Syringe size and flow rate affect drug delivery from syringe pumps.

M Weiss1, M I Hug, T Neff, J Fischer.   

Abstract

PURPOSE: To evaluate the effect of syringe size and infusion rate on drug delivery after vertical displacement of syringe pumps.
METHODS: Four syringes (10 ml, 20 ml, 30 ml, 50 ml) were studied at three infusion speeds (2 ml x hr(-1), 1 ml x hr(-1) and 0.5 ml x hr(-1)). Fluid delivery was measured gravimetrically using an electronic balance. Aspiration volume (amount of fluid retracted into the syringe-infusion line assembly) and zero-drug delivery time (time between the lowering of the syringe pump and reattainment of the initial weight) were determined after lowering the syringe pump 130 cm or 50 cm. Then, infusion bolus was measured after elevating the pump to its initial position. Syringe compliance was calculated from the occlusion release bolus at an occlusion pressure of 300 mmHg.
RESULTS: Zero-drug delivery time, aspiration volume and infusion bolus increased with syringe size (P < 0.0001). At 0.5 ml x hr(-1), a zero-drug delivery time of 29.7 +/- 0.6 min was recorded for the 50 ml syringe and 5.5 +/- 0.15 min for the 10 ml syringe. As infusion rates increased, zero-drug delivery times decreased following a 1/x ratio. Zero-drug delivery time highly correlated with calculated syringe compliance (Pearson's r = 0.97, P < 0.001), which was closely related to syringe plunger area (R2 = 0.93, P < 0.001).
CONCLUSIONS: Vertical displacement of syringe pumps causes flow irregularities which correlate directly with syringe size and indirectly with infusion rate. Extreme vertical maneuvering of syringe pumps should be avoided. Infusion rates > or = 1 ml x hr(-1) and low-compliant, small syringes are recommended when highly concentrated drugs are administered.

Mesh:

Year:  2000        PMID: 11032282     DOI: 10.1007/BF03024878

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  2 in total

1.  In-line filter included into the syringe infusion pump assembly reduces flow irregularities.

Authors:  B Brotschi; B Grass; M Weiss; C Doell; V Bernet
Journal:  Intensive Care Med       Date:  2012-01-12       Impact factor: 17.440

2.  Analytical method for calculation of deviations from intended dosages during multi-infusion.

Authors:  Maurits K Konings; Roland A Snijder; Joris H Radermacher; Annemoon M Timmerman
Journal:  Biomed Eng Online       Date:  2017-01-17       Impact factor: 2.819