BACKGROUND: An evaluation of the usefulness of target delineation based only on the two extreme phases of a four-dimensional computed tomography (4D CT) scan in lung stereotactic body radiation therapy (SBRT).Entities:
Keywords: 4D CT; lung cancer; planning target volume; stereotactic body radiation therapy
Year: 2015 PMID: 26273368 PMCID: PMC4448380 DOI: 10.1111/1759-7714.12170
Source DB: PubMed Journal: Thorac Cancer ISSN: 1759-7706 Impact factor: 3.500
Summary of the various ITVs and PTVs
| Target volumes | Definition |
|---|---|
| ITVall | Volume generated by combining the GTVs from all 8 phases of 4D CT |
| ITV2 | Volume generated by combining the GTVs from the 2 extreme phases of 4D CT |
| ITV_CBCT | Volume derived from the GTV on the CBCT image acquired for the first SBRT fraction |
| PTVall_3 mm | Volume generated by adding a 3 mm isotropic setup margin to the ITVall |
| PTV2_3 mm | Volume generated by adding a 3 mm isotropic setup margin to the ITV2 |
| PTVall_5 mm | Volume generated by adding a 5 mm isotropic setup margin to the ITVall |
| PTV2_5 mm | Volume generated by adding a 5 mm isotropic setup margin to the ITV2 |
4D CT, four-dimensional computed tomography; CBCT, cone-beam CT; GTV, gross tumor volume; ITV, internal target volume; PTV, planning target volume; SBRT, stereotactic body radiation therapy.
Comparisons of the target volumes obtained using the different ITV and PTV definitions and the POVs between pairs
|
Absolute volume | ITVall | ITV2 | PTVall_3 mm | PTV2_3 mm | PTVall_5 mm | PTV2_5 mm |
|---|---|---|---|---|---|---|
| All tumors ( | 14.4 ± 9.3 | 12.3 ± 8.1 | 29.2 ± 16.0 | 26.3 ± 14.7 | 41.1 ± 20.7 | 37.5 ± 19.3 |
3D, three dimensional mobility; GTV, gross tumor volume; ITV, internal target volume; POV, percentage of overlap volume, defined as the percentage ratio between the overlapping and encompassing volume; PTV, planning target volume; SD, standard deviation.
Figure 1The relation between planning target volume (PTV)all_3 mm and PTV2_5 mm. (a) Correlations between the percentage of overlap volume (POV) between two PTVs and the mean gross tumor volume (GTV); (b) correlations between the excess volume of PTV2_5 mm with respect to PTVall_3 mm and the mean GTV. Here, r represents the correlation coefficient.
Geometric coverage of the planning ITVs and PTVs with respect to the ITV generated from CBCT
| Target volumes | Percentage of ITV_CBCT missed by target volumes (%, Mean ± SD) | No. of tumors exhibiting any missing volumes with respect to ITV_CBCT |
|---|---|---|
| ITV2 | 19.2 ± 12.3 | 16/17 (94.1%) |
| ITVall | 13.9 ± 11.7 | 16/17 (94.1%) |
| PTV2_3 mm | 2.9 ± 5.5 | 12/17 (70.6%) |
| PTVall_3 mm | 2.4 ± 4.7 | 8/17 (47.1%) |
| PTV2_5 mm | 0.9 ± 2.1 | 6/17 (35.3%) |
| PTVall_5 mm | 0.7 ± 1.6 | 5/17 (29.4%) |
CBCT, cone-beam CT; ITV, internal target volume; PTV, planning target volume; SD, standard deviation.
Figure 2Example of the relations among the planning internal target volume (ITV) and planning target volume (PTV) and the ITV generated from cone-beam computed tomography (CBCT). (a) Transverse view of ITVall (red), ITV_CBCT (green), PTV2_5 mm (blue), and PTVall_5 mm (yellow) projected onto a modified slow CT scan; (b) delineation of ITV_CBCT on a CBCT image; (c and d) sagittal (c) and coronal (d) views of the target volumes. The geometric regions of ITV_CBCT that are missed by ITVall are displayed as green colored regions.
Absolute differences in the lung-dose parameters for the three SBRT plans (n = 17) based on the different PTV definitions
| Parameters | PTVall_5 mm | PTV2_5 mm | PTVall_3 mm |
PTVall_5 mm |
PTVall_5 mm
|
PTV2_5 mm | |||
|---|---|---|---|---|---|---|---|---|---|
| (Mean ± SD) | (Mean ± SD) | (Mean ± SD) | Mean (range) | Mean (range) | Mean (range) | ||||
| MLD (Gy) | 4.34 ± 1.14 | 4.24 ± 1.13 | 3.83 ± 1.06 | 0.10 (0.02–0.55) | 0.000 | 0.51 (0.32–0.76) | 0.000 | 0.41 (−0.04–0.69) | 0.000 |
| V5 (%) | 19.73 ± 5.05 | 19.41 ± 5.08 | 17.89 ± 4.56 | 0.32 (−0.06–2.15) | 0.001 | 1.84 (0.94–3.20) | 0.000 | 1.52 (0.10–3.05) | 0.000 |
| V10 (%) | 14.07 ± 3.83 | 13.81 ± 3.86 | 12.49 ± 3.58 | 0.26 (−0.05–2.06) | 0.001 | 1.58 (0.95–2.61) | 0.000 | 1.32 (−0.07–2.50) | 0.000 |
| V20 (%) | 6.77 ± 2.65 | 6.54 ± 2.63 | 5.61 ± 2.42 | 0.22 (0.04–0.99) | 0.000 | 1.15 (0.52–1.99) | 0.000 | 0.93 (−0.23–1.83) | 0.000 |
MLD, mean lung dose; PTV, planning target volume; SBRT, stereotactic body radiation therapy; V5, V10, and V20, percentage volumes of both lungs minus the PTVs receiving specific doses of 5, 10, and 20 Gy.