| Literature DB >> 22231220 |
A Kyle Jones1, Alexander S Pasciak.
Abstract
Part II of this review article presents four case studies for which peak skin doses are calculated following the methods outlined in Part I of this review. The data available in the cases ranges from proprietary dose reports to fluoroscopy time and number of digital acquisition frames only. Flowcharts are provided for each case. These flowcharts outline the calculation steps and data sources used to estimate the peak skin dose. The accuracy that can be achieved using these methods depends on several factors, including the calibration accuracy of dosimetric equipment, accuracy of information reported in the DICOM header and proprietary dose reports, accuracy of quantities measured by the medical physicist, and procedural factors such as rotation of the C-arm during a fluoroscopically-guided procedure.Entities:
Mesh:
Year: 2012 PMID: 22231220 PMCID: PMC5716137 DOI: 10.1120/jacmp.v13i1.3693
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Abbreviations used in this manuscript.
|
|
|
|
|---|---|---|
| Peak skin dose | PSD | The highest dose to a single area of the skin, units of Gy |
| Fluoroscopically‐guided intervention | FGI | Any medical intervention using fluoroscopy for guidance |
| Reference point air kerma |
| Cumulative air kerma at the interventional reference point, units of Gy |
| Kerma area product | KAP | Product of air kerma and x‐ray field size, units of Gy‐cm2 |
| Digital Imaging and Communications in Medicine | DICOM | Standard for communication used by medical imaging equipment |
| Interventional reference point | IRP | Point located 15 cm back towards the focal spot from the isocenter of a C‐arm fluoroscope(
|
| Source‐to‐image distance | SID | Distance from the focal spot to the center of the image receptor, units of mm |
| Field of view | FOV | Size of the x‐ray field at the image receptor, units of cm |
| Entrance skin air kerma | ESAK | Air kerma measured at the entrance surface of the patient, units of Gy |
| Source‐to‐patient distance | SPD | Distance from the focal spot to the entrance surface of the patient, units of mm |
| Backscatter factor | BSF | Factor that is applied to calculate entrance surface dose from the ESAK, accounts for the fact that many X Rays of diagnostic energy are backscattered in tissue, unitless |
| f‐factor | f | Factor used to convert exposure to dose in tissue, units of Gy/R. A similar, unitless quantity can be used to convert from air kerma to dose in tissue. |
| Digital acquisition series | DAS | A series of images generated using digital acquisition imaging |
| Fluoroscopic reference point air kerma |
| Total reference point air kerma resulting from fluoroscopic imaging, units of Gy |
| Digital acquisition reference point air kerma |
| Total reference point air kerma resulting from digital acquisition imaging, units of Gy |
| Total reference point air kerma |
| Total reference point air kerma resulting from the sum of fluoroscopic and digital acquisition imaging, units of Gy |
| Projected x‐ray field size on the skin |
| The area of the projected x‐ray field at the entrance surface of the patient, units of cm2 |
Figure 1Sources of information used to calculate PSD in Case 1: (a) for each DAS was listed in a proprietary dose report; (b) distances were listed in the DICOM header; (c) the location of the IRP was listed in the equipment operator's manual.
Figure 2Data used to estimate the BSF and f‐factor. kV and added filtration* were listed in both the (a) proprietary dose report and (b) the DICOM header. Data used to calculate the x‐ray field size (c) were available in the DICOM header.
*Cu filter thickness is in tag Private_0021_100a.
Figure 3Flowchart outlining the methods and sources of data used to estimate the PSD for a single digital acquisition series. The same method was used to estimate the PSD for other digital acquisition series and fluoroscopy.
Figure 4Flowchart outlining the methods and sources of data used to estimate the PSD for Case 2. The concept of the biologically effective dose (BED) is presented in Balter et al.( )
Figure 5Image from postprocedural CT in Case 3 that was used to measure patient thickness in the anteroposterior direction. Patient thickness was measured at the location of the newly created shunt, which is visible in the image.
Figure 6Simulation setup used to measure data for estimation of PSD when no dose metrics were available. The setup was based on recorded procedural information and a review of archived images: (a) source‐to‐chamber distance, which was equal to the SPD used during the actual procedure; (b) patient‐equivalent thickness of polymethylmethacrylate (PMMA); (c) SID as used during the actual procedure; (d) ionization chamber or solid‐state dosimeter; (e) indicated focal spot on X‐ray tube housing.
Figure 7Flowchart outlining the methods and sources of data used to estimate the PSD for Case 3.
Figure 8Flowchart outlining the methods and sources of data used to estimate the PSD for Case 4.