical TherapyBeta-blockers74 (57%)85 (64%)30 (60%)0.50
Angiotensin converting enzyme inhibitors43 (33%)59 (45%)15 (30%)0.08
Aspirin46 (36%)55 (42%)19 (38%)0.60
Lipid lowering medications66 (51%)75 (57%)32 (64%)0.30

represents the p-value of severe PPM vs. no PPM.

NYHA: New York Heart Association heart failure class.

Table 2

Operative Data.

VariablesNon PPM n = 129 (41%)Moderate PPM n = 132 (42%)Severe PPM n = 50 (16%)p-value
Elective129 (100%)129 (98%)50 (100%)0.12
Coronary artery bypass grafting53 (41%)48 (36%)21 (42%)0.67
Left main disease4 (3.1%)8 (6.0%)4 (8%)0.34
Mitral valve surgery16 (12%)23 (17%)17 (34%)0.003
Tricuspid valve surgery01 (0.8%)4 (8%)0.01
Aortic annular enlargement2 (1.6%)1 (0.8%)3 (6%)0.06
Intra-aortic balloon pump use19 (15%)27 (20%)13 (26%)0.19
Intraoperative blood use68 (52.7%)78 (59.1%)28 (56%)0.58
Perfusion time (minutes)118±38114±40100±420.0091
Cross-clamp time (minutes)87.2±3283.3±3372.2±280.0052
Minimally invasive surgery38 (32%)37 (28%)0<0.001
Complications
Total intensive care unit length of stay (hours)146±339130±137203±2500.0123
Prolonged ventilation (hours)73±26848±100112±3040.08
Permanent neurologic complications5 (3.9%)3 (2.3%)2 (4%)0.72
Transient neurologic complications01 (0.8%)1 (2%)0.31
Pneumonia10 (7.8%)6 (4.5%)3 (6%)0.13
Renal failure10 (7.8%)10 (7.6%)5 (10%)0.9
Short-term mortality3 (2.3%)8 (6.1%)4 (8%)0.19
Length of hospital stay (days)15±1614±1020±130.04

1p-value between severe PPM versus no PPM. p = 0.05 between severe PPM versus moderate PPM.

2p-value between severe PPM versus no PPM. p = 0.04 between severe PPM vs. moderate PPM.

3p-value between severe PPM versus moderate PPM. p = 0.27 between severe PPM and no PPM.

Measurements: The effective orifice area (EOA) of each valve was obtained from a list that the American Society of Echocardiography published in their 2009 Recommendations for the Evaluation of Prosthetic Valves with Echocardiography and Doppler Ultrasound (6), which was based on a published report of normally functioning aortic valve prostheses (7). The effective orifice area values obtained from in vivo measurements for the different bioprosthetic aortic valves used during this analysis are given in Table 3. PPM was calculated using the EOA of the aortic valve prosthesis, which was divided by the patient's body surface area to obtain the indexed EOA. We defined nonsignificant, moderate, and severe PPM as an indexed EOA of >0.85 cm2/m2, 0.66 cm2/m2 to ≤0.85 cm2/m2, and ≤0.65 cm2/m2, respectively (12). We defined the nonsignificant PPM group as our control because they received the optimal therapy. Mortality was analyzed at the time of discharge from the hospital.
Table 3

Reference values of effective orifice area for bioprosthetic aortic valves.

Size, mmProstheticValve
192123252729
No. of Patients54989051171
Mosaic Porcine1211.41.51.81.92.1
Freestyle41.41.72.12.5
Mitroflow11.11.31.51.8
Carpentier Edwards Standard1660.91.51.71.92.32.8
stented porcine
Carpentier Edwards Pericardial191.21.51.8
stented bovine pericardial
Statistical Analysis: Continuous variables are expressed as the mean ±1 SD; comparisons were conducted by t-tests/ANOVAs or nonparametric Mann-Whitney/Kruskal-Wallis tests if the normality assumption was violated when comparing severe PPM with no PPM or severe PPM with moderate PPM. Discrete variables are presented as percentages and relatives frequencies; comparisons were conducted by chi-squared statistics or Fisher's exact test, as appropriate. A p-value <0.05 was considered statistically significant. A stepwise logistic regression analysis (backward Wald) was used for the variables with a p-value <0.05 to assess the independent predictors of PPM and length of stay in the intensive care unit. We used a software system (Statistical Package for the Social Sciences, SPSS version 17 Chicago, Illinois) for all analyses.

