Literature DB >> 22159012

Brillouin optical microscopy for corneal biomechanics.

Giuliano Scarcelli1, Roberto Pineda, Seok Hyun Yun.   

Abstract

PURPOSE: The mechanical properties of corneal tissue are linked to prevalent ocular diseases and therapeutic procedures. Brillouin microscopy is a novel optical technology that enables three-dimensional mechanical imaging. In this study, the feasibility of this noncontact technique was tested for in situ quantitative assessment of the biomechanical properties of the cornea.
METHODS: Brillouin light-scattering involves a spectral shift proportional to the longitudinal modulus of elasticity of the tissue. A 532-nm single-frequency laser and a custom-developed ultrahigh-resolution spectrometer were used to measure the Brillouin frequency. Confocal scanning was used to perform Brillouin elasticity imaging of the corneas of whole bovine eyes. The longitudinal modulus of the bovine corneas was compared before and after riboflavin corneal collagen photo-cross-linking. The Brillouin measurements were then compared with conventional stress-strain mechanical test results.
RESULTS: High-resolution Brillouin images of the cornea were obtained, revealing a striking depth-dependent variation of the elastic modulus across the cornea. Along the central axis, the Brillouin frequency shift varied gradually from 8.2 GHz in the epithelium to 7.5 GHz near the endothelium. The coefficients of the down slope were measured to be approximately 1.09, 0.32, and 2.94 GHz/mm in the anterior, posterior, and innermost stroma, respectively. On riboflavin collagen cross-linking, marked changes in the axial Brillouin profiles (P < 0.001) were noted before and after cross-linking.
CONCLUSIONS: Brillouin imaging can assess the biomechanical properties of cornea in situ with high spatial resolution. This novel technique has the potential for use in clinical diagnostics and treatment monitoring.

Entities:  

Mesh:

Year:  2012        PMID: 22159012      PMCID: PMC3292356          DOI: 10.1167/iovs.11-8281

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  49 in total

1.  Confocal Brillouin microscopy for three-dimensional mechanical imaging.

Authors:  Giuliano Scarcelli; Seok Hyun Yun
Journal:  Nat Photonics       Date:  2007-12-09       Impact factor: 38.771

2.  Regional elastic performance of the human cornea.

Authors:  J O Hjortdal
Journal:  J Biomech       Date:  1996-07       Impact factor: 2.712

3.  Dynamic corneal imaging.

Authors:  Günther Grabner; Reinhard Eilmsteiner; Christian Steindl; Josef Ruckhofer; Renzo Mattioli; Wolfgang Husinsky
Journal:  J Cataract Refract Surg       Date:  2005-01       Impact factor: 3.351

4.  Ocular response analyzer measurements in keratoconus with normal central corneal thickness compared with matched normal control eyes.

Authors:  Bruno M Fontes; Renato Ambrósio; Guillermo Coca Velarde; Walton Nosé
Journal:  J Refract Surg       Date:  2010-05-03       Impact factor: 3.573

5.  Uneven swelling of the corneal stroma.

Authors:  Y Kikkawa; K Hirayama
Journal:  Invest Ophthalmol       Date:  1970-10

6.  Biomechanical evidence of the distribution of cross-links in corneas treated with riboflavin and ultraviolet A light.

Authors:  Markus Kohlhaas; Eberhard Spoerl; Thomas Schilde; Gabriele Unger; Christine Wittig; Lutz E Pillunat
Journal:  J Cataract Refract Surg       Date:  2006-02       Impact factor: 3.351

7.  Non-uniform swelling properties of the corneal stroma.

Authors:  D Lee; G Wilson
Journal:  Curr Eye Res       Date:  1981       Impact factor: 2.424

8.  Three-dimensional analysis of collagen lamellae in the anterior stroma of the human cornea visualized by second harmonic generation imaging microscopy.

Authors:  Naoyuki Morishige; Yuki Takagi; Tai-ichiro Chikama; Atsushi Takahara; Teruo Nishida
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-02-16       Impact factor: 4.799

9.  Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced cross-linking.

Authors:  Gregor Wollensak; Eberhard Spoerl; Theo Seiler
Journal:  J Cataract Refract Surg       Date:  2003-09       Impact factor: 3.351

Review 10.  The cornea is not a piece of plastic.

Authors:  C Roberts
Journal:  J Refract Surg       Date:  2000 Jul-Aug       Impact factor: 3.573

View more
  89 in total

1.  Shear Brillouin light scattering microscope.

Authors:  Moonseok Kim; Sebastien Besner; Antoine Ramier; Sheldon J J Kwok; Jeesoo An; Giuliano Scarcelli; Seok Hyun Yun
Journal:  Opt Express       Date:  2016-01-11       Impact factor: 3.894

2.  Imaging Corneal Biomechanical Responses to Ocular Pulse Using High-Frequency Ultrasound.

Authors:  Elias Pavlatos; Hong Chen; Keyton Clayson; Xueliang Pan; Jun Liu
Journal:  IEEE Trans Med Imaging       Date:  2018-02       Impact factor: 10.048

3.  Serial biomechanical comparison of edematous, normal, and collagen crosslinked human donor corneas using optical coherence elastography.

Authors:  Matthew R Ford; Abhijit Sinha Roy; Andrew M Rollins; William J Dupps
Journal:  J Cataract Refract Surg       Date:  2014-04-24       Impact factor: 3.351

4.  Spatially heterogeneous corneal mechanical responses before and after riboflavin-ultraviolet-A crosslinking.

Authors:  Joel R Palko; Junhua Tang; Benjamin Cruz Perez; Xueliang Pan; Jun Liu
Journal:  J Cataract Refract Surg       Date:  2014-04-18       Impact factor: 3.351

5.  Single etalon design for two-stage cross-axis VIPA spectroscopy.

Authors:  Antonio Fiore; Giuliano Scarcelli
Journal:  Biomed Opt Express       Date:  2019-02-28       Impact factor: 3.732

6.  Assessing the biomechanical properties of the porcine crystalline lens as a function of intraocular pressure with optical coherence elastography.

Authors:  Chen Wu; Salavat R Aglyamov; Zhaolong Han; Manmohan Singh; Chih-Hao Liu; Kirill V Larin
Journal:  Biomed Opt Express       Date:  2018-11-26       Impact factor: 3.732

7.  Integration of spectral coronagraphy within VIPA-based spectrometers for high extinction Brillouin imaging.

Authors:  Eitan Edrei; Malte C Gather; Giuliano Scarcelli
Journal:  Opt Express       Date:  2017-03-20       Impact factor: 3.894

8.  Optical coherence elastography for evaluating customized riboflavin/UV-A corneal collagen crosslinking.

Authors:  Manmohan Singh; Jiasong Li; Srilatha Vantipalli; Zhaolong Han; Kirill V Larin; Michael D Twa
Journal:  J Biomed Opt       Date:  2017-09-01       Impact factor: 3.170

9.  Noninvasive Assessment of Corneal Crosslinking With Phase-Decorrelation Optical Coherence Tomography.

Authors:  Brecken J Blackburn; Shi Gu; Matthew R Ford; Vinícius de Stefano; Michael W Jenkins; William J Dupps; Andrew M Rollins
Journal:  Invest Ophthalmol Vis Sci       Date:  2019-01-02       Impact factor: 4.799

10.  Biomechanical characterization of keratoconus corneas ex vivo with Brillouin microscopy.

Authors:  Giuliano Scarcelli; Sebastien Besner; Roberto Pineda; Seok Hyun Yun
Journal:  Invest Ophthalmol Vis Sci       Date:  2014-06-17       Impact factor: 4.799

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.