Downward movement of syringe pumps reduces syringe output.

H Kern1, A Kuring, U Redlich, U R Döpfmer, N M Sims, C D Spies, W J Kox.   

Abstract

We studied how lowering a syringe pump and changing the outflow pressure could affect syringe pump output. We experimentally reduced the height of three different syringe pump systems by 80 cm (adult setting) or 130 cm (neonatal setting), as can happen clinically, using five flow rates. We measured the time of backward flow, no flow and the total time without flow. An exponential negative correlation was present between infusion rate and time without flow (r2=0.809 to 0.972, P<0.01). Minimum flow rates of 4.4 and 2.6 ml h(-1) respectively were calculated to give 60 and 120 s without infusion. The compliance of the different syringe pumps and their infusion systems was linearly correlated with the effective time without infusion (r2=0.863, P<0.05). We conclude that the height of the syringe pumps should not be changed during transportation. If vertical movement of the syringe pump is necessary, the drugs should be diluted so that the flow rate is at least 5 ml h(-1).

Mesh:

Substances:

Year:  2001        PMID: 11573591     DOI: 10.1093/bja/86.6.828

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  5 in total

1.  In-line filter included into the syringe infusion pump assembly reduces flow irregularities.

Authors:  B Brotschi; B Grass; M Weiss; C Doell; V Bernet
Journal:  Intensive Care Med       Date:  2012-01-12       Impact factor: 17.440

2.  Siphon effects on continuous subcutaneous insulin infusion pump delivery performance.

Authors:  Howard C Zisser; Wendy Bevier; Eyal Dassau; Lois Jovanovic
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

3.  Predictors of haemodynamic instability during the changeover of norepinephrine infusion pumps.

Authors:  Martin Cour; Thomas Bénet; Romain Hernu; Marie Simon; Thomas Baudry; Philippe Vanhems; Laurent Argaud
Journal:  Ann Intensive Care       Date:  2016-04-22       Impact factor: 6.925

4.  Analytical method for calculation of deviations from intended dosages during multi-infusion.

Authors:  Maurits K Konings; Roland A Snijder; Joris H Radermacher; Annemoon M Timmerman
Journal:  Biomed Eng Online       Date:  2017-01-17       Impact factor: 2.819

5.  Changeovers of vasoactive drug infusion pumps: impact of a quality improvement program.

Authors:  Laurent Argaud; Martin Cour; Olivier Martin; Marc Saint-Denis; Tristan Ferry; Agnes Goyatton; Dominique Robert
Journal:  Crit Care       Date:  2007       Impact factor: 9.097

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.