RESULTS

A total of 311 patients (128 females and 183 males) with a diagnosis of severe aortic stenosis who underwent AVR with a bioprosthetic valve were identified. Based on the severity of the PPM, the cohort was divided in three groups: non-significant PPM (n = 129), 41%; moderate PPM (n = 132), 42%; and severe PPM, (n = 50) 16%. PPM was present in 58% of the total population, in 42% as moderate PPM and in 16% as severe PPM. Patient characteristics: The baseline characteristics are presented in Table 1. The mean age was 75±11, 73±9, and 77±7 years for nonsignificant, moderate, and severe PPM, respectively (p = 0.6). Females represented 82% of the severe PPM group compared with 34% and 32% of the nonsignificant and moderate PPM groups, respectively (p<0.001) (Figure 1). There was no discrepancy in age between the different groups of PPM, (p = 0.6). The body surface area was lowest in the severe PPM group (1.75±0.5 m2) compared with 1.97±0.2 m2 of the moderate PPM (p<0.001), but this was not significantly different when compared with the nonsignificant PPM group value of 1.80±0.2 m2 (p = 0.155).
Figure 1

Percentage of prosthesis-patient mismatch by gender.

Body mass index did not differ between the severe (27.8±6.2 Kg/m2) and moderate groups (29.4±5.4 Kg/m2), (p = 0.10), or when comparing nonsignificant (26.6±4.27 Kg/m2) and severe mismatch groups, (p = 0.12). Advanced heart failure (based on the New York Heart Association criteria) was found in 56% of patients in the severe group (p = 0.05). Diabetes mellitus was more prevalent in patients with moderate PPM (p = 0.008). Hypertension was prevalent in all of the groups, in particular, in those with moderate and severe mismatch (p = 0.04). Renal failure (serum creatinine >2 mg/dl) was higher in the severe PPM group, (p = 0.01). Coronary artery disease, with previous coronary artery bypass grafting (CABG) was similar in all groups as well, (p = 0.6). The echocardiographic basal characteristics of the study population presented no differences in the mean ejection fraction: no PPM, 52±12%; moderate PPM, 52±13%; and severe PPM, 54±12%, (p = 0.62). Likewise, the aortic valve gradient was not significantly different: 53±19 mmHg, 53±22 mmHg, and 51±18 mmHg, in that order, (p = 0.38). Surgical data: Most of the procedures were elective. Coronary artery bypass graft surgery was performed in 41% (nonsignificant PPM), 36% (moderate PPM), and 42% (severe PPM), respectively, (p = 0.67). Concomitant mitral valve surgery was performed in 34% of the severe PPM group, 17% of the moderate PPM group, and 12% in the no-PPM group (p = 0.003). Tricuspid valve surgery was completed in 8% of the severe PPM cases. A total of six (6) aortic annular enlargement procedures was performed; two in the no PPM, one in the moderate PPM, and three in the severe PPM groups. Interestingly, the operative data in Table 2 show that patients in the severe PPM group had the shortest perfusion (p = 0.009) and cross-clamp times (p = 0.005) (Figure 2). However, the same group had the highest total intensive care unit length of stay with a mean value of 203 hours compared with 146 and 130 hours in the nonsignificant and moderate PPM groups, respectively (Figure 3). This was statistically significant (p = 0.012) when comparing severe and moderate PPM. No other factors were associated with the amount of intensive care unit hours except for the PPM difference. However, there was a statistical trend (p = 0.08) for prolonged mechanical ventilation in hours in the severe mismatch group (Figure 4). Along with these findings, patients with severe mismatch had the longest hospital stay with a mean value of 20 days compared with 15 and 14 days in the nonsignificant and moderate PPM groups, respectively (p = 0.04) (Figure 5). Fifteen cases of in-hospital mortality were identified, three in nonsignificant PPM, eight in moderate PPM, and four in the severe PPM groups (p = 0.19). Seventy-five minimally invasive surgeries were performed, 38 (50%) had nonsignificant PPM and 37 (50%) had moderate PPM. No case of severe PPM was found to be associated with minimally invasive surgery.
Figure 2

Total amount of intensive care unit hours distributed according to patient-prosthesis mismatch.

