Literature DB >> 26627204

Accuracy of intracranial pressure monitoring: systematic review and meta-analysis.

Lucia Zacchetti1, Sandra Magnoni2, Federica Di Corte3, Elisa R Zanier4, Nino Stocchetti5,6.   

Abstract

INTRODUCTION: Intracranial pressure (ICP) measurement is used to tailor interventions and to assist in formulating the prognosis for traumatic brain injury patients. Accurate data are therefore essential. The aim of this study was to verify the accuracy of ICP monitoring systems on the basis of a literature review.
METHODS: A PubMed search was conducted from 1982 to 2014, plus additional references from the selected papers. Accuracy was defined as the degree of correspondence between the pressure read by the catheter and a reference "real" ICP measurement. Studies comparing simultaneous readings from at least two catheters were included. Drift was defined as the loss of accuracy over the monitoring period. Meta-analyses of data from the studies were used to estimate the overall mean difference between simultaneous ICP measurements and their variability. Individual studies were weighted using both a fixed and a random effects model.
RESULTS: Of 163 articles screened, 83 compared two intracranial catheters: 64 reported accuracy and 37 drift (some reported both). Of these, 10 and 17, respectively, fulfilled the inclusion criteria for accuracy and zero drift analysis. The combined mean differences between probes were 1.5 mmHg (95 % confidence interval (CI) 0.7-2.3) with the random effects model and 1.6 mmHg (95 % CI 1.3-1.9) with the fixed effects model. The reported mean drift over a long observation period was 0.75 mmHg. No relation was found with the duration of monitoring or differences between various probes.
CONCLUSIONS: This study confirms that the average error between ICP measures is clinically negligible. The random effects model, however, indicates that a high percentage of readings may vary over a wide range, with clinical implications both for future comparison studies and for daily care.

Entities:  

Mesh:

Year:  2015        PMID: 26627204      PMCID: PMC4667503          DOI: 10.1186/s13054-015-1137-9

Source DB:  PubMed          Journal:  Crit Care        ISSN: 1364-8535            Impact factor:   9.097


Introduction

Intracranial pressure (ICP) monitoring is widely used in neurointensive care, especially for the management of patients with traumatic brain injury (TBI). ICP levels are used to decide interventions, to verify the efficacy of therapeutic maneuvers, and in formulating a prognosis. For all these purposes accurate measurements are essential. Clinicians rely on numbers provided by different methods, and generally believe that those numbers reflect actual ICP with a high degree of accuracy. For instance, a recent trial on decompressive craniectomy randomized patients to different treatments when ICP exceeded 20 mmHg [1], assuming that the sensors used ensured a clear-cut threshold. In real life, however, ICP measurement is challenging. A basic assumption is that the ICP read at one point in the intracranial system properly reflects the average pressure throughout all other locations. Under normal conditions the cerebrospinal fluid (CSF) pressure should in fact be in equilibrium in the whole of the CSF space. Under pathological conditions (as when CSF circulation is obstructed or intracranial masses are expanding), however, ICP cones can develop leading to uneven ICP levels [2, 3]. Even when ICP is homogeneous in the whole intracranial space measurements may not all be accurate, depending on the devices used and the duration of the measurement [4-6]. Our hypothesis is that clinically used systems measure ICP with different degrees of accuracy, not always providing a precise measure. Therefore, our goal was to evaluate the accuracy and precision over time (drift) of different ICP measurements on the basis of a review of the literature.

Methods

Literature search

A PubMed (http://www.ncbi.nlm.nih.gov/pubmed) search was made, looking for studies published from January 1982 through November 2014. The following keywords were searched: “intracranial pressure” AND “Camino”, “intracranial pressure” AND “Codman”, “intracranial pressure” AND “fluid filled”, “intracranial pressure” AND “tip transducer”, “intracranial pressure” AND “drift”. The search was limited to articles in English and animal studies were included at this stage. Additional articles were screened through the references listed in the papers initially selected.

