Potential underestimation of the internal target volume (ITV) from free-breathing CBCT.

Irina Vergalasova1, Jacqueline Maurer, Fang-Fang Yin.   

Abstract

PURPOSE: Localization prior to delivery of SBRT to free-breathing patients is performed by aligning the planning internal target volume (ITV) from 4DCT with an on-board free-breathing cone-beam CT (FB-CBCT) image. The FB-CBCT image is assumed to also generate an ITV that captures the full range of motion, due to the acquisition spanning multiple respiratory cycles. However, the ITV could potentially be underestimated when the ratio of time spent in inspiration versus time spent in expiration (I/E ratio) deviates from unity. Therefore, the aim of this study was to investigate the effect of variable I/E ratios on the FB ITV generated from a FB-CBCT scan.
METHODS: This study employed both phantom and patient imaging data. For the phantom study, five periodic respiratory cycles were simulated with different I/E ratios. Six patient respiratory cycles with variable I/E ratios were also selected. All profiles were then programmed into a motion phantom for imaging and modified to exhibit three peak-to-peak motion amplitudes (0.5, 1.0, and 2.0 cm). Each profile was imaged using two spherical targets with 1.0 and 3.0 cm diameters. 2D projections were acquired with full gantry rotation of a kiloVoltage (kV) imager mounted onto the gantry of a modem linear accelerator. CBCT images were reconstructed from 2D projections using a standard filtered back-projection reconstruction algorithm. Quantitative analyses for the phantom study included computing the change in contrast along the direction of target motion as well as determining the area (which is proportional to the target volume) inside of the contour extracted using a Canny edge detector. For the patient study, projection data that were previously acquired under an investigational 4D CBCT slow-gantry imaging protocol were used to generate both FB-CBCT and 4D CBCT images. Volumes were then manually contoured from both datasets (using the same window and level) for quantitative comparison.
RESULTS: The phantom study indicated a reduction in contrast at the inferior edge of the ITV (corresponding to inspiration) as the ratio decreased, for both simulated and patient respiratory cycles. For the simulated phantom respiratory cycles, the contrast reduction of the smallest I/E ratio was 27.6% for the largest target with the smallest amplitude and 89.7% for the smallest target with the largest amplitude. For patient respiratory cycles, these numbers were 22.3% and 94.0%, respectively. The extracted area from inside of the target contours showed a decreasing trend as the I/E ratio decreased. In the patient study, the FB-CBCT ITVs of both lung tumors studied were underestimated when compared with their corresponding 4D CBCT ITV. The underestimations found were 40.1% for the smaller tumor and 24.2% for the larger tumor.
CONCLUSIONS: The ITV may be underestimated in a FB-CBCT image when a patient's respiratory pattern is characterized by a disparate length of time spent in inspiration versus expiration. Missing the full target motion information during on-board verification imaging may result in localization errors.

Entities:  

Mesh:

Year:  2011        PMID: 21928643      PMCID: PMC3161505          DOI: 10.1118/1.3613153

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  50 in total

1.  The correlation evaluation of a tumor tracking system using multiple external markers.

Authors:  Hui Yan; Fang-Fang Yin; Guo-Pei Zhu; Munther Ajlouni; Jae Ho Kim
Journal:  Med Phys       Date:  2006-11       Impact factor: 4.071

2.  Four-dimensional cone beam CT with adaptive gantry rotation and adaptive data sampling.

Authors:  Jun Lu; Thomas M Guerrero; Peter Munro; Andrew Jeung; Pai-Chun M Chi; Peter Balter; X Ronald Zhu; Radhe Mohan; Tinsu Pan
Journal:  Med Phys       Date:  2007-09       Impact factor: 4.071

3.  Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position.

Authors:  Thomas G Purdie; Jean-Pierre Bissonnette; Kevin Franks; Andrea Bezjak; David Payne; Fanny Sie; Michael B Sharpe; David A Jaffray
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-02-27       Impact factor: 7.038

4.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

5.  A simulation study of irregular respiratory motion and its dosimetric impact on lung tumors.

Authors:  Y D Mutaf; C J Scicutella; D Michalski; K Fallon; E D Brandner; G Bednarz; M S Huq
Journal:  Phys Med Biol       Date:  2011-01-17       Impact factor: 3.609

6.  Deep-inspiration breath-hold kilovoltage cone-beam CT for setup of stereotactic body radiation therapy for lung tumors: initial experience.

