| Literature DB >> 21554671 |
Tamara K Berthoud1, Maria Nelia Manaca, Diana Quelhas, Ruth Aguilar, Caterina Guinovart, Laura Puyol, Arnoldo Barbosa, Pedro L Alonso, Carlota Dobaño.
Abstract
BACKGROUND: Multiplex cytokine profiling systems are useful tools for investigating correlates of protective immunity. Several Luminex and flow cytometry methods are commercially available but there is limited information on the relative performance of different kits. A series of comparison experiments were carried out to determine the most appropriate method for our subsequent studies.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21554671 PMCID: PMC3112452 DOI: 10.1186/1475-2875-10-115
Source DB: PubMed Journal: Malar J ISSN: 1475-2875 Impact factor: 2.979
Number of positive responses detected by the Bio-Rad Plate 1 and Plate 2
| Cytokine | Plate 1 | Plate 2 |
|---|---|---|
| IL-1β | 32/32 | 31/32 |
| IL-2 | 17/32 | 12/32 |
| IL-4 | 26/32 | 21/32 |
| IL-5 | 8/32 | 11/32 |
| IL-6 | 19/32 | 32/32 |
| IL-7 | 3/32 | 4/32 |
| IL-8 | 0/32 | 0/32 |
| IL-10 | 8/32 | 30/32 |
| IL-12p70 | 2/32 | 2/32 |
| IL-13 | 26/32 | 19/32 |
| Il-17 | 25/32 | 1/32 |
| G-CSF | 30/32 | 12/32 |
| GM-CSF | 0/32 | 1/32 |
| IFN-γ | 19/32 | 4/32 |
| MCP-1 | 0/32 | 0/32 |
| MIP-1β | 15/32 | 31/32 |
| TNF | 17/32 | 9/32 |
| Total All Cytokines | 247 | 220 |
Comparisons between two identical plates tested with the Bio-Rad Human 17-plex cytokine panel.
| Cytokine | Number of positive observations | Spearman's Rho | GM (pg/ml) plate 1 | GM (pg/ml) | % difference in GM (95% CI) between 2 plates |
|---|---|---|---|---|---|
| IL-1 β | 31 | 0.88** | 32.47 | 21.34 | 52 (32, 75) |
| IL-2 | 12 | 0.96** | 30.86 | 43.76 | 29 (16, 41) |
| IL-4 | 19 | 0.63** | 1.88 | 1.21 | 55 (35, 78) |
| IL-5 | 7 | 0.71NS | 2.16 | 2.52 | 14 (-23, 40) |
| IL-6 | 19 | 0.95** | 345.36 | 248.76 | 38 (20, 60) |
| IL-10 | 8 | 0.86** | 21.65 | 15.06 | 44 (-19, 155) |
| IL-13 | 18 | 0.86** | 5.30 | 4.49 | 18 (-8, 52) |
| G-CSF | 12 | 0.62* | 118.08 | 51.06 | 131 (92, 179) |
| IFN-γ | 4 | -0.60NS | 85.55 | 103.95 | 18 (-50, 55) |
| MIP-1β | 15 | 0.89** | 156.73 | 59.85 | 162 (117, 217) |
| TNF | 9 | -0.13NS | 61.10 | 40.35 | 51 (2, 126) |
Spearman's Rho was used to test if a significant correlation existed between the samples on the two plates. **= Spearman's Rho test with significance of P < 0.01, * = Spearman's Rho test with significance of P < 0.05, NS = Spearman's Rho test which is non-significant. The cytokines with too few positive observations to enable spearman's Rho tests were not included on this table. GM = geometric mean.
Number of positive responses detected by the Bio-Rad and Invitrogen kits.
| Cytokine | Bio-Rad | Invitrogen | Number (%) of concordant positive observations |
|---|---|---|---|
| IL-1β | 17/17 | 17/17 | 17 (100%) |
| IL-2 | 10/17 | 1/17 | 1 (5.8%) |
| IL-4 | 15/17 | 0/17 | 0 (0%) |
| IL-5 | 5/17 | 0/17 | 0 (0%) |
| IL-6 | 7/17 | 16/17 | 7 (41.2%) |
| IL-8 | 0/17 | 0/17 | 0 (0%) |
| IL-10 | 9/17 | 1/17 | 1 (5.8%) |
| GM-CSF | 1/17 | 13/17 | 1 (5.8%) |
| IFN-γ | 12/17 | 0/17 | 0 (0%) |
| TNF | 12/17 | 9/17 | 9 (52.9%) |
| TOTAL all cytokines | 88 | 57 | |
Comparison of the Bio-Rad and Invitrogen kits.
| Cytokine | Number of positive observations | Spearman's Rho | GM (pg/ml) Invitrogen | GM (pg/ml) Bio-Rad | % difference in GM | Wilcoxon Signed Rank Test |
|---|---|---|---|---|---|---|
| IL-1β | 17 | 0.6348** | 185.86 | 48.51 | 283 (174, 436) | 0.0004 |
| IL-6 | 7 | 0.9286** | 408.19 | 316.03 | 29 (-3, 73) | 0.0008 |
| TNF | 9 | -0.4017NS | 52.02 | 76.48 | 32 (-60,15) | 0.09 |
Number of positive observation concordant between both assays with the Spearman's Rho to test for significant correlations between the assays. **= Spearman's Rho test with significance of P =< 0.01, * = Spearman's Rho test with significance of P =< 0.05, NS = Spearman's Rho test which in non-significant. Geometric means (GM) of both assays with the % difference in the geometric means. The table also shows Wilcoxon Rank Sum Significant differences in the quantity of cytokine detected in samples between the Bio-Rad and Invitrogen assays. GM = geometric mean.