Accuracy

Accuracy was defined as the degree of correspondence between the pressure read by the catheter and a reference “real” ICP measurement. In order to assess accuracy in vivo, studies comparing simultaneous readings from at least two catheters were included, divided into two groups: Group 1: studies comparing a ventricular fluid-coupled catheter with external transducer (VFC) with another type of device. According to the guidelines of the Brain Trauma Foundation [7], this is the most accurate method for measuring ICP so we took it as the reference standard. Group 2: studies comparing catheters other than VFC. Some articles included substudies with different sets of paired readings from probes. These were analyzed as separate studies. Studies on intracranial probes located in the epidural, subdural, or subarachnoid space or comparing infratentorial versus supratentorial probes were excluded from the analysis. Studies were screened for numerical data, and only those reporting the limits of agreement and bias (or data from which they could be calculated) were included. Animal studies were not included in this analysis.

Zero drift

Drift was defined as the loss of accuracy over the monitoring time, a problem related to catheters such as intraparenchymal probes that cannot be re-zeroed during monitoring. In the selected papers drift was assessed by two methods: a) verifying the pressure read by the catheter at atmospheric pressure once it was removed from the patient [8, 9]; and b) measuring pressure changes over time in an artificial hydraulic model [10, 11]. Animal studies were included in this analysis.

Statistics

Mean ICP differences between the various probes, standard deviation and limits of agreement were reported in most papers. If these summary measures were not available, but mean ICP readings per patient were presented, as read with different probes, we calculated the mean ICP differences and standard deviations (Table 1). Studies for which the results were only given in plots or summary tables were not suitable for the planned analysis, and therefore they were not included on the basis of inadequate data reporting (Fig. 1). The authors of five relevant papers reporting incomplete information [12-16] were asked to provide the original data, but this was not accessible (mainly because it was collected decades ago), and these studies were not included either.
Table 1

Main features of the studies included in the meta-analysis

Author/year of publicationProbe 1Probe 2Probe placementPatients (n)Calibration/zeroinga Mean differenceStandard deviationCorrelationDatab
Group 1
1Schickner 1992 [4]CaminoVFCP–V10+/+9.207.80NAReported
2Chambers 1993 [21]CaminoVFCV–V10+/+ (EAM) 1.432.260.98Calculated
3Statham 1993 [22]CaminoVFCP–V11+/+2.734.220.98Reported
4Gopinath 1995 [23]CodmanVFCV–V25+/+0.502.600.97Reported
5Signorini 1998-2 [24]CodmanVFCP–V2+/+ (EAM) 1.813.75NACalculated
6Chambers 2001 [25]SpiegelbergVFCP–V11+/+ (EAM) 0.104.99NAReported
7Koskinen 2005 [8]CodmanVFCP–V22+/+ (EAM) 1.203.320.79Reported
8Lescot 2011-1 [26]PressioVFCP–V15+/+ (EAM) –0.603.83NAReported
9Lescot 2011-2 [26]CodmanVFCP–V15+/+ (EAM) 0.303.52NAReported
10Eide 2012-2 [18]CodmanVFCP–V5NA/NA4.5213.97NACalculated
Group 2
11Signorini 1998-1 [24]CodmanCaminoP–P5+1.583.36NACalculated
12Sahuquillo 1999-1 [2]CaminoCaminoP–P33NA1.801.100.95Reported
13Sahuquillo 1999-2 [2]CaminoCaminoP–P16NA7.506.400.85Reported
14Eide 2012-1 [18]CodmanCodmanP–P5NA0.645.59NACalculated
15Eide 2012-3 [18]CodmanSpiegelbergP–P7+0.702.63NACalculated

Studies comparing simultaneous readings from at least two catheters for intracranial pressure monitoring classified depending on the reported use (Group 1) or not use (Group 2) of a reference standard (i.e.VFC) and reported in chronological order. Some papers reported more than one study; four are cited repeatedly. In two studies a single catheter was placed in the ventricles (V-V). This gave a typical reading through a fluid-filled system and a simultaneous measurement through a solid transducer at the tip

aDescription of calibration (of the solid probe) and zeroing (of the ventricular fluid-filled catheter). (+) Calibration or zeroing is reported, (++) both calibration and zeroing are reported; NA information on calibration and zeroing maneuvers not available, zeroing described at the external auditory meatus

bIndicates how the mean intracranial pressure differences between probes and standard deviations were determined (see Methods)

