Comparison of IOPen rebound tonometer with Goldmann applanation tonometer at different IOP levels.

Fereydoun Farrahi1, Farideh Sharifipour, Mohammad Malekahmadi, Bahman Cheraghian.   

Abstract

AIM: To compare the accuracy of IOPen rebound tonometer with Goldmann applanation tonometer (GAT) in individuals with low, normal and high intraocular pressure (IOP) and to evaluate the effect of central corneal thickness (CCT) on IOP measurements.
METHODS: This cross-sectional study consisted of 159 participants. IOP of one eye of each subject was measured consecutively with IOPen and GAT. Then CCT was measured using an ultrasonic pachymeter. Based on GAT IOP readings, participants were divided into low, normal and high IOP groups. Correlation between tonometers and CCT was calculated by spearman's correlation coefficient. Agreement between tonometers was evaluated using Bland-Altman method.
RESULTS: Non-significant underestimation of IOP by IOPen was observed in low IOP group (Mean difference: 0.20mmHg; P=0.454) and also in normal IOP group (Mean difference: 0.56mmHg; P=0.065). However, IOPen significantly overestimated IOP in high IOP group (Mean difference: 1.06mmHg; P=0.038). The 95% limits of agreement (LoA) width between IOPen and GAT IOPs were 7.84, 8.57 and 14.27mmHg in low, normal and high IOP groups, respectively. Low IOP group had thinner corneas compared to high IOP group (P=0.034). IOP measurements taken by IOPen were not influenced by CCT (P=0.099) while poor correlation between CCT and GAT was found (R=0.17, P=0.032). Using receiver operating characteristic (ROC) curve, cutoff value of 18.75mmHg was determined for IOPen with sensitivity of 98.1 and specificity of 97.2%.
CONCLUSION: Accuracy of IOPen is comparable to GAT in patients with low or normal IOP but IOPen overestimates IOP at high IOP levels. CCT does not affect IOP readings with IOPen.

Entities:  

Keywords:  Goldmann applanation tonometry; IOPen; intraocular pressure; rebound tonometry

Year:  2013        PMID: 24195039      PMCID: PMC3808911          DOI: 10.3980/j.issn.2222-3959.2013.05.15

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  25 in total

1.  Comparation of the new rebound tonometer IOPen and the Goldmann tonometer, and their relationship to corneal properties.

Authors:  J Moreno-Montañés; I Gosende; J Caire; M García-Granero
Journal:  Eye (Lond)       Date:  2010-10-01       Impact factor: 3.775

2.  A comparison of four methods of tonometry: method agreement and interobserver variability.

Authors:  P-A Tonnu; T Ho; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

3.  Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer.

Authors:  Kelechi C Ogbuehi
Journal:  Clin Exp Optom       Date:  2006-09       Impact factor: 2.742

4.  The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat.

Authors:  A I Kontiola; D Goldblum; T Mittag; J Danias
Journal:  Exp Eye Res       Date:  2001-12       Impact factor: 3.467

5.  Contact lens tonometry--application in humans.

Authors:  B Entenmann; Y C Robert; P Pirani; H Kanngiesser; P W Dekker
Journal:  Invest Ophthalmol Vis Sci       Date:  1997-11       Impact factor: 4.799

6.  Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients.

Authors:  Paolo Brusini; Maria Letizia Salvetat; Marco Zeppieri; Claudia Tosoni; Lucia Parisi
Journal:  J Glaucoma       Date:  2006-06       Impact factor: 2.503

7.  Effect of corneal thickness on dynamic contour, rebound, and goldmann tonometry.

Authors:  Jose M Martinez-de-la-Casa; Julian Garcia-Feijoo; Eva Vico; Ana Fernandez-Vidal; Jose M Benitez del Castillo; Mohamed Wasfi; J Garcia-Sanchez
Journal:  Ophthalmology       Date:  2006-09-25       Impact factor: 12.079

8.  Dynamic tonometry.

Authors:  H M Dekking; H D Coster
Journal:  Ophthalmologica       Date:  1967       Impact factor: 3.250

9.  Comparison of the rebound tonometer (ICare) to the applanation tonometer (Tonopen XL) in normotensive dogs.

Authors:  M Leiva; C Naranjo; M T Peña
Journal:  Vet Ophthalmol       Date:  2006 Jan-Feb       Impact factor: 1.644

10.  Comparison of the rebound tonometer with the Goldmann applanation tonometer in glaucoma patients.

Authors:  Afsun Sahin; Leyla Niyaz; Nilgun Yildirim
Journal:  Clin Exp Ophthalmol       Date:  2007 May-Jun       Impact factor: 4.207

View more
  2 in total

1.  Comparison of ICare and IOPen vs Goldmann applanation tonometry according to international standards 8612 in glaucoma patients.

Authors:  Milena Pahlitzsch; Jeanette Brünner; Johannes Gonnermann; Anna-Karina B Maier; Necip Torun; Eckart Bertelmann; Matthias Kj Klamann
Journal:  Int J Ophthalmol       Date:  2016-11-18       Impact factor: 1.779

2.  Does patient comfort influence the choice of tonometer for the measurement of intraocular pressure?

Authors:  Mary O Ugalahi; Mukaila A Seidu; Bolutife A Olusanya; Aderonke M Baiyeroju
Journal:  Int Ophthalmol       Date:  2015-10-16       Impact factor: 2.031

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.