Accuracy of the transpulmonary ultrasound dilution method for detection of small anatomic shunts.

R Saxena1, N Krivitski, K Peacock, A Durward, J M Simpson, S M Tibby.   

Abstract

The purpose of this study was to investigate the qualitative and quantitative accuracy of transpulmonary ultrasound dilution (UD) (COstatus™, Transonic Systems) for the detection of small anatomic shunts. It was a prospective, observational study in a multi-disciplinary pediatric intensive care unit. Seventy-three critically ill children (67 post cardiac surgery), with a median (IQR) age of 10 (3-50.3) months and a median (IQR) weight of 8 (3.43-13) kg were enrolled. Ultrasound dilution (UD) measurements were performed on patients within 1 h of undergoing two-dimensional echocardiography, which was used as the comparator technique. Shunt was diagnosed by characteristic changes on the UD curve shape, and was considered "test-positive" only if two or more measurements suggested the presence of the shunt. The UD technology also provided an estimate of pulmonary to systemic blood flow ratio (Qp:Qs). 12/73 (16.4 %) patients had a shunt identified by both UD and echocardiography. The overall accuracy (95 % CI) was 86.1 % (75.6-96.6 %), with a sensitivity of 85.7 % (57.2-98.2 %) and specificity of 86.4 % (75.0-94.0 %). The estimated Qp:Qs ranged from 0.7 to 1.4, which was consistent qualitatively with the echocardiographic findings on color flow doppler. Shunt was detected by UD alone in eight children; six of these had clinical conditions known to compromise dilution curve analysis (valve regurgitation, asymmetric pulmonary blood flow). Shunt was detected by echocardiography alone in two children; in both cases the shunt was tiny. UD is an accurate method for the detection of small anatomical shunts, both qualitatively and quantitatively.

Entities:  

Mesh:

Year:  2014        PMID: 25240251     DOI: 10.1007/s10877-014-9618-y

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  26 in total

1.  Relative contribution of blood from each lung to the left-to-right shunt in atrial septal defect; demonstration by indicatordilution technics.

Authors:  H B BURCHELL; P S HETZEL; H J SWAN; E H WOOD
Journal:  Circulation       Date:  1956-08       Impact factor: 29.690

2.  Quantitative estimation by indicator-dilution technics of the contribution of blood from each lung to the left-to-right shunt in atrial septal defect.

Authors:  P S HETZEL; H J SWAN; E H WOOD
Journal:  Circulation       Date:  1956-08       Impact factor: 29.690

3.  Ultrasound dilution: an accurate means of determining cardiac output in children.

Authors:  Ivory Crittendon; William J Dreyer; Jamie A Decker; Jeffrey J Kim
Journal:  Pediatr Crit Care Med       Date:  2012-01       Impact factor: 3.624

4.  Quantitative evaluation of intracardiac shunts from arterial dye dilution curves. Demonstration of very small shunts.

Authors:  G A MOOK; W G ZIJLSTRA
Journal:  Acta Med Scand       Date:  1961-12

5.  Flow-regulated extracorporeal arteriovenous tubing loop for cardiac output measurements by ultrasound velocity dilution: validation in post-cardiac surgery intensive care unit patients.

Authors:  Alexsandr A Eremenko; Perviz N Safarov
Journal:  ASAIO J       Date:  2010 Nov-Dec       Impact factor: 2.872

6.  Cardiac output measurement using an ultrasound dilution method: a validation study in ventilated piglets.

Authors:  Willem P de Boode; Arno F J van Heijst; Jeroen C W Hopman; Ronald B Tanke; Hans G van der Hoeven; Kian D Liem
Journal:  Pediatr Crit Care Med       Date:  2010-01       Impact factor: 3.624

7.  The assessment of operative results in congenital heart disease by intraoperative indicator-dilution curves.

Authors:  A G Morrow; H N Oldham; G M Callard; E Braunwald
Journal:  Circulation       Date:  1966-02       Impact factor: 29.690

8.  Transpulmonary thermodilution in a pediatric patient with an intracardiac left-to-right shunt.

Authors:  Geoffray Keller; Olivier Desebbe; Roland Henaine; Jean-Jacques Lehot
Journal:  J Clin Monit Comput       Date:  2011-06-09       Impact factor: 2.502

9.  Theory and in vitro validation of a new extracorporeal arteriovenous loop approach for hemodynamic assessment in pediatric and neonatal intensive care unit patients.

Authors:  Nikolai M Krivitski; Victor V Kislukhin; Naveen V Thuramalla
Journal:  Pediatr Crit Care Med       Date:  2008-07       Impact factor: 3.624

10.  Comparison of a new cardiac output ultrasound dilution method with thermodilution technique in adult patients under general anesthesia.

Authors:  Masato Tsutsui; Nobuhiro Matsuoka; Takehiko Ikeda; Yoshimitsu Sanjo; Tomiei Kazama
Journal:  J Cardiothorac Vasc Anesth       Date:  2009-05-22       Impact factor: 2.628

View more
  7 in total

Review 1.  Journal of Clinical Monitoring and Computing 2015 end of year summary: cardiovascular and hemodynamic monitoring.

Authors:  Karim Bendjelid; Steffen Rex; Thomas Scheeren; Bernd Saugel
Journal:  J Clin Monit Comput       Date:  2016-02-12       Impact factor: 2.502

2.  Feasibility and Accuracy of Cardiac Right-to-Left-Shunt Detection in Children by New Transpulmonary Ultrasound Dilution Method.

Authors:  Martin Boehne; Mathias Baustert; Verena Paetzel; Dietmar Boethig; Harald Köditz; Nils Dennhardt; Philipp Beerbaum; Harald Bertram
Journal:  Pediatr Cardiol       Date:  2016-11-08       Impact factor: 1.655

3.  Predicting fluid responsiveness in 100 critically ill children: the effect of baseline contractility.

Authors:  Rohit Saxena; Andrew Durward; Sarah Steeley; Ian A Murdoch; Shane M Tibby
Journal:  Intensive Care Med       Date:  2015-09-28       Impact factor: 17.440

Review 4.  Neonatal Hemodynamics: From Developmental Physiology to Comprehensive Monitoring.

Authors:  Sabine L Vrancken; Arno F van Heijst; Willem P de Boode
Journal:  Front Pediatr       Date:  2018-04-05       Impact factor: 3.418

5.  Indirect Calorimetry Overestimates Oxygen Consumption in Young Children: Caution is Advised Using Direct Fick Method as a Reference Method in Cardiac Output Comparison Studies.

Authors:  Theodor S Sigurdsson; Lars Lindberg
Journal:  Pediatr Cardiol       Date:  2019-11-18       Impact factor: 1.655

6.  Estimation of intracardiac shunts in young children with a novel indicator dilution technology.

Authors:  Theodor Skuli Sigurdsson; Lars Lindberg
Journal:  Sci Rep       Date:  2020-01-28       Impact factor: 4.379

7.  Indexing haemodynamic variables in young children.

Authors:  Theodor S Sigurdsson; Lars Lindberg
Journal:  Acta Anaesthesiol Scand       Date:  2020-10-30       Impact factor: 2.105

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.