Figure 1Correlations between IL-1, IL-6 and TNF measured by Luminex. Correlation of pg/ml of IL-1 (A), IL-6 (B) and TNF (C) detected between the Bio-Rad and Invitrogen kits. The two kits significantly correlated with one another for the detection of IL-1 (Spearman's Rho test with significance of P < 0.01 = 0.6348) and IL-6 (Spearman's Rho test with significance of P < 0.01 = 0.9286).
Number of positive responses detected with the BD and the Bender MedSystems kits.
| Cytokine | BD CBA | Bender MedSystems |
|---|---|---|
| IFN-γ | 23/73 | 28/73 |
| IL-2 | 19/73 | 29/73 |
| IL-4 | 0/73 | 0/73 |
| IL-6 | 73/73 | 70/73 |
| IL-10 | 49/73 | 20/73 |
| IL-12 | 0/73 | 1/73 |
| TNF | 47/73 | 44/73 |
| TOTAL all cytokines | 204 | 160 |
The two flow cytometry methods were compared using 73 supernatant samples collected from PBMC stimulated with either P. falciparum schizont lysate or iRBC
Comparison of the BD and Bender MedSystems Kits
| Cytokine | Number of positive observations | Spearman's Rho | GM (pg/ml) | GM (pg/ml) | % difference in GM between 2 plates (95% CI) | Wilcoxon signed rank test |
|---|---|---|---|---|---|---|
| IFN-γ | 23 | 0.75** | 80.00 | 145.69 | 45 (62, 21) | 0.043 |
| IL-2 | 13 | 0.86** | 80.87 | 113.10 | 29 (57, 19) | 0.339 |
| IL-6 | 70 | 0.94** | 867.64 | 433.89 | 99.9 (72,133) | <0.0001 |
| IL-10 | 20 | 0.88** | 40.84 | 29.87 | 37 (-5, 96) | 0.002 |
| TNF | 44 | 0.7** | 36.83 | 56.64 | 35 (51, 13) | 0.010 |
Number of positive observation concordant between both assays, with the Spearman's Rho to test for significant correlations between the assays. **= Spearman's Rho test with significance of P < 0.01, * = Spearman's Rho test with significance of P < 0.05, NS = Spearman's Rho test which in non-significant. Geometric means of both assays with the % difference in geometric means. The table also shows Wilcoxon Rank Sum Significant differences in the quantity of cytokine detected in samples between the BD and Bender MedSystems kits. GM = geometric mean.
Figure 2Correlations between IFN-γ, IL-2, TNF, IL-10 and IL-6 measured by Flow Cytometry. Scatter plots to show the correlation between the BD and Bender MedSystems kits for IFN-γ (A), IL-2 (B), TNF (C) IL-10 (D), IL-6 (E). The two kits significantly correlated with one another for the detection of all cytokines (Spearman's Rho test with significance of P < 0.01).
Spearman's Rho correlation between the duplicates within the Bio-Rad, Invitrogen, BD and Bender MedSystems kits.
| Cytokine | Bio-Rad | Invitrogen | BD | Bender MedSystems |
|---|---|---|---|---|
| IFN-γ | 0.97 | 0.77 | 0.99 | 0.96 |
| IL-2 | 0.95 | 0.71 | 0.99 | 0.85 |
| IL-6 | 0.95 | 0.94 | 0.94 | 0.98 |
| IL-10 | 0.98 | 0.90 | 0.89 | 1 |
| TNF | 0.93 | 0.78 | 0.96 | 0.96 |
| IL-4 | 0.96 | 0.83 | No responses | No Responses |
| IL-12p70 | 0.93 | Not Done | No responses | No Responses |
| Mean | 0.96 | 0.95 | 0.95 | 0.82 |
Summary of the four assays
| Measure | Bio-Rad | Invitrogen | BD | Bender MedSystems |
|---|---|---|---|---|
| Cost (1 = highest cost) | 2 | 1 | 3 | 4 |
| Time taken for assay completion | 2 h 10 min | 5 h | 4 h | 4 h |
| Time taken for reading samples | 30 min | 30 min | 2 1/2 h | 2 1/2 h |
| Mean correlation (duplicate wells) | 0.95 | 0.82 | 0.95 | 0.96 |
| Number of cytokines tested | 17 | 10 | 11 | 7 |
| Volume of sample needed | 100 μl | 100 μl | 50 μl | 25 μl |
The kits were compared in terms of cost, time for assay completion, correlation between duplicate wells, variability, number of cytokines detected and sample volume. Cost was based on quotes received from Spanish distributers (2007).