UNLABELLED: Silicon photomultiplier (SiPM; also called a Geiger-mode avalanche photodiode) is a promising semiconductor photosensor in PET and PET/MRI because it is intrinsically MRI-compatible and has internal gain and timing properties comparable to those of a photomultiplier tube. In this study, we have developed a small-animal PET system using SiPMs and lutetium gadolinium oxyorthosilicate (LGSO) crystals and performed physical evaluation and animal imaging studies to show the feasibility of this system.
METHODS: The SiPM PET system consists of 8 detectors, each of which comprises 2 × 6 SiPMs and 4 × 13 LGSO crystals. Each crystal has dimensions of 1.5 × 1.5 × 7 mm. The crystal face-to-face diameter and axial field of view are 6.0 cm and 6.5 mm, respectively. Bias voltage is applied to each SiPM using a finely controlled voltage supply because the gain of the SiPM strongly depends on the supply voltage. The physical characteristics were studied by measuring energy resolution, sensitivity, and spatial resolution. Various mouse and rat images were obtained to study the feasibility of the SiPM PET system in in vivo animal studies. Reconstructed PET images using a maximum-likelihood expectation maximization algorithm were coregistered with animal CT images.
RESULTS: All individual LGSO crystals within the detectors were clearly distinguishable in flood images obtained by irradiating the detector using a (22)Na point source. The energy resolution for individual crystals was 25.8% ± 2.6% on average for 511-keV photopeaks. The spatial resolution measured with the (22)Na point source in a warm background was 1.0 mm (2 mm off-center) and 1.4 mm (16 mm off-center) when the maximum-likelihood expectation maximization algorithm was applied. A myocardial (18)F-FDG study in mice and a skeletal (18)F study in rats demonstrated the fine spatial resolution of the scanner. The feasibility of the SiPM PET system was also confirmed in the tumor images of mice using (18)F-FDG and (68)Ga-RGD and in the brain images of rats using (18)F-FDG.
CONCLUSION: These results indicate that it is possible to develop a PET system using a promising semiconductor photosensor, which yielded reasonable PET performance in phantom and animal studies.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21421723     DOI: 10.2967/jnumed.110.079707

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  12 in total

1.  Development of 1.45-mm resolution four-layer DOI-PET detector for simultaneous measurement in 3T MRI.

Authors:  Fumihiko Nishikido; Atsushi Tachibana; Takayuki Obata; Naoko Inadama; Eiji Yoshida; Mikio Suga; Hideo Murayama; Taiga Yamaya
Journal:  Radiol Phys Technol       Date:  2014-10-28

2.  Small animal, positron emission tomography-magnetic resonance imaging system based on a clinical magnetic resonance imaging scanner: evaluation of basic imaging performance.

Authors:  Raymond R Raylman; Patrick Ledden; Alexander V Stolin; Bob Hou; Ganghadar Jaliparthi; Peter F Martone
Journal:  J Med Imaging (Bellingham)       Date:  2018-09-08

3.  Advances in imaging instrumentation for nuclear cardiology.

Authors:  Jae Sung Lee; Gil Kovalski; Tali Sharir; Dong Soo Lee
Journal:  J Nucl Cardiol       Date:  2017-07-17       Impact factor: 5.952

4.  Reaching 200-ps timing resolution in a time-of-flight and depth-of-interaction positron emission tomography detector using phosphor-coated crystals and high-density silicon photomultipliers.

Authors:  Sun Il Kwon; Alessandro Ferri; Alberto Gola; Eric Berg; Claudio Piemonte; Simon R Cherry; Emilie Roncali
Journal:  J Med Imaging (Bellingham)       Date:  2016-11-23

Review 5.  Advances in PET/MR instrumentation and image reconstruction.

Authors:  Jorge Cabello; Sibylle I Ziegler
Journal:  Br J Radiol       Date:  2016-07-22       Impact factor: 3.039

6.  After-pulsing, cross-talk, dark-count, and gain of MPPC under 7-T static magnetic field.

Authors:  Yoshiyuki Hirano; Fumihiko Nishikido; Daisuke Kokuryo; Taiga Yamaya
Journal:  Radiol Phys Technol       Date:  2016-05-17

7.  Realization of a time-correlated photon counting technique for fluorescence analysis.

Authors:  Lei Dai; Jian Liu; Kun Liang; Ru Yang; Dejun Han; Bo Lu
Journal:  Biomed Opt Express       Date:  2020-03-25       Impact factor: 3.732

8.  A time-based single transmission-line readout with position multiplexing.

Authors:  Minseok Yi; Jae Sung Lee
Journal:  Biomed Eng Lett       Date:  2022-01-17

9.  Simulating Silicon Photomultiplier Response to Scintillation Light.

Authors:  Abhinav K Jha; Herman T van Dam; Matthew A Kupinski; Eric Clarkson
Journal:  IEEE Trans Nucl Sci       Date:  2013-02       Impact factor: 1.679

10.  Performance evaluation of the MOLECUBES β-CUBE-a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors.

Authors:  Srilalan Krishnamoorthy; Eric Blankemeyer; Pieter Mollet; Suleman Surti; Roel Van Holen; Joel S Karp
Journal:  Phys Med Biol       Date:  2018-07-27       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.