Investigations of interference between electromagnetic transponders and wireless MOSFET dosimeters: a phantom study.

Zhong Su1, Lisha Zhang, V Ramakrishnan, Michael Hagan, Mitchell Anscher.   

Abstract

PURPOSE: To evaluate both the Calypso Systems' (Calypso Medical Technologies, Inc., Seattle, WA) localization accuracy in the presence of wireless metal-oxide-semiconductor field-effect transistor (MOSFET) dosimeters of dose verification system (DVS, Sicel Technologies, Inc., Morrisville, NC) and the dosimeters' reading accuracy in the presence of wireless electromagnetic transponders inside a phantom.
METHODS: A custom-made, solid-water phantom was fabricated with space for transponders and dosimeters. Two inserts were machined with positioning grooves precisely matching the dimensions of the transponders and dosimeters and were arranged in orthogonal and parallel orientations, respectively. To test the transponder localization accuracy with/without presence of dosimeters (hypothesis 1), multivariate analyses were performed on transponder-derived localization data with and without dosimeters at each preset distance to detect statistically significant localization differences between the control and test sets. To test dosimeter dose-reading accuracy with/without presence of transponders (hypothesis 2), an approach of alternating the transponder presence in seven identical fraction dose (100 cGy) deliveries and measurements was implemented. Two-way analysis of variance was performed to examine statistically significant dose-reading differences between the two groups and the different fractions. A relative-dose analysis method was also used to evaluate transponder impact on dose-reading accuracy after dose-fading effect was removed by a second-order polynomial fit.
RESULTS: Multivariate analysis indicated that hypothesis 1 was false; there was a statistically significant difference between the localization data from the control and test sets. However, the upper and lower bounds of the 95% confidence intervals of the localized positional differences between the control and test sets were less than 0.1 mm, which was significantly smaller than the minimum clinical localization resolution of 0.5 mm. For hypothesis 2, analysis of variance indicated that there was no statistically significant difference between the dosimeter readings with and without the presence of transponders. Both orthogonal and parallel configurations had difference of polynomial-fit dose to measured dose values within 1.75%.
CONCLUSIONS: The phantom study indicated that the Calypso System's localization accuracy was not affected clinically due to the presence of DVS wireless MOSFET dosimeters and the dosimeter-measured doses were not affected by the presence of transponders. Thus, the same patients could be implanted with both transponders and dosimeters to benefit from improved accuracy of radiotherapy treatments offered by conjunctional use of the two systems.

Entities:  

Mesh:

Year:  2011        PMID: 21776780      PMCID: PMC3188601          DOI: 10.1118/1.3578602

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

Review 1.  Organ motion and its management.

Authors:  K M Langen; D T Jones
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-05-01       Impact factor: 7.038

Review 2.  A review of prostate motion with considerations for the treatment of prostate cancer.

Authors:  Thomas E Byrne
Journal:  Med Dosim       Date:  2005       Impact factor: 1.482

3.  Preliminary evaluation of implantable MOSFET radiation dosimeters.

Authors:  A S Beddar; M Salehpour; T M Briere; H Hamidian; M T Gillin
Journal:  Phys Med Biol       Date:  2005-01-07       Impact factor: 3.609

4.  Accuracy of a wireless localization system for radiotherapy.

Authors:  James M Balter; J Nelson Wright; Laurence J Newell; Barry Friemel; Steven Dimmer; Yuki Cheng; John Wong; Edward Vertatschitsch; Timothy P Mate
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-03-01       Impact factor: 7.038

Review 5.  Dose-guided radiation therapy with megavoltage cone-beam CT.

Authors:  J Chen; O Morin; M Aubin; M K Bucci; C F Chuang; J Pouliot
Journal:  Br J Radiol       Date:  2006-09       Impact factor: 3.039

6.  Implantable MOSFET detectors: evaluation of a new design.

Authors:  Tina Marie Briere; Michael T Gillin; A Sam Beddar
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

7.  The observed variance between predicted and measured radiation dose in breast and prostate patients utilizing an in vivo dosimeter.

Authors:  Charles W Scarantino; Bradley R Prestidge; Mitchel S Anscher; Carolyn R Ferree; William T Kearns; Robert D Black; Natasha G Bolick; Gloria P Beyer
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-10-01       Impact factor: 7.038

8.  An implantable radiation dosimeter for use in external beam radiation therapy.

Authors:  Charles W Scarantino; David M Ruslander; Christopher J Rini; Gregory G Mann; H Troy Nagle; Robert D Black
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

9.  Influence of intrafraction motion on margins for prostate radiotherapy.

Authors:  Dale W Litzenberg; James M Balter; Scott W Hadley; Howard M Sandler; Twyla R Willoughby; Patrick A Kupelian; Lisa Levine
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-03-20       Impact factor: 7.038

10.  Technical evaluation of radiation dose delivered in prostate cancer patients as measured by an implantable MOSFET dosimeter.

Authors:  Gloria P Beyer; Charles W Scarantino; Bradley R Prestidge; Amir G Sadeghi; Mitchell S Anscher; Moyed Miften; Tammy B Carrea; Marianne Sims; Robert D Black
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-11-01       Impact factor: 7.038

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.