Brian Wang1, Prema Rassiah-Szegedi, Hui Zhao, Y Jessica Huang, Vikren Sarkar, Martin Szegedi, Kristine E Kokeny, Christopher J Anker, Dennis C Shrieve, Bill J Salter.
Abstract
For Stereotactic Body Radiation Therapy (SBRT) treatment of lung and liver, we quantified the differences between two image guidance methods: 4DCT and ExacTrac respiratory-triggered imaging. Five different patients with five liver lesions and one lung lesion for a total of 19 SBRT delivered fractions were studied. For the 4DCT method, a manual registration process was used between the 4DCT image sets from initial simulation and treatment day to determine the required daily image-guided corrections. We also used the ExacTrac respiratory-triggered imaging capability to verify the target positioning, and calculated the differences in image guidance shifts between these two methods. The mean (standard deviation) of the observed differences in image-guided shifts between 4DCT and ExacTrac respiratory-triggered image guidance was left/right (L/R) = 0.4 (2.0) mm, anterior/posterior (A/P) = 1.4 (1.7) mm, superior/inferior (S/I) = 2.2 (2.0) mm, with no difference larger than 5.0 mm in any given direction for any individual case. The largest error occurred in the S/I direction, with a mean of 2.2 mm for the six lesions. This seems reasonable, because respiratory motion and the resulting imaging uncertainties are most pronounced in this S/I direction. Image guidance shifts derived from ExacTrac triggered imaging at two extreme breathing phases (i.e., full exhale vs. full inhale), agreed well (less than 2.0 mm) with each other. In summary, two very promising image guidance methods of 4DCT and ExacTrac respiratory-triggered imaging were presented and the image guidance shifts were comparable for the patients evaluated in this study.Entities:
Mesh:
Year: 2011 PMID: 21844853 PMCID: PMC5718634 DOI: 10.1120/jacmp.v12i3.3429
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 14DCT images: (a) at full exhale respiratory phase on simulation day, and (b) at full exhale respiratory phase on treatment day #4 for one patient (case #6) in this study. The contoured ITV contained the tumor motion envelope on the day of simulation, but the tumor can be seen to have shifted for treatment day #4. Image guidance shifts were carried out to align to the target's day #4 location prior to treatment delivery. In this example, the corrective shifts were 2 mm posteriorly and 9 mm to patient left.
Figure 2One of two ExacTrac images taken at the full exhale reference level (arrow) with three fiducial markers (indicated by circles).
Figure 3Two orthogonal X‐ray images acquired at the full exhale reference level of a liver tumor patient with the implanted fiducial markers before (a) and after (b) fusion. Markers (M1, M2, and M3) were reconstructed from planning CT of the location of the implanted fiducial markers and then were projected onto the X‐ray images.
Mean differences between two image guidance methods: 4DCT and ExacTrac respiratory‐triggered images.
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| 1 | liver |
| 4.1 |
| 5.9 |
| 2 | liver |
| 4.0 |
| 4.2 |
| 3 | liver |
|
| 3.2 | 4.7 |
| 4 | liver |
| 1.9 | 1.7 | 3.6 |
| 5 | liver |
| 3.2 | 0.8 | 4.9 |
| 6 | lung | 1.4 | 0.7 |
| 2.5 |
|
|
|
|
|
| |
; ; .
The directions of ExacTrac image guidance shifts relative to 4DCT shifts were: ; ; .
Figure 4Mean differences between two image guidance methods: 4DCT and ExacTrac respiratory‐triggered images.
Differences between two breathing phases: full exhale and full inhale, determined by ExacTrac respiratory‐triggered images.
|
|
|
|
|
|---|---|---|---|
| case1 : fx3 |
|
|
|
| case3 : fx1 | 0.5 | 0.1 |
|
| case3 : fx2 | 0.7 |
| 0.0 |
| case3 : fx3 | 0.8 | 0.7 | 2.0 |
| case3 : fx4 | 1.0 |
| 0.8 |
| case4 : fx1 | 0.4 |
| 1.2 |
| Mean | 0.3 |
| 0.4 |
; ; .
The directions of ExacTrac image guidance shifts at full exhale phase relative to full inhale phase shifts were: ; ; .
Differences between pre‐ and post‐treatment, determined by ExacTrac respiratory‐triggered images at full exhale phase.
|
|
|
|
|
|---|---|---|---|
| case1 : fx2 |
|
| 0.2 |
| case3 : fx2 |
| 1.3 | 0.1 |
| case4 : fx1 |
| 0.1 |
|
| case6 : fx5 |
|
| 0.0 |
| Mean |
| 0.1 |
|
; ; .
The directions of ExacTrac image guidance shifts at pre‐treatment relative to post‐treatment shifts were: ; ; .