NA Not Available, P Parenchymal, V Ventricular, VFC Ventricular Fluid-Coupled catheter with external transducer

Fig. 1

Flow diagram summarizing article selection; 83 articles were identified:64 reported accuracy, 37 reported drift (some reported both). Articles on animals, with probes placed in the epidural, subdural, or infratentorial spaces and with inadequate data reporting were excluded. High-quality articles on drift reported the range, mean and median of drift; the 17 papers refer to 20 studies. High-quality articles on accuracy included comparisons of two ICP probes, with Bland-Altman analysis and adequate data reporting; the 10 papers listed refer to 15 studies

Main features of the studies included in the meta-analysis Studies comparing simultaneous readings from at least two catheters for intracranial pressure monitoring classified depending on the reported use (Group 1) or not use (Group 2) of a reference standard (i.e.VFC) and reported in chronological order. Some papers reported more than one study; four are cited repeatedly. In two studies a single catheter was placed in the ventricles (V-V). This gave a typical reading through a fluid-filled system and a simultaneous measurement through a solid transducer at the tip aDescription of calibration (of the solid probe) and zeroing (of the ventricular fluid-filled catheter). (+) Calibration or zeroing is reported, (++) both calibration and zeroing are reported; NA information on calibration and zeroing maneuvers not available, zeroing described at the external auditory meatus bIndicates how the mean intracranial pressure differences between probes and standard deviations were determined (see Methods) NA Not Available, P Parenchymal, V Ventricular, VFC Ventricular Fluid-Coupled catheter with external transducer Flow diagram summarizing article selection; 83 articles were identified:64 reported accuracy, 37 reported drift (some reported both). Articles on animals, with probes placed in the epidural, subdural, or infratentorial spaces and with inadequate data reporting were excluded. High-quality articles on drift reported the range, mean and median of drift; the 17 papers refer to 20 studies. High-quality articles on accuracy included comparisons of two ICP probes, with Bland-Altman analysis and adequate data reporting; the 10 papers listed refer to 15 studies The limits of agreement between simultaneous ICP measurements were calculated using the following formula: where equals the mean difference, s the standard deviation of the differences, and t(0.025, n–1) the value from Student’s t-distribution corresponding to a two-sided alpha of 0.05 and (n–1) degrees of freedom. While still based on the assumption of normal distribution of probe differences around a mean (“bias” or “offset”), this method compensates for the size of the sample, includes an adjustment for the expected differences in the mean, and it is therefore more widely applicable than the large sample approximation of ( ± 1.96 s). Meta-analyses of data from the studies were performed to calculate the overall mean difference between simultaneous ICP measurements and the variability. In calculating the overall mean difference across the combined studies, results from individual studies were weighted, assigning the weights using two standard approaches: a fixed effects model and a random effects model. A fixed effects approach assumes there is one true mean ICP difference between paired probe values. The overall mean difference is calculated by weighting individual study results by the inverse of the observed variance. The random effects approach assumes that the true mean ICP difference between paired probes can vary from one study to another. The overall mean difference is calculated by weighting individual studies according to within-study and between-study variability. Summary statistical analysis was done using the software program SPSS (Version 21, IBM/SPSS, Inc.) and for meta-analyses we used the software package Comprehensive Meta Analysis (Version 2.2.064, Biostat, Inc.).