Figure 3

Hours of mechanical ventilation according to the severity of patient-prosthesis mismatch.

Figure 4

Total amount of mechanical ventilation in hours distributed according to patient-prosthesis mismatch.

Figure 5

Total length of in-hospital stay in days distributed according to patient-prosthesis mismatch.

Table 3 shows the different types of bioprosthetic valves used. The Carpentier Edwards Standard (Edwards Lifesciences, Irvine, CA), and the Mosaic Porcine (Medtronic, Minneapolis, MN) were the most commonly used, 53% and 39% of the time, respectively. A valve size of 21 mm was implanted in 31% of the surgeries, making it the most frequently used, while a valve size of 23 mm was the second most common, being used 29% of the time. Independent Risk Factors: Univariate analysis from preoperative data showed that female gender and renal failure were statistically associated with severe PPM. Diabetes mellitus and ACE inhibitors use were associated with the moderate PPM group but not with the severe PPM group. Further multivariable analysis demonstrated that female gender was associated with severe mismatch (p<0.001), even when compared with the nonsignificant as well as the moderate PPM groups.

DISCUSSION

In patients undergoing AVR, PPM is a commonly encountered problem that leads to worsened hemodynamic function, less regression of left ventricular hypertrophy, and more cardiac events with lower survival rates (4,11,12). Other studies have demonstrated an association between female gender and higher operative mortality after valvular heart surgery (15). Our results indicate that female gender is an independent risk factor for severe prosthesis-patient mismatch. This association was observed after the body surface area of the severe mismatch group proved not to be significantly different when compared with the group without any mismatch. When compared with men, women in general have a smaller body surface area and a smaller aortic annulus. Accordingly, PPM has proven to be more common in older females with concomitant coronary artery disease (16-18). Our study demonstrated that 82 percent of severe mismatches corresponded to females without differences in age or any other baseline characteristics. Interestingly, this particular group had the shortest operative times. It is our impression that the use of smaller prosthetic valves for small and severely calcified annuli may have shortened the surgical times. It is possible that the reason is associated with the fact that surgeons did not have to enlarge the aortic root. PPM is therefore more frequently encountered in females, and the discrepancy between the inadequate size of both the prosthesis and the patient have proven to be risk factors for mortality (4,12,13). At the same time, the female heart responds differently than the male heart does for the same degree of left ventricular outflow obstruction, resulting in smaller end-systolic chamber size and higher pressures in the female heart response (18). This is particularly common with elderly patients with an excessive or inappropriate degree of hypertrophy; wall thickness is greater than necessary to counterbalance the high intracavitary pressures. As a result, the systolic wall stress is lower, and the ejection fraction is higher; such inappropriate LV hypertrophy has been associated with higher perioperative morbidity and mortality (17-19). We found a significant impact of severe PPM on the intensive care unit length of stay and on the total hospital length of stay. Because of the nature of the study, we cannot establish a causal relationship between the degree of PPM and the findings. However, it has been demonstrated that abnormal intracavitary flow accelerations (defined as the presence of a dagger-shaped intracavitary flow signal on continuous-wave Doppler) after aortic valve replacement for severe aortic stenosis are associated with concentric left ventricular hypertrophy and supernormal systolic function. The abnormal intracavitary flow acceleration and concentric left ventricular hypertrophy were associated with higher in-hospital mortality, morbidity and prolonged intubation (17-20). Although not directly investigated in our study, this may have played a role in our patients. We demonstrated that minimally invasive surgery may be related to a higher frequency of moderate PPM, but also to a lower rate of severe PPM; we believe that if this is confirmed, it is likely to have clinical implications with respect to the selection of a particular surgical technique. For this reason, we are currently prospectively exploring this possibility in our institution. This study highlights the importance of avoiding PPM, particularly in females who react differently from a pathophysiologic standpoint. Aortic PPM can be avoided through optimal prosthesis selection in the individual patient by calculating the necessary indexed EOA prior to surgery (3). This is particularly important in patients with left ventricular dysfunction and/or severe left ventricular hypertrophy (3). Study Limitations: The current study was based on data obtained from a patient registry with retrospective analysis and was not able to establish a causal relationship. The postoperative intracavitary flow accelerations were not available but may have helped to confirm our hypothesis. The assessment of these parameters along with the patient's outcome may have an impact on selecting the appropriate therapy.
  19 in total

Review 1.  Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention.

