Antonio Buño Soto1, Katell Peoc'h2,3, Tommaso Fasano4, Jorge Diaz-Garzon1, Bernardino González de la Presa5, Valerie Chicha-Cattoir2, Simone Canovi4, Maria Sanz de Pedro1, Nayra Rico5, Tiphaine Robert3, Efrem Bonelli4, Pilar Fernández Calle1, Aurea Mira5, Guillaume Lefevre6, Luigi Vecchia4, Jose Luis Bedini5.
Abstract
Introduction: The Fourth Universal Definition of Myocardial Infarction Global Taskforce recommends the use of high sensitive troponin (hs-Tn) assays in the diagnosis of acute myocardial infarction. We evaluated the analytical performance of the Atellica IM High-sensitivity Troponin I Assay (hs-TnI) (Siemens Healthcare Diagnostics Inc., Tarrytown, USA) and compared its performance to other hs-TnI assays (Siemens Advia Centaur, Dimension Vista, Dimension EXL, and Abbott Architect (Wiesbaden, Germany)) at one or more sites across Europe. Materials and methods: Precision, detection limit, linearity, method comparison, and interference studies were performed according to Clinical and Laboratory Standards Institute protocols. Values in 40 healthy individuals were compared to the manufacturer's cut-offs. Sample turnaround time (TAT) was examined.Entities:
Keywords: analytical techniques and equipment; high-sensitive troponin I; immunoassay; performance evaluation
Mesh:
Substances:
Year: 2022 PMID: 35799986 PMCID: PMC9195607 DOI: 10.11613/BM.2022.020709
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.515
Summary of assay characteristics according to manufacturer
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| Siemens Healthcare Diagnostics Inc. | High-Sensitivity | Atellica IM analyser | 3-site CLIA | 2.5–25,000 | 10 |
| Siemens Healthcare Diagnostics Inc. | High-Sensitivity | Advia Centaur | Same as for hs-TnI assay on Atellica IM analyser above | 2.5-25,000 | 18 |
| Siemens Healthcare Diagnostics Inc. | High-Sensitivity | Dimension Vista | Homogeneous sandwich CLIA based on LOCI® technology. | 3.0–25,000 | 10 |
| Siemens Healthcare Diagnostics Inc. | High-Sensitivity | Dimension EXL | Same as for hs-TnI assay on Dimension Vista | 4.0–25,000 | 10 |
| Abbott | STAT High Sensitive Troponin-1 | Architect | 2-step immunoassay using CMIA technology. | 10–50,000 | 16 |
| CLIA – Chemiluminescent immunoassay technology. LOCI – Luminescent oxygen channeling assay. CMIA – Chemiluminescent microparticle immunoassay. STAT – Short turn around testing. hs-TnI – high sensitive troponin I. | |||||
Repeatability and within-lab (total) imprecision for the hs-TnI assay on the Atellica IM analyser at different sites
|
|
|
|
|
| ||
|---|---|---|---|---|---|---|
|
|
|
|
| |||
| A | SP1 | 11.9 | 0.5 | 4.1 | 0.9 | 7.6 |
| B | SP1 | 10.7 | 0.4 | 3.6 | 0.7 | 6.8 |
| C | SP1 | 11.0 | 0.4 | 3.6 | 0.6 | 5.5 |
| D | SP1 | 11.2 | 0.5 | 4.7 | 0.8 | 6.7 |
| hs-TnI – high sensitive troponin I. SD – standard deviation. CV – coefficient of variation. Conc. – concentration. | ||||||
Correlation and Passing-Bablok regression results for hs-TnI assays on the Atellica IM 1600 Analyser and established systems
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|
| A | Advia Centaur XP‡ | 39 | 11.84–18,397 | 0.88 | 0.86–0.90 | 0.77 | -0.33–2.61 | 0.997 |
| B | Advia Centaur XP* | 87 | 2.21–15,989 | 1.02 | 1.00–1.03 | 0.98 | 0.67–1.41 | 0.999 |
| C | Advia Centaur XP§ | 117 | 2.19–7,223 | 1.06 | 1.04–1.08 | 0.87 | 0.66–1.10 | 0.999 |
| D | Advia Centaur XP§ | 72 | 1.53–20,212 | 1.05 | 1.03–1.06 | -2.31 | -2.70–(-1.63) | 0.999 |
| B | Architect STAT* | 91 | 1.34–15,989 | 0.99 | 0.88–1.14 | 1.46 | 0.28–2.71 | 0.964 |
| C | Dimension Vista† | 121 | 1.49–7,223 | 1.04 | 1.00–1.07 | -1.37 | -1.63–(-0.76) | 0.984 |
| A | Dimension EXL* | 40 | 11.84–19,566 | 0.97 | 0.93–1.01 | -0.24 | -6.12–9.48 | 0.998 |
| CI – coefficient of variation. hs-Tn – high sensitive troponin. *Intercept 95% CI includes value zero and slope 95% CI includes value one; †Intercept 95% CI does not include value zero and slope 95% CI includes value one; ‡Intercept 95% CI includes value zero and slope 95% CI does not include value one; §Intercept 95% CI does not include value zero and slope 95% CI does not include value one | ||||||||
Interference of biotin and haemolysis on the hs-assay on the Atellica IM analyser
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
| Control (100) | 50 | 50.90 | / | Control (1.5) | 50 | 46.86 | / |
| Spike (100) | 50 | 50.38 | -1.03 | Spike (1.5) | 50 | 48.21 | 2.81 |
| Control (1500) | 50 | 48.49 | / | Control (5.0) | 50 | 49.79 | / |
| Spike (1500) | 50 | 48.80 | 0.65 | Spike (5.0) | 50 | 44.70 | -11.39 |
| Control (100) | 100 | 99.19 | / | Control (1.5) | 100 | 96.31 | / |
| Spike (100) | 100 | 97.76 | -1.46 | Spike (1.5) | 100 | 95.86 | -0.46 |
| Control (1500) | 100 | 93.99 | / | Control (5.0) | 100 | 96.43 | / |
| Spike (1500) | 100 | 92.455 | -1.67 | Spike (5.0) | 100 | 94.375 | -2.18 |
| *Hb – Haemoglobin. hs-Tn – high sensitive troponin. conc. – concentration. Recovery was expressed in %. | |||||||
Results of the Atellica IM 1600 turnaround time (TAT) for the hs-TnI assay during testing of 1461 routine and 13 hs-TnI STAT samples
|
|
|
|
|
| |
|---|---|---|---|---|---|
| Mean (min.) | 44.4 | 6.8 | 19.3 | 27.2 | 10.7 |
| SD (min.) | 12.4 | 5.5 | 3.0 | 8.0 | 0.4 |
| P90 (min.) | 63.1 | 14.0 | 23.6 | 38.4 | 11.2 |
| CV (%) | 28 | 80 | 16 | 29 | 4 |
| Minimum | 18.2 | 0.0 | 13.5 | 7.0 | 10.1 |
| Maximum | 74.7 | 29.5 | 29.8 | 44.8 | 15.8 |
| *IOM – input/output module from the APTIO Automation Solution (Siemens Healthcare Diagnostics Inc., Tarrytown, USA) used with the Atellica IM analyser. TAT – Turnaround time. SD – standard deviation. CV – coefficient of variation. Min – minutes. STAT – short turnaround time. hs-TnI – high sensitive troponin I. P90 – 90th percentile. | |||||