| Literature DB >> 21587173 |
Ryan L McMahon1, Nicole A Larrier, Q Jackie Wu.
Abstract
We present a technique for planning and verification of craniospinal treatment with the patient in the supine position. Treatment delivery and verification is streamlined through the use of modern imaging techniques. Treatments use two lateral brain fields abutted to a single or pair of posterior spine fields. Treatment delivery is simplified by aligning all isocenters in the anterior-posterior and lateral directions. Patient positioning is accomplished via on-board kV imaging. Verification of field shape and junctions is accomplished with BB placement and MV portal imaging. Daily treatment is simplified by using only longitundinal couch shifts, which are recorded in the patient chart and RV database. The technique is simple to implement in a clinic that is already using a similar beam arrangement with the patient prone. It requires no additional devices to be fabricated (for immobilization or QA), and it takes advantage of all the existing elements of a modern linac.Entities:
Mesh:
Year: 2011 PMID: 21587173 PMCID: PMC5718685 DOI: 10.1120/jacmp.v12i2.3310
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Illustration of isocenter alignment of multiple CSI fields. Isocenters of all fields share the same anterior‐posterior and lateral coordinates. Once the brain isocenter is chosen, the upper spine isocenter is shifted longitudinally by a multiple of 1 cm. The lower spine isocenter is then shifted from the upper spine by another multiple of 1 cm. This alignment requires only longitudinal couch shifts between isocenters for treatment.
Figure 2Illustration of gaps used to verify field junctions. Gap 1 represents the gap between the brain fields and upper spine. Gaps 2–3 are used to verify the abutment of the two spine fields. Gaps 1–2 (using BBs) are only verified on the first treatment day and on days after a gap shift has been performed. Gap 3 is verified on a daily basis.
Figure 3Distribution of measured gap errors for six patients treated with supine technique. Measurements done after planned couch shifts with BB placement described in Section II: a) gap 1: ); b) gap 2: ).