Validation of Method Performance of pH, PCO2, PO2, Na(+), K(+) of Cobas b121 ABG Analyser.

Sunil Kumar Nanda1, Lopamudra Ray2, Asha Dinakaran3.   

Abstract

INTRODUCTION: The introduction of a new method or new analyser is a common occurrence in clinical biochemistry laboratory. Blood gas measurements and electrolytes are often performed in Point-of-Care (POC) settings. When a new POC analyser is obtained, the performance of the analyser should be evaluated by comparison to the measurements with the reference analyser in the laboratory.
OBJECTIVES: Evaluation of method performance of pH, PCO2, PO2, Na(+), K(+) of cobas b121 ABG analyser. MATERIALS &
METHODS: The evaluation of method performance of pH, PO2, PCO2, Na(+), K(+) of cobas b121 ABG analyser was done by comparing the results of 50 patient samples run on cobas b121 with the results obtained from Rapid lab values (reference analyser). Correlation coefficient was calculated from the results obtained from both the analysers. Precision was calculated by running biorad ABG control samples.
RESULTS: The correlation coefficient values obtained for parameters were close to 1.0 indicating good correlation. The CV obtained for all the parameters were less than 5 indicating good precision.
CONCLUSION: The new ABG analyser, Cobas b121 correlated well with the reference ABG analyser (Rapid Lab) and could be used to run on patient samples.

Entities:  

Keywords:  Blood gas; Electrolytes; Point of care

Year:  2014        PMID: 25120973      PMCID: PMC4129322          DOI: 10.7860/JCDR/2014/9218.4442

Source DB:  PubMed          Journal:  J Clin Diagn Res        ISSN: 0973-709X


  4 in total

1.  Validation of methods performance for routine biochemistry analytes at Cobas 6000 analyzer series module c501.

Authors:  Vesna Supak Smolcic; Lidija Bilic-Zulle; Elizabeta Fisic
Journal:  Biochem Med (Zagreb)       Date:  2011       Impact factor: 2.313

2.  Validation Protocol: First Step of a Lean-Total Quality Management Principle in a New Laboratory Set-up in a Tertiary Care Hospital in India.

Authors:  Barnali Das
Journal:  Indian J Clin Biochem       Date:  2011-02-18

3.  Error tracking in a clinical biochemistry laboratory.

Authors:  Pal Bela Szecsi; Lars Ødum
Journal:  Clin Chem Lab Med       Date:  2009       Impact factor: 3.694

4.  The truth about quality: medical usefulness and analytical reliability of laboratory tests.

Authors:  James O Westgard; Teresa Darcy
Journal:  Clin Chim Acta       Date:  2004-08-02       Impact factor: 3.786

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.