Literature DB >> 24169818

New developed urological protocols for the Uro Dyna-CT reduce radiation exposure of endourological patients below the levels of the low dose standard CT scans.

M-C Rassweiler1, R Banckwitz, C Koehler, B Mueller-Allissat, M-S Michel, A Häcker, M Ritter.   

Abstract

PURPOSE: Cross-sectional imaging by computed tomography (CT) is associated with higher radiation dose compared to plain X-ray. The Uro Dyna-CT provides CT-like images in the endourological operating room. Our aim was to reduce the radiation exposure of endourological patients with the Uro Dyna-CT and optimize the cross-sectional image quality.
MATERIALS AND METHODS: For the hard contrast protocol, two artificial stones were placed in a Rando-Alderson phantom's left kidney region. Relevant parameters of the standard abdomen protocol were changed. After each modification, two urologists subjectively evaluated the image quality. We developed two customized protocols (standard, low-dose) for hard contrast imaging. To optimize the examination protocol for soft tissue imaging a standardized cone beam phantom was used. Parameters of the preset high-resolution protocol were changed to develop a protocol with similar objective image quality but lower radiation dose. To evaluate the effective radiation dose we embedded 129 thermoluminescence dosimeters in the kidney and ureter region of the Rando-Alderson phantom and performed each protocol five times (stone, soft tissue) and ten times (low-dose protocol). Mean effective dose values per 3D-examination were calculated.
RESULTS: We detected a dose area product (DAP) 776.2 (standard) and 163.5 μGym(2) (low-dose) for the stone protocols with an effective dose of 1.96 and 0.33 mSv, respectively. The soft tissue protocol produced a DAP of 5,070 μGym(2) and an effective dose of 7.76 mSv.
CONCLUSION: Our newly developed examination protocols for the Uro Dyna-CT provide CT-like image quality during urological interventions with low radiation dose.

Entities:  

Mesh:

Year:  2013        PMID: 24169818     DOI: 10.1007/s00345-013-1195-z

Source DB:  PubMed          Journal:  World J Urol        ISSN: 0724-4983            Impact factor:   4.226


  20 in total

1.  Laser-guided percutaneous kidney access with the Uro Dyna-CT: first experience of three-dimensional puncture planning with an ex vivo model.

Authors:  M Ritter; M-C Rassweiler; A Häcker; M S Michel
Journal:  World J Urol       Date:  2012-03-04       Impact factor: 4.226

2.  Organ-specific radiation dose rates and effective dose rates during percutaneous nephrolithotomy.

Authors:  Michael E Lipkin; John G Mancini; Greta Toncheva; Agnes J Wang; Colin Anderson-Evans; W Neal Simmons; Michael N Ferrandino; Terry T Yoshizumi; Glenn M Preminger
Journal:  J Endourol       Date:  2011-09-23       Impact factor: 2.942

3.  Thermoluminescence in medical dosimetry.

Authors:  T Rivera
Journal:  Appl Radiat Isot       Date:  2012-04-30       Impact factor: 1.513

Review 4.  The impact of new methods of imaging on radiation dosage delivered to patients.

Authors:  John G Mancini; Michael N Ferrandino
Journal:  Curr Opin Urol       Date:  2010-03       Impact factor: 2.309

5.  Determination of patient radiation dose during ureteroscopic treatment of urolithiasis using a validated model.

Authors:  Michael E Lipkin; Agnes J Wang; Greta Toncheva; Michael N Ferrandino; Terry T Yoshizumi; Glenn M Preminger
Journal:  J Urol       Date:  2012-01-20       Impact factor: 7.450

6.  Efficacy of DynaCT for surgical navigation during complex laparoscopic surgery: an initial experience.

Authors:  Tetsuo Nozaki; Yasuyoshi Fujiuchi; Akira Komiya; Hideki Fuse
Journal:  Surg Endosc       Date:  2012-10-06       Impact factor: 4.584

7.  [The urological Dyna-CT. Urological cross-sectional imaging on a newly developed urological intervention table].

Authors:  M S Michel; M Ritter; S Schönberg; A Häcker; G Weisser
Journal:  Urologe A       Date:  2012-06       Impact factor: 0.639

Review 8.  Evolution of urological imaging.

Authors:  Anton J Bueschen; Mark E Lockhart
Journal:  Int J Urol       Date:  2010-11-29       Impact factor: 3.369

9.  Influence of surgeon's experience on fluoroscopy time during endourological interventions.

Authors:  M Ritter; F Siegel; P Krombach; A Martinschek; C Weiss; A Häcker; A E Pelzer
Journal:  World J Urol       Date:  2012-08-12       Impact factor: 4.226

10.  Dose optimisation for intraoperative cone-beam flat-detector CT in paediatric spinal surgery.

Authors:  Asger Greval Petersen; Søren Eiskjær; Jon Kaspersen
Journal:  Pediatr Radiol       Date:  2012-06-06
View more
  2 in total

1.  Accuracy in detecting and measuring residual fragments with the Uro Dyna-CT.

Authors:  B Meister; M C Rassweiler; C Weiß; A Häcker; M Ritter
Journal:  World J Urol       Date:  2014-08-28       Impact factor: 4.226

2.  Intraoperative cone beam computed tomography for detecting residual stones in percutaneous nephrolithotomy: a feasibility study.

Authors:  R A Kingma; M J H Voskamp; B H J Doornweerd; I J de Jong; S Roemeling
Journal:  Urolithiasis       Date:  2021-03-08       Impact factor: 3.436

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.