| Literature DB >> 21656261 |
Geoffray Keller1, Olivier Desebbe, Roland Henaine, Jean-Jacques Lehot.
Abstract
Monitoring of cardiac output (CO) in the perioperative period and in seriously ill pediatric patients is of major importance for medical management. Hemodynamic monitoring, using transpulmonary thermodilution (TPTD) via a single thermal indicator injection, allows for measurements of CO, volumetric variables and extravascular lung water (EVLW). We describe and explain the influence of a left-to-right shunt on TPTD curve characteristics and EVLW measurements in a young child undergoing a surgical atrial septal defect repair. We suggest that these specific changes in the TPTD curve and the overestimation of EVLW detected by current device, in absence of gas exchange abnormalities, could be indicators of existing circulatory shunts in pediatric patients.Entities:
Mesh:
Year: 2011 PMID: 21656261 DOI: 10.1007/s10877-011-9281-5
Source DB: PubMed Journal: J Clin Monit Comput ISSN: 1387-1307 Impact factor: 2.502