Pooneh Saidi1, Mahdi Sadeghi, S Hamed Hosseini, Claudio Tenreiro.
Abstract
This work presents experimental dosimetry results for a new 103Pd brachytherapy seed, in accordance with the AAPMEntities:
Mesh:
Substances:
Year: 2011 PMID: 22089014 PMCID: PMC5718737 DOI: 10.1120/jacmp.v12i4.3581
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Schematic diagram of the IR06‐ seed.
Figure 2Central slabs of Perspex phantoms used for the experimental determination of: (a) radial dose function values, (b) anisotropy function values.
Comparison of dose rate constant, Λ, of the IR06‐ seed with the measured and calculated values of model MED3633, Theragenics200, and Best double‐wall sources.
|
|
|
|
|
|
|---|---|---|---|---|
| IR06‐ | Monte Carlo simulation (MCNP5) | Liquid water |
| (Present work) |
| Monte Carlo simulation (MCNP5) | Perspex |
| (Present work) | |
| TLD dosimetry | Perspex |
| (Present work) | |
| Consensus value | Perspex | 0.690 | (Present work) | |
| Theragenics 200 | Monte Carlo simulation (MCNP5) | Liquid water |
| (Present work) |
| Monte Carlo simulation | Liquid water |
| 19 | |
| TLD dosimetry | Solid water |
| 21 | |
| MED3633 | Monte Carlo simulation | Liquid water |
| 22 |
| TLD dosimetry | Liquid water |
| 20 | |
| Best Double‐wall | TLD dosimetry | Solid water |
| 23 |
| Monte Carlo simulation | Liquid water |
| 23 |
Figure 3Comparison of the measured and the Monte Carlo‐calculated radial dose function of the IR06‐ source in Perspex phantom. The error bars indicate uncertainty in the measurement.
Measured and calculated radial dose function of the IR06‐ source in Perspex phantom and also calculated in water.
|
|
| ||
|---|---|---|---|
|
|
|
|
|
| 0.5 | 1.448 | 1.49 | 1.42 |
| 1 | 1.000 | 1.000 | 1.000 |
| 1.5 | 0.817 | 0.841 | 0.822 |
| 2 | 0.668 | 0.689 | 0.668 |
| 3 | 0.422 | 0.431 | 0.426 |
| 4 | 0.264 | 0.276 | 0.264 |
| 5 | 0.156 | 0.168 | 0.161 |
Figure 4Comparison of the consensus radial dose function value of IR06‐ seeds versus three other available sources.( , )
Measured and calculated anisotropy function, F(r,θ), for IR06‐ seed in Perspex.
|
| |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
| ||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| 0.25 | 0.053 | 0.074 | 0.616 | 0.854 | 0.924 | 0.959 | 0.979 | 0.992 | 0.998 | 1.000 | |||
| 0.5 | 0.132 | 0.165 | 0.472 | 0.701 | 0.845 | 0.927 | 0.964 | 0.985 | 0.996 | 1.000 | |||
| 1 | 0.190 | 0.242 | 0.491 | 0.686 | 0.816 | 0.903 | 0.956 | 0.982 | 0.996 | 1.000 | |||
| 1.5 | 0.229 | 0.281 | 0.512 | 0.692 | 0.813 | 0.899 | 0.952 | 0.982 | 0.995 | 1.000 | 0.209 0.656 | 0.920 | |
| 2 | 0.265 | 0.309 | 0.526 | 0.696 | 0.814 | 0.896 | 0.950 | 0.982 | 0.996 | 1.000 | 0.253 0.675 | 0.921 | |
| 3 | 0.294 | 0.345 | 0.545 | 0.705 | 0.815 | 0.897 | 0.949 | 0.981 | 0.997 | 1.000 | 0.270 0.681 | 0.925 | |
| 4 | 0.319 | 0.369 | 0.558 | 0.709 | 0.818 | 0.895 | 0.948 | 0.979 | 0.995 | 1.000 | |||
| 5 | 0.337 | 0.392 | 0.568 | 0.715 | 0.820 | 0.900 | 0.948 | 0.981 | 0.995 | 1.000 | 0.301 0.682 | 0.905 | |
Figure 5Comparison of the calculated anisotropy function of the IR06‐ seed versus measured values in Perspex phantom at radial distances of: (a) 1.5 cm, (b) 3 cm, (c) 5 cm.
Figure 6Comparison of the consensus anisotropy function values of the IR06‐ seed versus other available sources at 2 cm.( , )