Literature DB >> 11228293

Disinfection of Goldmann tonometers after contamination with hepatitis C virus.

W A Segal1, J R Pirnazar, M Arens, J S Pepose.   

Abstract

PURPOSE: To compare methods of disinfecting Goldmann tonometer tips inoculated with hepatitis C virus.
METHODS: Hepatitis C virus was placed on Goldmann tonometer tips, air dried, and then disinfected by dry gauze wipes, isopropyl alcohol wipes, cold water washes, povidone iodine 10% wipes, and hydrogen peroxide or isopropyl alcohol soaks followed by a cold water wash and dry. Hydrogen peroxide and isopropyl alcohol disinfection techniques followed the Centers for Disease Control and Prevention guidelines for prevention of possible transmission of human immunodeficiency virus (HIV). After disinfection, samples from tonometer tips were amplified by polymerase chain reaction to quantitate the amount of hepatitis C virus RNA remaining.
RESULTS: Percentage of hepatitis C virus RNA remaining after disinfection: dry gauze wipes 95.65%, isopropyl alcohol 5-second wipes 88.91%, cold water wash 4.78%, povidone iodine 10% 5-second wipes 0.72%, hydrogen peroxide soak with cold water wash 0.07%, and isopropyl alcohol soak and cold water wash 0.02%.
CONCLUSIONS: After inoculation of Goldmann tonometer tips with hepatitis C virus, a 5-minute soak in 3% hydrogen peroxide or 70% isopropyl alcohol followed by washing in cold water resulted in the greatest reduction in hepatitis C virus RNA.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11228293     DOI: 10.1016/s0002-9394(00)00784-4

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  4 in total

1.  Reliability of tonosafe disposable tonometer prisms: clinical implications from the Veterans Affairs Boston Healthcare System Quality Assurance Study.

Authors:  V Thomas; M K Daly; T Cakiner-Egilmez; E Baker
Journal:  Eye (Lond)       Date:  2011-04-01       Impact factor: 3.775

2.  Correspondence.

Authors:  Bozho Todorich; Aristomenis Thanos; Yoshihiro Yonekawa; Benjamin J Thomas; Lisa J Faia; Emmanuel Chang; Julia Shulman; Karl R Olsen; Michael P Blair; Michael P Shapiro; Philip Ferrone; Lejla Vajzovic; Cynthia A Toth; Thomas C Lee; Joshua Robinson; Baker Hubbard; Hiroyuki Kondo; Cagri G Besirli; Eric Nudleman; Sui Chien Wong; Shunji Kusaka; Mark Walsh; R V Paul Chan; Audina Berrocal; George Caputo; Timothy G Murray; Jonathan Sears; Roberto Schunemann; Clio A Harper; Andres Kychental; Paola Dorta; Linda A Cernichiaro-Espinosa; Wei-Chi Wu; J Peter Campbell; Maria A Martinez-Castellanos; Hugo Quiroz-Mercado; Hidyuki Hayashi; Polly Quiram; Atchara Amphornphruet; Mary E Hartnett; Irena Tsui; Anna Ells; Vishak John; Darius Moshfeghi; Antonio Capone; Kimberly A Drenser; Michael T Trese
Journal:  Retina       Date:  2017-05       Impact factor: 4.256

3.  Efficacy of detergent and water versus bleach for disinfection of direct contact ophthalmic lenses.

Authors:  Ashkan M Abbey; Ninel Z Gregori; Krishna Surapaneni; Darlene Miller
Journal:  Cornea       Date:  2014-06       Impact factor: 2.651

Review 4.  A Systematic Review Regarding Tonometry and the Transmission of Infectious Diseases.

Authors:  Nicole Atkins; William Hodge; Bruce Li
Journal:  J Clin Med Res       Date:  2018-01-26
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.