The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps.

M Weiss1, J Fischer, T Neff, O Baenziger.   

Abstract

Fluid delivery from four types of commercially available 50-ml syringes was measured using an electronic balance at an infusion rate of 1 ml.h(-1). Retrograde aspiration volume and zero-drug delivery time were recorded after lowering the syringe pump by 50 cm. Syringe compliance was calculated from the volume of bolus released after occlusion at 100 mmHg. Zero-drug delivery times differed significantly between syringes, ranging from [mean (SD)] 3.26 (0.40) min to 6.38 (0.56) min (F = 55.5, d.f. = 3/20, p < 0.0001). Syringe compliance correlated well with aspiration volume (Pearson r(2) = 0.92, p < 0.001) and zero-drug delivery time (r(2) = 0.90, p < 0.001). Syringe design affected the internal syringe compliance. All syringes were associated with potentially relevant zero-drug delivery times after moderate vertical displacement. To minimise this risk, vertical displacement of syringe pumps delivering highly vasoactive drugs should be avoided.

Mesh:

Year:  2000        PMID: 11069336     DOI: 10.1046/j.1365-2044.2000.01550.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  5 in total

1.  In-line filter included into the syringe infusion pump assembly reduces flow irregularities.

Authors:  B Brotschi; B Grass; M Weiss; C Doell; V Bernet
Journal:  Intensive Care Med       Date:  2012-01-12       Impact factor: 17.440

2.  Siphon effects on continuous subcutaneous insulin infusion pump delivery performance.

Authors:  Howard C Zisser; Wendy Bevier; Eyal Dassau; Lois Jovanovic
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

3.  Impact of priming the infusion system on the performance of target-controlled infusion of remifentanil.

Authors:  Jong-Yeop Kim; Bong-Ki Moon; Jong Hyuk Lee; Youn Yi Jo; Sang-Kee Min
Journal:  Korean J Anesthesiol       Date:  2013-05-24

4.  Arrangements of the intravenous parallel infusions with anti-reflux valves decreasing occlusion alarm delay.

Authors:  Han Bum Joe; Bong-Ki Moon; Yeon-Ju Lee; Sang-Kee Min
Journal:  Korean J Anesthesiol       Date:  2014-04-28

5.  Analytical method for calculation of deviations from intended dosages during multi-infusion.

Authors:  Maurits K Konings; Roland A Snijder; Joris H Radermacher; Annemoon M Timmerman
Journal:  Biomed Eng Online       Date:  2017-01-17       Impact factor: 2.819

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.