CT scanner x-ray spectrum estimation from transmission measurements.

Xinhui Duan1, Jia Wang, Lifeng Yu, Shuai Leng, Cynthia H McCollough.   

Abstract

PURPOSE: In diagnostic CT imaging, multiple important applications depend on the knowledge of the x-ray spectrum, including Monte Carlo dose calculations and dual-energy material decomposition analysis. Due to the high photon flux involved, it is difficult to directly measure spectra from the x-ray tube of a CT scanner. One potential method for indirect measurement involves estimating the spectrum from transmission measurements. The expectation maximization (EM) method is an accurate and robust method to solve this problem. In this article, this method was evaluated in a commercial CT scanner.
METHODS: Two step-wedges (polycarbonate and aluminum) were used to produce different attenuation levels. Transmission measurements were performed on the scanner and the measured data from the scanner were exported to an external computer to calculate the spectra. The EM method was applied to solve the equations that represent the attenuation processes of polychromatic x-ray photons. Estimated spectra were compared to the spectra simulated using a software provided by the manufacturer of the scanner. To test the accuracy of the spectra, a verification experiment was performed using a phantom containing different depths of water. The measured transmission data were compared to the transmission values calculated using the estimated spectra.
RESULTS: Spectra of 80, 100, 120, and 140 kVp from a dual-source CT scanner were estimated. The estimated and simulated spectra were well matched. The differences of mean energies were less than 1 keV. In the verification experiment, the measured and calculated transmission values were in excellent agreement.
CONCLUSIONS: Spectrum estimation using transmission data and the EM method is a quantitatively accurate and robust technique to estimate the spectrum of a CT system. This method could benefit studies relying on accurate knowledge of the x-ray spectra from CT scanner.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21452736      PMCID: PMC3041810          DOI: 10.1118/1.3547718

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  17 in total

1.  Modification and benchmarking of MCNP for low-energy tungsten spectra.

Authors:  J R Mercier; D T Kopp; W D McDavid; S B Dove; J L Lancaster; D M Tucker
Journal:  Med Phys       Date:  2000-12       Impact factor: 4.071

2.  Evaluation of the use of six diagnostic X-ray spectra computer codes.

Authors:  P Meyer; E Buffard; L Mertz; C Kennel; A Constantinesco; P Siffert
Journal:  Br J Radiol       Date:  2004-03       Impact factor: 3.039

3.  Semiempirical model for generating tungsten target x-ray spectra.

Authors:  D M Tucker; G T Barnes; D P Chakraborty
Journal:  Med Phys       Date:  1991 Mar-Apr       Impact factor: 4.071

4.  Calculation of 10 MV x-ray spectra emitted by a medical linear accelerator using the BFGS quasi-Newton method.

Authors:  T Shimozato; K Tabushi; S Kitoh; Y Shiota; C Hirayama; S Suzuki
Journal:  Phys Med Biol       Date:  2006-12-29       Impact factor: 3.609

5.  Spectroscopy of diagnostic x rays by a Compton-scatter method.

Authors:  M Yaffe; K W Taylor; H E Johns
Journal:  Med Phys       Date:  1976 SEP-OCT       Impact factor: 4.071

6.  Flat and curved crystal spectrography for mammographic X-ray sources.

Authors:  C T Chantler; R D Deslattes; A Henins; L T Hudson
Journal:  Br J Radiol       Date:  1996-07       Impact factor: 3.039

7.  Compton spectroscopy in the diagnostic x-ray energy range. I. Spectrometer design.

Authors:  G Matscheko; G A Carlsson
Journal:  Phys Med Biol       Date:  1989-02       Impact factor: 3.609

8.  Compton spectroscopy in the diagnostic x-ray energy range. II. Effects of scattering material and energy resolution.

Authors:  G Matscheko; G A Carlsson; R Ribberfors
Journal:  Phys Med Biol       Date:  1989-02       Impact factor: 3.609

9.  A Compton scattering spectrometer for determining X-ray photon energy spectra.

Authors:  G Matscheko; R Ribberfors
Journal:  Phys Med Biol       Date:  1987-05       Impact factor: 3.609

10.  Computation of bremsstrahlung X-ray spectra and comparison with spectra measured with a Ge(Li) detector.

Authors:  R Birch; M Marshall
Journal:  Phys Med Biol       Date:  1979-05       Impact factor: 3.609

View more
  12 in total

1.  Segmentation-free x-ray energy spectrum estimation for computed tomography using dual-energy material decomposition.

Authors:  Wei Zhao; Lei Xing; Qiude Zhang; Qingguo Xie; Tianye Niu
Journal:  J Med Imaging (Bellingham)       Date:  2017-06-30

2.  Gemstone spectral imaging for measuring adult bone mineral density.

Authors:  Wei-Guang Shao; Dian-Mei Liu
Journal:  Exp Ther Med       Date:  2016-09-01       Impact factor: 2.447

3.  Technical Note: Insertion of digital lesions in the projection domain for dual-source, dual-energy CT.

Authors:  Andrea Ferrero; Baiyu Chen; Zhoubo Li; Lifeng Yu; Cynthia McCollough
Journal:  Med Phys       Date:  2017-04-17       Impact factor: 4.071

4.  Monte Carlo modeling of a conventional X-ray computed tomography scanner for gel dosimetry purposes.

Authors:  Homa Hayati; Asghar Mesbahi; Mahmood Nazarpoor
Journal:  Radiol Phys Technol       Date:  2015-07-25

5.  Physical density estimations of single- and dual-energy CT using material-based forward projection algorithm: a simulation study.

Authors:  Kai-Wen Li; Daiyu Fujiwara; Akihiro Haga; Huisheng Liu; Li-Sheng Geng
Journal:  Br J Radiol       Date:  2021-09-29       Impact factor: 3.039

6.  Lesion insertion in the projection domain: Methods and initial results.

Authors:  Baiyu Chen; Shuai Leng; Lifeng Yu; Zhicong Yu; Chi Ma; Cynthia McCollough
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

7.  Estimating the spectrum in computed tomography via Kullback-Leibler divergence constrained optimization.

Authors:  Wooseok Ha; Emil Y Sidky; Rina Foygel Barber; Taly Gilat Schmidt; Xiaochuan Pan
Journal:  Med Phys       Date:  2018-12-13       Impact factor: 4.071

8.  Semiempirical, parameterized spectrum estimation for x-ray computed tomography.

Authors:  Paul FitzGerald; Stephen Araujo; Mingye Wu; Bruno De Man
Journal:  Med Phys       Date:  2021-03-16       Impact factor: 4.071

9.  Evaluation of energy spectrum CT for the measurement of thyroid iodine content.

Authors:  Weiguang Shao; Jingang Liu; Dianmei Liu
Journal:  BMC Med Imaging       Date:  2016-08-12       Impact factor: 1.930

10.  Cone-Beam CT image contrast and attenuation-map linearity improvement (CALI) for brain stereotactic radiosurgery procedures.

Authors:  SayedMasoud Hashemi; Christopher Huynh; Arjun Sahgal; William Y Song; Håkan Nordström; Markus Eriksson; James G Mainprize; Young Lee; Mark Ruschin
Journal:  J Appl Clin Med Phys       Date:  2018-10-19       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.