Reliability of tonosafe disposable tonometer prisms: clinical implications from the Veterans Affairs Boston Healthcare System Quality Assurance Study.

V Thomas1, M K Daly, T Cakiner-Egilmez, E Baker.   

Abstract

PURPOSE: Given the Veterans Affairs Boston Healthcare System's recent introduction of single-use Tonosafe disposable tonometer prisms as an alternative to Goldmann applanation tonometers (GATs), this study had two aims: to conduct a large-scale quality assurance trial to assess the reliability of intraocular pressure (IOP) measurements of the Tonosafe disposable tonometer compared with GAT, particularly at extremes of pressure; to evaluate the suitability of Tonosafe disposable tonometer prisms as an acceptable substitute for GATs and for clinic-wide implementation in an academic tertiary referral setting.
METHODS: Ophthalmology resident physicians measured the IOPs of patients in general and specialty eye clinics with the Tonosafe disposable tonometer and GAT. Tonosafe test-retest reliability data were also collected. A retrospective review of patient charts and data analysis were performed to determine the reliability of measurements.
RESULTS: The IOPs of 652 eyes (326 patients) were measured with both GAT and Tonosafe, with a range of 3-34 mm Hg. Linear regression analysis showed R=0.93, slope=0.91, both of which supported the proposed hypothesis, and the y-intercept=-1.05 was significantly different from the hypothesized value. The Tonosafe test-retest repeatability (40 eyes of 40 patients), r=0.977, was very high, which was further supported by linear regression slope=0.993, y-intercept=0.118, and a Tonosafe repeatability coefficient of 2.06, similar to GAT repeatability.
CONCLUSIONS: The IOP measurements by Tonosafe disposable prisms correlated closely with Goldmann measurements, with similar repeated measurement variability to GAT. This suggests that the Tonosafe is an acceptable substitute for GAT to measure IOP in ophthalmology clinic settings.

Entities:  

Mesh:

Year:  2011        PMID: 21455241      PMCID: PMC3171273          DOI: 10.1038/eye.2011.40

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  24 in total

1.  Comparison between standard Goldmann applanation prism and disposable applanation prism in tonometry.

Authors:  S Goel; C Chua; B Dong; M Butcher; F Ahfat; S K Hindi; S Kotta
Journal:  Eye (Lond)       Date:  2004-02       Impact factor: 3.775

2.  Epidemic keratoconjunctivitis in an ophthalmology clinic--California.

Authors: 
Journal:  MMWR Morb Mortal Wkly Rep       Date:  1990-09-07       Impact factor: 17.586

3.  Corneal damage from improperly cleaned tonometer tips.

Authors:  J E Levenson
Journal:  Arch Ophthalmol       Date:  1989-08

4.  Repeated applanation tonometry.

Authors:  R A Moses; C H Liu
Journal:  Am J Ophthalmol       Date:  1968-07       Impact factor: 5.258

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Disinfection of Goldmann tonometers after contamination with hepatitis C virus.

Authors:  W A Segal; J R Pirnazar; M Arens; J S Pepose
Journal:  Am J Ophthalmol       Date:  2001-02       Impact factor: 5.258

7.  Intraocular pressure measurement precision with the Goldmann applanation, dynamic contour, and ocular response analyzer tonometers.

Authors:  Aachal Kotecha; Edward White; Patricio G Schlottmann; David F Garway-Heath
Journal:  Ophthalmology       Date:  2010-02-01       Impact factor: 12.079

8.  A large outbreak of epidemic keratoconjunctivitis: problems in controlling nosocomial spread.

Authors:  D Warren; K E Nelson; J A Farrar; E Hurwitz; J Hierholzer; E Ford; L J Anderson
Journal:  J Infect Dis       Date:  1989-12       Impact factor: 5.226

9.  Epidemic keratoconjunctivitis: report of an outbreak in an ophthalmology practice and recommendations for prevention.

Authors:  J W Buehler; R J Finton; R A Goodman; K Choi; J C Hierholzer; R K Sikes; W R Elsea
Journal:  Infect Control       Date:  1984-08

10.  Accuracy of the Tonosafe disposable tonometer head compared to the Goldmann tonometer alone.

Authors:  Peter Kim; Somsak Lertsumitkul; Meaghan Clark; Leeanne Gardner; Peter Macken
Journal:  Clin Exp Ophthalmol       Date:  2004-08       Impact factor: 4.207

View more
  2 in total

1.  Disposable versus non-disposable tonometer prisms: a UK national survey.

Authors:  Kirti M Jasani; Christine Putri; Amy Pearl; Nayeem Sattar; Karl Mercieca; George Spaeth; Archana Bhan-Bhargava
Journal:  BMJ Open Ophthalmol       Date:  2017-03-09

Review 2.  Guidelines and Recommendations for Tonometry Use during the COVID-19 Era.

Authors:  Enmar M Almazyad; Sally Ameen; Mohammad A Khan; Rizwan Malik
Journal:  Middle East Afr J Ophthalmol       Date:  2020-07-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.