Indirect computed tomography venography with a low-tube-voltage technique: reduction in the radiation and contrast material dose--a prospective randomized study.

Seitaro Oda1, Daisuke Utsunomiya, Kazuo Awai, Hiroko Takaoka, Takeshi Nakaura, Kazuhiro Katahira, Syoji Morishita, Tomohiro Namimoto, Yasuyuki Yamashita.   

Abstract

OBJECTIVE: The objective of the study was to investigate the effects of low-tube-voltage computed tomography (CT) venography on qualitative and quantitative image parameters and the radiation dose.
METHODS: Eighty-eight studies on 84 patients underwent pelvic and lower-extremity CT venography under protocol A (standard 120 kV with 150 mL of contrast material, n = 44) or protocol B (80 kV with 100 mL of contrast material, n = 44) on a 64-detector CT scanner. We compared the dose length product in the 2 protocols. Two blinded observers measured CT attenuation in the veins, the image noise, contrast-to-noise ratio, and figure of merit.
RESULTS: The mean dose length product was significantly lower under protocol B than A (603.2 [SD, 67.2] vs 1131.7 [SD, 67.0] mGy × cm) (P < 0.01). Mean CT attenuation of the veins was significantly greater with protocol B (125.3 [SD, 16.2] vs 106.1 [SD, 16.0] Hounsfield units) (P < 0.01), and the mean image noise was also significantly higher under protocol B (6.6 [SD, 0.8] vs 4.9 [SD, 0.7] Hounsfield units) (P < 0.01). There was no difference in contrast-to-noise ratio (P = 0.46). Figure of merit was significantly higher under protocol B (P < 0.01).
CONCLUSIONS: Computed tomography venography with a low-tube-voltage technique allows reducing the radiation dose and the amount of contrast material without image quality degradation.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21926861     DOI: 10.1097/RCT.0b013e31822a563d

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  12 in total

1.  Comparison of the image qualities of filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction for CT venography at 80 kVp.

Authors:  Jin Hyeok Kim; Ki Seok Choo; Tae Yong Moon; Jun Woo Lee; Ung Bae Jeon; Tae Un Kim; Jae Yeon Hwang; Myeong-Ja Yun; Dong Wook Jeong; Soo Jin Lim
Journal:  Eur Radiol       Date:  2015-10-20       Impact factor: 5.315

2.  Image quality and radiation dose of 128-slice dual-source CT venography using low kilovoltage combined with high-pitch scanning and automatic tube current modulation.

Authors:  Chan Kue Park; Ki Seok Choo; Ung Bae Jeon; Seung Kug Baik; Yong Woo Kim; Tae Un Kim; Chang Won Kim; Yeon Ju Jeong; Dong Wook Jeong; Soo Jin Lim
Journal:  Int J Cardiovasc Imaging       Date:  2013-06-09       Impact factor: 2.357

3.  Automatic spectral imaging protocol and iterative reconstruction for radiation dose reduction in typical hepatic hemangioma computed tomography with reduced iodine load: a preliminary study.

Authors:  Wei Li; Aiyin Li; Bin Wang; Xiuyuan Niu; Xin Cao; Xinyi Wang; Hao Shi
Journal:  Br J Radiol       Date:  2018-05-17       Impact factor: 3.039

4.  Impact of dual-energy CT post-processing to differentiate venous thrombosis from iodine flux artefacts.

Authors:  Jakob Weiss; Christoph Schabel; Ahmed E Othman; Georg Bier; Konstantin Nikolaou; Fabian Bamberg; Malte N Bongers
Journal:  Eur Radiol       Date:  2018-06-04       Impact factor: 5.315

5.  Image quality at low tube voltage (70 kV) and sinogram-affirmed iterative reconstruction for computed tomography in infants with congenital heart disease.

Authors:  Motoo Nakagawa; Yoshiyuki Ozawa; Keita Sakurai; Masashi Shimohira; Kazuya Ohashi; Miki Asano; Sachiko Yamaguchi; Yuta Shibamoto
Journal:  Pediatr Radiol       Date:  2015-06-27

6.  CT angiography of the head-and-neck vessels acquired with low tube voltage, low iodine, and iterative image reconstruction: clinical evaluation of radiation dose and image quality.

Authors:  Wei-lan Zhang; Min Li; Bo Zhang; Hai-yang Geng; Yin-qiang Liang; Ke Xu; Song-bai Li
Journal:  PLoS One       Date:  2013-12-05       Impact factor: 3.240

Review 7.  Low radiation dose in computed tomography: the role of iodine.

Authors:  Andrik J Aschoff; Carlo Catalano; Miles A Kirchin; Martin Krix; Thomas Albrecht
Journal:  Br J Radiol       Date:  2017-05-04       Impact factor: 3.039

8.  CT venography for deep vein thrombosis using a low tube voltage (100 kVp) setting could increase venous enhancement and reduce the amount of administered iodine.

Authors:  Eun-Suk Cho; Jae-Joon Chung; Sungjun Kim; Joo Hee Kim; Jeong-Sik Yu; Choon-Sik Yoon
Journal:  Korean J Radiol       Date:  2013-02-22       Impact factor: 3.500

9.  Indirect CT Venography at 80 kVp with Sinogram-Affirmed Iterative Reconstruction Compared to 120 kVp with Filtered Back Projection: Assessment of Image Quality and Radiation Dose.

Authors:  Inyoung Song; Jeong Geun Yi; Jeong Hee Park; Sung Min Ko
Journal:  PLoS One       Date:  2016-09-23       Impact factor: 3.240

10.  Low-tube-voltage combined with adaptive statistical iterative reconstruction-V technique in CT venography of lower limb deep vein thrombosis.

Authors:  Dan Chen; Jiahui Zhou; Peixi Wang; Quanxu Ge; Min Xu; Wei Qiu; Xinnan Li; Xiaodong Wang
Journal:  Sci Rep       Date:  2018-07-24       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.