Reduced image noise at low-dose multidetector CT of the abdomen with prior image constrained compressed sensing algorithm.

Meghan G Lubner1, Perry J Pickhardt, Jie Tang, Guang-Hong Chen.   

Abstract

PURPOSE: To assess the effect of prior image constrained compressed sensing (PICCS) on noise reduction and image quality at low-dose computed tomography (CT).
MATERIALS AND METHODS: This HIPAA-compliant institutional review board-approved retrospective study was performed by using DICOM CT colonography data sets obtained in 20 adult patients. Informed consent was waived. Low-dose CT colonography was performed with 64-detector CT by using the standard protocol with mean effective dose per series of 3.06 mSv (range, 1.4-7.7 mSv). PICCS was applied to standard filtered back-projection (FBP) series. For FBP and PICCS series, mean and standard deviation (SD) of attenuation were obtained with 100-mm(2) circular region of interest (ROI) at six sites (240 soft-tissue, colonic gas, and subcutaneous fat measurements). Two abdominal radiologists reviewed two- and three-dimensional CT colonography displays and graded image quality with a five-point scale. Phantom studies were performed to compare spatial resolution and image quality between FBP and PICCS. Mean image noise and image quality scores were calculated and compared for clinical and phantom data sets. Bland-Altman, generalized estimating equation regression model, and Student t tests were used to obtain limits of agreement and to compare noise ratios and subjective image quality.
RESULTS: Mean SD of attenuation (image noise) for ROIs was 38.0 for FBP and 12.2 for PICCS, corresponding to a noise-reduction factor of 3.1 (P < .001). Average noise reduction was 3.3 for soft tissue, 2.8 for air, and 3.0 for fat attenuation. Attenuation did not substantially change between FBP and PICCS images. Average two-dimensional image quality was 2.45 for FBP and 3.4 for PICCS (P < .001). Average three-dimensional image quality at three sites in the colon was 3.5 for FBP and 3.7 for PICCS (P = .34). Phantom data sets revealed no loss of spatial resolution in a line phantom and reduced noise in a liver tumor phantom when PICCS was compared with FBP.
CONCLUSION: Application of PICCS to standard FBP low-dose multidetector CT abdominal images results in substantial noise reduction and improved image quality.

Entities:  

Mesh:

Year:  2011        PMID: 21436086     DOI: 10.1148/radiol.11101380

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  27 in total

1.  Prior image constrained compressed sensing: implementation and performance evaluation.

Authors:  Pascal Thériault Lauzier; Jie Tang; Guang-Hong Chen
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

2.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT.

Authors:  Cynthia H McCollough; Guang Hong Chen; Willi Kalender; Shuai Leng; Ehsan Samei; Katsuyuki Taguchi; Ge Wang; Lifeng Yu; Roderic I Pettigrew
Journal:  Radiology       Date:  2012-06-12       Impact factor: 11.105

Review 3.  Missed lesions at CT colonography: lessons learned.

Authors:  Perry J Pickhardt
Journal:  Abdom Imaging       Date:  2013-02

4.  Sub-milliSievert (sub-mSv) CT colonography: a prospective comparison of image quality and polyp conspicuity at reduced-dose versus standard-dose imaging.

Authors:  Meghan G Lubner; B Dustin Pooler; Douglas R Kitchin; Jie Tang; Ke Li; David H Kim; Alejandro Munoz del Rio; Guang-Hong Chen; Perry J Pickhardt
Journal:  Eur Radiol       Date:  2015-04-23       Impact factor: 5.315

5.  [High-quality reconstruction of four-dimensional cone beam CT from motion registration prior image].

Authors:  Meiling Chen; Yi Huang; Wufan Chen; Xin Chen; Hua Zhang
Journal:  Nan Fang Yi Ke Da Xue Xue Bao       Date:  2019-02-28

6.  Single-scan patient-specific scatter correction in computed tomography using peripheral detection of scatter and compressed sensing scatter retrieval.

Authors:  Bowen Meng; Ho Lee; Lei Xing; Benjamin P Fahimian
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

7.  Characterization of statistical prior image constrained compressed sensing. I. Applications to time-resolved contrast-enhanced CT.

Authors:  Pascal Theriault Lauzier; Guang-Hong Chen
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

8.  Characterization of statistical prior image constrained compressed sensing (PICCS): II. Application to dose reduction.

Authors:  Pascal Theriault Lauzier; Guang-Hong Chen
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

Review 9.  CT colonography for population screening: ready for prime time?

Authors:  Perry J Pickhardt
Journal:  Dig Dis Sci       Date:  2014-12-10       Impact factor: 3.199

10.  PET image reconstruction using kernel method.

Authors:  Guobao Wang; Jinyi Qi
Journal:  IEEE Trans Med Imaging       Date:  2014-07-30       Impact factor: 10.048

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.