Compton spectroscopy in the diagnostic x-ray energy range. I. Spectrometer design.

G Matscheko1, G A Carlsson.   

Abstract

The optimal design of a Compton spectrometer for measuring photon energy spectra from x-ray tubes in a clinical laboratory is analysed. The demands are: (i) coherent and multiple scattering distort the measurements and must be avoided; (ii) the measuring time should be as short as possible to avoid unnecessary wear on the x-ray tube; and (iii) the impairment in energy resolution due to the scattering geometry should be kept minimal. A scattering angle of 90 degrees is advocated. Scatterers (of low-atomic-number material) in the shape of long circular rods (0.5-4 mm diameter, 20-40 mm long) are preferable to scattering foils. Use of a short focus-scatterer distance (approximately 200 mm) is to be preferred compared to using a large detector area (greater than or equal to 4 mm diameter) in order to establish a sufficiently high count rate in the detector. Short focal distances and a 90 degrees scattering angle are advantages in measuring energy spectra in the gantry of CT machines where the available space is limited. To limit the geometrical energy broadening to less than 1 keV, the spread in scattering angles of registered photons must not exceed 1-2 degrees for incident photon energies of 100-150 keV.

Entities:  

Mesh:

Year:  1989        PMID: 2928383     DOI: 10.1088/0031-9155/34/2/003

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  CT scanner x-ray spectrum estimation from transmission measurements.

Authors:  Xinhui Duan; Jia Wang; Lifeng Yu; Shuai Leng; Cynthia H McCollough
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

2.  Experimental separation of diagnostic X-ray spectra into scatter and primary components.

Authors:  H Kubota; Y Ozaki; M Matsumoto; H Kanamori
Journal:  Med Biol Eng Comput       Date:  1994-07       Impact factor: 2.602

3.  Shaping X-ray spectra with filters in X-ray diagnostics.

Authors:  M Sandborg; C A Carlsson; G A Carlsson
Journal:  Med Biol Eng Comput       Date:  1994-07       Impact factor: 2.602

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.