Authors:  P Pibarot; J G Dumesnil
Journal:  J Am Coll Cardiol       Date:  2000-10       Impact factor: 24.094

Review 2.  Prosthesis-patient mismatch: definition, clinical impact, and prevention.

Authors:  P Pibarot; J G Dumesnil
Journal:  Heart       Date:  2005-10-26       Impact factor: 5.994

Review 3.  Doppler echocardiography in normally functioning replacement aortic valves: a review of 129 studies.

Authors:  Ronak Rajani; Dayal Mukherjee; John B Chambers
Journal:  J Heart Valve Dis       Date:  2007-09

4.  Prediction of operative mortality after valve replacement surgery.

Authors:  F H Edwards; E D Peterson; L P Coombs; E R DeLong; W R Jamieson; F L Grover
Journal:  J Am Coll Cardiol       Date:  2001-03-01       Impact factor: 24.094

5.  Malignant natural history of asymptomatic severe aortic stenosis: benefit of aortic valve replacement.

Authors:  Ramdas G Pai; Nikhil Kapoor; Ramesh C Bansal; Padmini Varadarajan
Journal:  Ann Thorac Surg       Date:  2006-12       Impact factor: 4.330

6.  Impact of valve prosthesis-patient mismatch on short-term mortality after aortic valve replacement.

Authors:  Claudia Blais; Jean G Dumesnil; Richard Baillot; Serge Simard; Daniel Doyle; Philippe Pibarot
Journal:  Circulation       Date:  2003-08-11       Impact factor: 29.690

Review 7.  The problem of valve prosthesis-patient mismatch.

Authors:  S H Rahimtoola
Journal:  Circulation       Date:  1978-07       Impact factor: 29.690

8.  Sex-associated differences in left ventricular function in aortic stenosis of the elderly.

Authors:  J D Carroll; E P Carroll; T Feldman; D M Ward; R M Lang; D McGaughey; R B Karp
Journal:  Circulation       Date:  1992-10       Impact factor: 29.690

9.  Influence of concentric left ventricular remodeling on early mortality after aortic valve replacement.

Authors:  Andra Ibrahim Duncan; Boris S Lowe; Mario J Garcia; Meng Xu; A Marc Gillinov; Tomislav Mihaljevic; Colleen G Koch
Journal:  Ann Thorac Surg       Date:  2008-06       Impact factor: 4.330

10.  Abnormal left ventricular intracavitary flow acceleration in patients undergoing aortic valve replacement for aortic stenosis. A marker for high postoperative morbidity and mortality.

Authors:  G Aurigemma; S Battista; D Orsinelli; A Sweeney; L Pape; H Cuénoud
Journal:  Circulation       Date:  1992-09       Impact factor: 29.690

View more
  4 in total

Review 1.  On the Mechanics of Transcatheter Aortic Valve Replacement.

Authors:  Lakshmi P Dasi; Hoda Hatoum; Arash Kheradvar; Ramin Zareian; S Hamed Alavi; Wei Sun; Caitlin Martin; Thuy Pham; Qian Wang; Prem A Midha; Vrishank Raghav; Ajit P Yoganathan
Journal:  Ann Biomed Eng       Date:  2016-11-21       Impact factor: 3.934

2.  Early and mid-term haemodynamic performance and clinical outcomes of St. Jude Medical Trifecta™ valve.

Authors:  Renata Raimundo; Soraia Moreira; Francisca Saraiva; Rui J Cerqueira; Pedro Teixeira; Elson Salgueiro; André Lourenço; Mário J Amorim; Jorge Almeida; Paulo Pinho; Adelino F Leite-Moreira
Journal:  J Thorac Dis       Date:  2018-02       Impact factor: 2.895

3.  Cardiovascular research in CLINICS.

Authors:  Mauricio Rocha-e-Silva
Journal:  Clinics (Sao Paulo)       Date:  2013-06       Impact factor: 2.365

4.  Early outcomes of patient-prosthesis mismatch following aortic valve replacement.

Authors:  Serik Aitaliyev; Egle Rumbinaitė; Karolina Mėlinytė-Ankudavičė; Rokas Nekrošius; Vytenis Keturakis; Rimantas Benetis
Journal:  Perfusion       Date:  2021-06-03       Impact factor: 1.581

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.