Literature DB >> 31100485

The effect of manufacturing tolerances on the mechanical environment of taper junctions in modular TKR.

Kyle Snethen1, Jorge Hernandez1, Melinda Harman2.   

Abstract

Taper design is known to influence corrosive behavior in taper junctions used in modular orthopaedic devices. Manufacturing tolerance of bore-cone tapers is a critical design parameter due to the effect on taper fit, but the effect of variations in manufacturing tolerance on the mechanics of taper junctions has not been well characterized, particularly in modular total knee replacement (TKR). The purpose of this study was to investigate the effect of manufacturing tolerance on stress and micromotion of modular TKR taper junctions. A 3D finite element (FE) model of a modular TKR taper junction was developed and assigned elastoplastic material properties. Model taper geometry was varied by perturbing the angle mismatch by 0.05° between ±0.25° and represented expected variation in manufacturing tolerance. Stress and micromotion were calculated during dynamic FE simulations for each taper junction geometry under varying activity loads and material combinations. Although an increase in angle mismatch generally resulted in higher stress and micromotion, plastic material behavior disrupted this trend for larger angle mismatches. Model predictions corresponded with corrosion behavior evident in vitro. If the FE results obtained here apply in vivo, the absence of elastoplastic material properties in a taper model may grossly overestimate the micromotion and underestimate corrosion behavior and ion release. It is recommended that manufacturing tolerances of bore-cone tapers in modular TKR designs should produce angle mismatches within 0.1° at the taper junction.
Copyright © 2019. Published by Elsevier Ltd.

Entities:  

Keywords:  Angle mismatch; Finite element analysis; Manufacturing tolerance; Micromotion; Modular TKR; Stress

Mesh:

Year:  2019        PMID: 31100485      PMCID: PMC7274368          DOI: 10.1016/j.jmbbm.2019.04.018

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  40 in total

1.  New technology for mechanical property improvement of cast Co-Cr-Mo-C surgical implants.

Authors:  R Hollander; J Wulff
Journal:  J Biomed Mater Res       Date:  1975-05

2.  Adverse local tissue response lesion of the knee associated with Morse taper corrosion.

Authors:  William C McMaster; Jay Patel
Journal:  J Arthroplasty       Date:  2012-06-15       Impact factor: 4.757

3.  Stress-enhanced ion release--the effect of static loading.

Authors:  K J Bundy; C J Williams; R E Luedemann
Journal:  Biomaterials       Date:  1991-09       Impact factor: 12.479

4.  Stemmed implants improve stability in augmented constrained condylar knees.

Authors:  Jeremy J Rawlinson; Robert F Closkey; Nicole Davis; Timothy M Wright; Russell Windsor
Journal:  Clin Orthop Relat Res       Date:  2008-08-15       Impact factor: 4.176

5.  Fretting corrosion of CoCrMo and Ti6Al4V interfaces.

Authors:  Viswanathan Swaminathan; Jeremy L Gilbert
Journal:  Biomaterials       Date:  2012-05-09       Impact factor: 12.479

6.  One hundred and fifty years of history of the Morse taper: from Stephen A. Morse in 1864 to complications related to modularity in hip arthroplasty.

Authors:  Philippe Hernigou; Steffen Queinnec; Charles Henri Flouzat Lachaniette
Journal:  Int Orthop       Date:  2013-05-30       Impact factor: 3.075

7.  Influence of assembly procedure and material combination on the strength of the taper connection at the head-neck junction of modular hip endoprostheses.

Authors:  Annelie Rehmer; Nicholas E Bishop; Michael M Morlock
Journal:  Clin Biomech (Bristol, Avon)       Date:  2011-09-08       Impact factor: 2.063

8.  Focal osteolysis at the junctions of a modular stainless-steel femoral intramedullary nail.

Authors:  D M Jones; J L Marsh; J V Nepola; J J Jacobs; A K Skipor; R M Urban; J L Gilbert; J A Buckwalter
Journal:  J Bone Joint Surg Am       Date:  2001-04       Impact factor: 5.284

9.  Total hip arthroplasty head-neck contact mechanics: a stochastic investigation of key parameters.

Authors:  Finn E Donaldson; James C Coburn; Karen Lohmann Siegel
Journal:  J Biomech       Date:  2014-03-05       Impact factor: 2.712

10.  Taper junction failure in large-diameter metal-on-metal bearings.

Authors:  D J Langton; R Sidaginamale; J K Lord; A V F Nargol; T J Joyce
Journal:  Bone Joint Res       Date:  2012-04-01       Impact factor: 5.853

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.