Home tonometry for management of pediatric glaucoma.

Meghan S Flemmons1, Ya-Chuan Hsiao1, Jacqueline Dzau1, Sanjay Asrani1, Sarah Jones1, Sharon F Freedman2.   

Abstract

PURPOSE: To use iCare rebound tonometry in the home setting for documentation of diurnal intraocular pressure (IOP) fluctuations in children.
DESIGN: Nonrandomized, prospective clinical study.
METHODS: Pediatric ophthalmology clinic patients were recruited between October 2009 and February 2010 who were able to cooperate with IOP measurement by iCare rebound tonometry and whose caregiver was willing and able to obtain iCare measurements at home. The child's IOP was measured first by iCare tonometry followed by a second method (Goldmann applanation [GAT]). The caregiver was instructed on the use of the iCare tonometer. The subject's IOP was measured by the caregiver at home at designated time periods for at least 2 consecutive days.
RESULTS: Seventeen children (17 eyes) with known or suspected glaucoma and 11 normal children were included. Excellent reliability was obtained by caregivers in 70% of iCare measurements. Mean difference between iCare and GAT in clinic was 2.0 ± 4.0 mm Hg, P = .08. Daily IOP fluctuation occurred in both subjects with glaucoma and normal subjects. In children with known or suspected glaucoma, relative peak and trough IOPs occurred in the early morning (45%) and late evening (43.5%), respectively. Comparison of the peak IOP measured at home vs in the clinic was >6 mm Hg in 5 of 16 subjects (31%) and affected glaucoma management in several subjects.
CONCLUSIONS: In selected children with glaucoma, home tonometry by iCare rebound tonometry was reliable, easily performed by caregivers, and well tolerated, and offered IOP information valuable in clinical management.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21689808     DOI: 10.1016/j.ajo.2011.03.001

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  11 in total

1.  Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study.

Authors:  Senay Asik Nacaroglu; Emine Seker Un; Mehmet Giray Ersoz; Yelda Tasci
Journal:  Int J Ophthalmol       Date:  2015-06-18       Impact factor: 1.779

2.  [Evaluation of the self-tonometer Icare ONE in comparison to Goldmann applanation tonometry].

Authors:  V Witte; Ä Glass; R Beck; R Guthoff
Journal:  Ophthalmologe       Date:  2012-10       Impact factor: 1.059

3.  Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses.

Authors:  Alexandra Anton; Matthias Neuburger; Daniel Böhringer; Jens F Jordan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-03-28       Impact factor: 3.117

4.  Comparison of IOPen rebound tonometer with Goldmann applanation tonometer at different IOP levels.

Authors:  Fereydoun Farrahi; Farideh Sharifipour; Mohammad Malekahmadi; Bahman Cheraghian
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

5.  Intraocular pressure in children after congenital heart surgery: A single-center study.

Authors:  Sunali Goyal; Paul H Phillips; Lamonda A Corder; Michael J Robertson; Xiomara Garcia; Michael L Schmitz; Punkaj Gupta
Journal:  Ann Pediatr Cardiol       Date:  2017 Sep-Dec

6.  Comparison of self-measured diurnal intraocular pressure profiles using rebound tonometry between primary angle closure glaucoma and primary open angle glaucoma patients.

Authors:  Shaoying Tan; Nafees Baig; Linda Hansapinyo; Vishal Jhanji; Shihui Wei; Clement C Tham
Journal:  PLoS One       Date:  2017-03-23       Impact factor: 3.240

7.  Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients.

Authors:  Shaoying Tan; Marco Yu; Nafees Baig; Linda Hansapinyo; Clement C Tham
Journal:  Sci Rep       Date:  2017-02-06       Impact factor: 4.379

Review 8.  Teleglaucoma: ready to go?

Authors:  N G Strouthidis; G Chandrasekharan; J P Diamond; I E Murdoch
Journal:  Br J Ophthalmol       Date:  2014-04-10       Impact factor: 4.638

9.  The measurement of intraocular pressure over positive soft contact lenses by rebound tonometry.

Authors:  Fabrizio Zeri; Mario De Cusatis; Luigi Lupelli; Peter Graham Swann
Journal:  J Optom       Date:  2015-10-21

10.  Inaccuracy of intraocular pressure measurement in congenital corneal opacity: three case reports.

Authors:  Byeong Soo Kang; Jin Wook Jeoung; Joo Youn Oh
Journal:  BMC Ophthalmol       Date:  2020-01-02       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.