High resolution gel-dosimetry by optical-CT and MR scanning.

M Oldham1, J H Siewerdsen, A Shetty, D A Jaffray.   

Abstract

The increased intricacy of Intensity-Modulated-Radiation-Therapy (IMRT) delivery has created the need for a high-resolution 3D-dosimetry (three-dimensional) system capable of measuring and verifying the complex delivery. Present clinical methods are inadequate being restricted to single points (e.g., ion-chambers) or to 2D planes (e.g., film), and are labor intensive. In this paper we show that gel-dosimetry in conjunction with optical-CT scanning can yield maps of dose that are of sufficient accuracy, resolution and precision to allow verification of complex radiosurgery deliveries, and by extension IMRT deliveries. The radiosurgery dose-distribution represents the most challenging case encountered in external beam therapy by virtue of the steep dose-gradients and high resolution of delivery. We characterize the stringent radiosurgery requirements by the RTAP (Resolution-Time-Accuracy-Precision) criteria defined as < or = 1 mm3 spatial resolution, < or = 1 hour imaging time, accurate to within 3%, and within -1% precision. The RTAP criteria is applied to an in-house laser-based optical-CT scanning system presented here, and evaluated using gel-flasks containing BANG3 gel. The same gel flasks were subsequently imaged using the MR imaging protocol recommended by the gel manufacturer, but modified to match as closely as possible the RTAP. The resulting dose-maps demonstrate the high precision (< 1.3% noise at high dose) achievable with optical CT scanning while preserving high spatial resolution (<1 mm3). Using the sequence above, the MR gel-dose maps were found to have poorer precision by a factor of 5, under the strict conditions of the RTAP. The optical CT gel-dosimetry system was further evaluated for the verification of a complex 3-isocenter radiosurgery delivery. In conclusion, this work demonstrates that gel-dosimetry and optical-CT scanning approach an important long-term goal of radiation dosimetry, as specified by the RTAP criteria, and have potential to impact the clinic by improving and facilitating clinical dose verification for the most complex external beam radiation treatments.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11488576     DOI: 10.1118/1.1380430

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  39 in total

1.  Optical-CT gel-dosimetry I: basic investigations.

Authors:  Mark Oldham; Jeffrey H Siewerdsen; Sai Kumar; John Wong; David A Jaffray
Journal:  Med Phys       Date:  2003-04       Impact factor: 4.071

2.  Optical-CT gel-dosimetry. II: Optical artifacts and geometrical distortion.

Authors:  Mark Oldham; Leonard Kim
Journal:  Med Phys       Date:  2004-05       Impact factor: 4.071

3.  Commissioning a small-field biological irradiator using point, 2D, and 3D dosimetry techniques.

Authors:  Joseph Newton; Mark Oldham; Andrew Thomas; Yifan Li; John Adamovics; David G Kirsch; Shiva Das
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

4.  Characterization of a new radiochromic three-dimensional dosimeter.

Authors:  P Y Guo; J A Adamovics; M Oldham
Journal:  Med Phys       Date:  2006-05       Impact factor: 4.071

5.  Optical-CT imaging of complex 3D dose distributions.

Authors:  Mark Oldham; Leonard Kim; Geoffrey Hugo
Journal:  J Phys Conf Ser       Date:  2005-04

6.  Optical-CT scanning of polymer gels.

Authors:  M Oldham
Journal:  J Phys Conf Ser       Date:  2004

7.  A practical three-dimensional dosimetry system for radiation therapy.

Authors:  Pengyi Guo; John Adamovics; Mark Oldham
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

8.  Fast, high-resolution 3D dosimetry utilizing a novel optical-CT scanner incorporating tertiary telecentric collimation.

Authors:  H S Sakhalkar; M Oldham
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

9.  Quality assurance in 3D dosimetry by optical-CT.

Authors:  Pengyi Guo; John Adamovics; Mark Oldham
Journal:  J Phys Conf Ser       Date:  2006-12-01

10.  A dual-purpose CCD based micro-optical-CT scanning system.

Authors:  M Oldham; H Sakhalkar; P Guo
Journal:  J Phys Conf Ser       Date:  2006-12-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.