Authors:  Dennis M Duggan; George X Ding; Charles W Coffey; Wyndee Kirby; Dennis E Hallahan; Arnold Malcolm; Bo Lu
Journal:  Lung Cancer       Date:  2006-12-13       Impact factor: 5.705

7.  Internal-external correlation investigations of respiratory induced motion of lung tumors.

Authors:  Dan Ionascu; Steve B Jiang; Seiko Nishioka; Hiroki Shirato; Ross I Berbeco
Journal:  Med Phys       Date:  2007-10       Impact factor: 4.071

8.  Cone-beam CT localization of internal target volumes for stereotactic body radiotherapy of lung lesions.

Authors:  Zhiheng Wang; Q Jackie Wu; Lawrence B Marks; Nicole Larrier; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-12-01       Impact factor: 7.038

9.  Effects of organ motion on IMRT treatments with segments of few monitor units.

Authors:  J Seco; G C Sharp; J Turcotte; D Gierga; T Bortfeld; H Paganetti
Journal:  Med Phys       Date:  2007-03       Impact factor: 4.071

10.  Measurement of the interplay effect in lung IMRT treatment using EDR2 films.

Authors:  Ross I Berbeco; Cynthia J Pope; Steve B Jiang
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

View more
  22 in total

1.  An assessment of cone beam CT in the adaptive radiotherapy planning process for non-small-cell lung cancer patients.

Authors:  Aileen Duffton; Stephen Harrow; Carolynn Lamb; Mark McJury
Journal:  Br J Radiol       Date:  2016-04-07       Impact factor: 3.039

2.  Optimization of acquisition parameters and accuracy of target motion trajectory for four-dimensional cone-beam computed tomography with a dynamic thorax phantom.

Authors:  Yoshinobu Shimohigashi; Fujio Araki; Masato Maruyama; Yuji Nakaguchi; Kengo Nakato; Nozomu Nagasue; Yudai Kai
Journal:  Radiol Phys Technol       Date:  2014-10-07

3.  Evaluating the four-dimensional cone beam computed tomography with varying gantry rotation speed.

Authors:  S A Yoganathan; K J Maria Das; Shajahan Mohamed Ali; Arpita Agarwal; Surendra P Mishra; Shaleen Kumar
Journal:  Br J Radiol       Date:  2016-02-26       Impact factor: 3.039

4.  Evaluation of the effect of respiratory and anatomical variables on a Fourier technique for markerless, self-sorted 4D-CBCT.

Authors:  I Vergalasova; J Cai; W Giles; W P Segars; F F Yin
Journal:  Phys Med Biol       Date:  2013-09-24       Impact factor: 3.609

5.  Dosimetric verification of lung cancer treatment using the CBCTs estimated from limited-angle on-board projections.

Authors:  You Zhang; Fang-Fang Yin; Lei Ren
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

6.  Usefulness of target delineation based on the two extreme phases of a four-dimensional computed tomography scan in stereotactic body radiation therapy for lung cancer.

Authors:  Seong Soon Jang; Gil Ja Huh; Suk Young Park; Po Song Yang; EunYoun Cho
Journal:  Thorac Cancer       Date:  2015-04-24       Impact factor: 3.500

7.  Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition.

Authors:  Andrew P Santoso; Kwang H Song; Yujiao Qin; Stephen J Gardner; Chang Liu; Indrin J Chetty; Benjamin Movsas; Munther Ajlouni; Ning Wen
Journal:  Radiat Oncol       Date:  2016-07-29       Impact factor: 3.481

8.  Registration uncertainties between 3D cone beam computed tomography and different reference CT datasets in lung stereotactic body radiation therapy.

Authors:  Markus Oechsner; Barbara Chizzali; Michal Devecka; Stephanie Elisabeth Combs; Jan Jakob Wilkens; Marciana Nona Duma
Journal:  Radiat Oncol       Date:  2016-10-26       Impact factor: 3.481

9.  Comparison of internal target volumes defined on 3-dimensional, 4-dimensonal, and cone-beam CT images of non-small-cell lung cancer.

Authors:  Fengxiang Li; Jianbin Li; Zhifang Ma; Yingjie Zhang; Jun Xing; Huanpeng Qi; Dongping Shang
Journal:  Onco Targets Ther       Date:  2016-11-17       Impact factor: 4.147

10.  Dosimetric effect of respiratory motion on volumetric-modulated arc therapy-based lung SBRT treatment delivered by TrueBeam machine with flattening filter-free beam.

Authors:  Xiang Li; Yong Yang; Tianfang Li; Kevin Fallon; Dwight E Heron; M Saiful Huq
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.