Single-scan patient-specific scatter correction in computed tomography using peripheral detection of scatter and compressed sensing scatter retrieval.

Bowen Meng1, Ho Lee, Lei Xing, Benjamin P Fahimian.   

Abstract

PURPOSE: X-ray scatter results in a significant degradation of image quality in computed tomography (CT), representing a major limitation in cone-beam CT (CBCT) and large field-of-view diagnostic scanners. In this work, a novel scatter estimation and correction technique is proposed that utilizes peripheral detection of scatter during the patient scan to simultaneously acquire image and patient-specific scatter information in a single scan, and in conjunction with a proposed compressed sensing scatter recovery technique to reconstruct and correct for the patient-specific scatter in the projection space.
METHODS: The method consists of the detection of patient scatter at the edges of the field of view (FOV) followed by measurement based compressed sensing recovery of the scatter through-out the projection space. In the prototype implementation, the kV x-ray source of the Varian TrueBeam OBI system was blocked at the edges of the projection FOV, and the image detector in the corresponding blocked region was used for scatter detection. The design enables image data acquisition of the projection data on the unblocked central region of and scatter data at the blocked boundary regions. For the initial scatter estimation on the central FOV, a prior consisting of a hybrid scatter model that combines the scatter interpolation method and scatter convolution model is estimated using the acquired scatter distribution on boundary region. With the hybrid scatter estimation model, compressed sensing optimization is performed to generate the scatter map by penalizing the L1 norm of the discrete cosine transform of scatter signal. The estimated scatter is subtracted from the projection data by soft-tuning, and the scatter-corrected CBCT volume is obtained by the conventional Feldkamp-Davis-Kress algorithm. Experimental studies using image quality and anthropomorphic phantoms on a Varian TrueBeam system were carried out to evaluate the performance of the proposed scheme.
RESULTS: The scatter shading artifacts were markedly suppressed in the reconstructed images using the proposed method. On the Catphan©504 phantom, the proposed method reduced the error of CT number to 13 Hounsfield units, 10% of that without scatter correction, and increased the image contrast by a factor of 2 in high-contrast regions. On the anthropomorphic phantom, the spatial nonuniformity decreased from 10.8% to 6.8% after correction.
CONCLUSIONS: A novel scatter correction method, enabling unobstructed acquisition of the high frequency image data and concurrent detection of the patient-specific low frequency scatter data at the edges of the FOV, is proposed and validated in this work. Relative to blocker based techniques, rather than obstructing the central portion of the FOV which degrades and limits the image reconstruction, compressed sensing is used to solve for the scatter from detection of scatter at the periphery of the FOV, enabling for the highest quality reconstruction in the central region and robust patient-specific scatter correction.

Entities:  

Mesh:

Year:  2013        PMID: 23298098      PMCID: PMC3543379          DOI: 10.1118/1.4769421

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  X-ray scatter correction algorithm for cone beam CT imaging.

Authors:  Ruola Ning; Xiangyang Tang; David Conover
Journal:  Med Phys       Date:  2004-05       Impact factor: 4.071

2.  The influence of antiscatter grids on soft-tissue detectability in cone-beam computed tomography with flat-panel detectors.

Authors:  J H Siewerdsen; D J Moseley; B Bakhtiar; S Richard; D A Jaffray
Journal:  Med Phys       Date:  2004-12       Impact factor: 4.071

3.  Scatter correction method for X-ray CT using primary modulation: theory and preliminary results.

Authors:  Lei Zhu; N Robert Bennett; Rebecca Fahrig
Journal:  IEEE Trans Med Imaging       Date:  2006-12       Impact factor: 10.048

4.  Characterization of scattered radiation in kV CBCT images using Monte Carlo simulations.

Authors:  Geneviève Jarry; Sean A Graham; Douglas J Moseley; David J Jaffray; Jeffrey H Siewerdsen; Frank Verhaegen
Journal:  Med Phys       Date:  2006-11       Impact factor: 4.071

5.  Scatter kernel estimation with an edge-spread function method for cone-beam computed tomography imaging.

Authors:  Heng Li; Radhe Mohan; X Ronald Zhu
Journal:  Phys Med Biol       Date:  2008-11-07       Impact factor: 3.609

6.  Scatter correction in cone-beam CT via a half beam blocker technique allowing simultaneous acquisition of scatter and image information.

Authors:  Ho Lee; Lei Xing; Rena Lee; Benjamin P Fahimian
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

7.  An analytical model of the scattered radiation distribution in diagnostic radiology.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1988 Sep-Oct       Impact factor: 4.071

8.  Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimization.

Authors:  Emil Y Sidky; Xiaochuan Pan
Journal:  Phys Med Biol       Date:  2008-08-13       Impact factor: 3.609

9.  A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT.

Authors:  J H Siewerdsen; M J Daly; B Bakhtiar; D J Moseley; S Richard; H Keller; D A Jaffray
Journal:  Med Phys       Date:  2006-01       Impact factor: 4.071

10.  Initial evaluation of coronary images from 320-detector row computed tomography.

Authors:  Frank J Rybicki; Hansel J Otero; Michael L Steigner; Gabriel Vorobiof; Leelakrishna Nallamshetty; Dimitrios Mitsouras; Hale Ersoy; Richard T Mather; Philip F Judy; Tianxi Cai; Karl Coyner; Kurt Schultz; Amanda G Whitmore; Marcelo F Di Carli
Journal:  Int J Cardiovasc Imaging       Date:  2008-03-27       Impact factor: 2.357

View more
  4 in total

1.  Optimization of an on-board imaging system for extremely rapid radiation therapy.

Authors:  Erica M Cherry Kemmerling; Meng Wu; He Yang; Peter G Maxim; Billy W Loo; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

2.  A model-based scatter artifacts correction for cone beam CT.

Authors:  Wei Zhao; Don Vernekohl; Jun Zhu; Luyao Wang; Lei Xing
Journal:  Med Phys       Date:  2016-04       Impact factor: 4.071

3.  Optimization of the geometry and speed of a moving blocker system for cone-beam computed tomography scatter correction.

Authors:  Xi Chen; Luo Ouyang; Hao Yan; Xun Jia; Bin Li; Qingwen Lyu; You Zhang; Jing Wang
Journal:  Med Phys       Date:  2017-09       Impact factor: 4.071

4.  Radiation dose to procedural personnel and patients from an X-ray volume imaging system.

Authors:  Jijo Paul; Emmanuel C Mbalisike; Thomas J Vogl
Journal:  Eur Radiol       Date:  2013-06-29       Impact factor: 5.315

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.