In-line filter included into the syringe infusion pump assembly reduces flow irregularities.

B Brotschi1, B Grass, M Weiss, C Doell, V Bernet.   

Abstract

PURPOSE: To evaluate whether an in-line filter inserted in the syringe pump infusion line assembly influences start-up times and flow irregularities during vertical pump displacement at low infusion rates.
METHODS: Fluid delivery after syringe pump start-up and after vertical displacement of the syringe pump by -50 cm was determined gravimetrically at flow rates of 0.5, 1.0 and 2.0 ml h(-1). Measurements were repeated for each flow rate four times with two different syringe pumps with and without an in-line filter incorporated. Data are shown as median and range.
RESULTS: Start-up times were reduced by an in-line filter at 0.5 ml h(-1) flow rate from 355.5 s (0-660) to 115 s (0-320), whereas the effect was attenuated at higher flow rates. Pooling of fluid into the infusion system after lowering the infusion syringe pump was halved in all flow rates tested. Amount of infusion bolus after elevating the syringe pump by 50 cm was not affected by an in-line filter.
CONCLUSION: In the evaluated model in-line filters help to reduce flow irregularities and delay in drug delivery of syringe pumps at low flow rates and represent an option to optimize continuous administration of highly concentrated short-acting drugs at very small infusion rates.

Mesh:

Year:  2012        PMID: 22237747     DOI: 10.1007/s00134-011-2452-5

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  14 in total

1.  Infusion pump performance with vertical displacement: effect of syringe pump and assembly type.

Authors:  T A Neff; J E Fischer; G Schulz; O Baenziger; M Weiss
Journal:  Intensive Care Med       Date:  2001-01       Impact factor: 17.440

2.  Syringe size and flow rate affect drug delivery from syringe pumps.

Authors:  M Weiss; M I Hug; T Neff; J Fischer
Journal:  Can J Anaesth       Date:  2000-10       Impact factor: 5.063

3.  The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps.

Authors:  M Weiss; J Fischer; T Neff; O Baenziger
Journal:  Anaesthesia       Date:  2000-11       Impact factor: 6.955

4.  Downward movement of syringe pumps reduces syringe output.

Authors:  H Kern; A Kuring; U Redlich; U R Döpfmer; N M Sims; C D Spies; W J Kox
Journal:  Br J Anaesth       Date:  2001-06       Impact factor: 9.166

5.  Evaluation of the FASTSTART mode for reducing start-up delay in syringe pump infusion systems.

Authors:  T Neff; J Fischer; S Fehr; O Baenziger; M Weiss
Journal:  Swiss Med Wkly       Date:  2001-04-21       Impact factor: 2.193

Review 6.  Intravenous in-line filters: filtering the evidence.

Authors:  Patrick A Ball
Journal:  Curr Opin Clin Nutr Metab Care       Date:  2003-05       Impact factor: 4.294

7.  Bacterial colonization and endotoxin contamination of intravenous infusion fluids.

Authors:  M Trautmann; B Zauser; H Wiedeck; K Buttenschön; R Marre
Journal:  J Hosp Infect       Date:  1997-11       Impact factor: 3.926

8.  A gravimetric technique for evaluating flow continuity from two infusion devices.

Authors:  R D Leff; W R True; R J Roberts
Journal:  Am J Hosp Pharm       Date:  1987-06

9.  Syringe pumps for infusion of vasoactive drugs: mechanical idiosyncrasies and recommended operating procedures.

Authors:  G A Rooke; T A Bowdle
Journal:  Anesth Analg       Date:  1994-01       Impact factor: 5.108

10.  Analysis of particulate contaminations of infusion solutions in a pediatric intensive care unit.

Authors:  Thomas Jack; Bernadette E Brent; Martin Boehne; Meike Müller; Katherina Sewald; Armin Braun; Armin Wessel; Michael Sasse
Journal:  Intensive Care Med       Date:  2010-02-18       Impact factor: 17.440

View more
  4 in total

Review 1.  Strategies to prevent drug incompatibility during simultaneous multi-drug infusion in intensive care units: a literature review.

Authors:  Laura Négrier; Anthony Martin Mena; Gilles Lebuffe; Pascal Odou; Stéphanie Genay; Bertrand Décaudin
Journal:  Eur J Clin Pharmacol       Date:  2021-03-25       Impact factor: 2.953

2.  Analytical method for calculation of deviations from intended dosages during multi-infusion.

Authors:  Maurits K Konings; Roland A Snijder; Joris H Radermacher; Annemoon M Timmerman
Journal:  Biomed Eng Online       Date:  2017-01-17       Impact factor: 2.819

Review 3.  Year in review in Intensive Care Medicine 2012: III. Noninvasive ventilation, monitoring and patient-ventilator interactions, acute respiratory distress syndrome, sedation, paediatrics and miscellanea.

Authors:  Massimo Antonelli; Marc Bonten; Maurizio Cecconi; Jean Chastre; Giuseppe Citerio; Giorgio Conti; J R Curtis; Goran Hedenstierna; Michael Joannidis; Duncan Macrae; Salvatore M Maggiore; Jordi Mancebo; Alexandre Mebazaa; Jean-Charles Preiser; Patricia Rocco; Jean-François Timsit; Jan Wernerman; Haibo Zhang
Journal:  Intensive Care Med       Date:  2013-01-22       Impact factor: 17.440

4.  Effectiveness of in-Line Filters to Completely Remove Particulate Contamination During a Pediatric Multidrug Infusion Protocol.

Authors:  Maxime Perez; Bertrand Décaudin; Wadih Abou Chahla; Brigitte Nelken; Laurent Storme; Morgane Masse; Christine Barthélémy; Gilles Lebuffe; Pascal Odou
Journal:  Sci Rep       Date:  2018-05-16       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.