Results

The literature search identified 163 studies; 80 were excluded because they were not related to the aim of this study. Of the remaining 83, only 64 reported comparisons of two paired intracranial catheters (Fig. 1). After studies based on catheters placed in the epidural space, or comparing the infratentorial and supratentorial spaces, or with inadequate data reporting were excluded, 10 articles were left for accuracy analysis (Fig. 1). They reported 15 studies, 10 in Group 1, and 5 in Group 2. The main features of the studies are reported in Table 1, including details on calibration of solid probes before insertion, and zeroing of ventricular fluid-filled catheters. Calibration and/or VFC zeroing maneuvers were reported in the vast majority of Group 1 studies, whereas in Group 2 these maneuvers were less accurately or not reported at all. ICP recordings for individual studies covered a wide range (up to 100 mmHg) and all studies included cases with ICP >20 mmHg. Limits of agreement for individual studies are summarized in Table 2.
Table 2

Limits of agreement of the studies included in the meta-analysis

Author and year of publicationPatients (n)Lower limit of agreementMean differenceUpper limit of agreement
Group 1
1Schickner 1992 [4]10−9.319.2027.71
2Chambers 1993 [21]10−3.921.436.78
3Statham 1993 [22]11−7.092.7312.55
4Gopinath 1995 [23]25−4.970.505.97
5Signorini 1998-2 [24]5−9.581.8113.21
6Chambers 2001 [25]11−11.510.1011.71
7Koskinen 2005 [8]22−5.861.208.26
8Lescot 2011-1 [26]15−9.08–0.607.88
9Lescot 2011-2 [26]15−7.500.308.10
10Eide 2012-2 [18]5−37.974.5247.01
Group 2
11Signorini 1998-1 [24]5−8.651.5811.82
12Sahuquillo 1999-1 [2]33–0.471.804.07
13Sahuquillo 1999-2 [2]16−6.567.5021.56
14Eide 2012-1 [18]5−16.370.6417.65
15Eide 2012-3 [18]7−6.170.707.57
Limits of agreement of the studies included in the meta-analysis Both positive and negative differences between probes were observed in individual studies. The combined estimate is a positive difference for probe 1 minus probe 2 readings, partly because of the ordering of the probes in the calculation. All comparisons of the two locations in Group 1 were standardized so that the difference was always parenchymal–ventricular, leaving only one study with negative differences (Table 1). For other paired readings between catheters in similar locations (parenchymal–parenchymal or ventricular–ventricular), the catheter with the highest average reading was identified as probe 1, so as to give consistent positive (and not offsetting) mean differences. The combined mean difference between probes, using a fixed model,was 0.9 mmHg, with a 95 % confidence interval (CI) for the mean of 0.4–1.5 mmHg across Group 1 studies and 1.8 mmHg (95 % CI 1.5–2.2 mmHg) across Group 2 studies. The combined mean difference between all probes, assuming the fixed model, was 1.6 mmHg (95 % CI 1.3–1.9 mmHg) (Fig. 2).
Fig. 2

Meta-analysis. Meta-analysis data with the corresponding forest plot for individual data sets (combined Groups 1 and 2). The black horizontal line divides Group 1 (above) from Group 2 (below) studies, as reported in methods. CI Confidence Interval

Meta-analysis. Meta-analysis data with the corresponding forest plot for individual data sets (combined Groups 1 and 2). The black horizontal line divides Group 1 (above) from Group 2 (below) studies, as reported in methods. CI Confidence Interval The combined mean difference between probes, using a random model, was 1.2 mmHg (95 % CI 0.2–2.1 mmHg) in Group 1 studies and 2.3 mmHg (95 % CI 0.5–4.2 mmHg) in Group 2 studies. The combined mean difference between all probes, pooling all studies and assuming the random effects model, was 1.5 mmHg (95 % CI 0.7–2.3 mmHg) (Fig. 2). The zero drift analysis was reported in 37 articles but only 17 had adequate data reporting (Table 3). Although some papers reported a wide range of values, the mean drift over a long observation period was 0.75 mmHg.
Table 3

Main features of the studies reporting drift

Author and year of publicationProbe typeReadings (n)a Test duration (days)ICP range (mmHg)
Clinical studies
1Statham 1993 [22]Fiberoptic111–110/+4
2Bavetta 1997 [27]Fiberoptic831–12−12/+14
3Münch1998 [28]Fiberoptic951–17−15/+12
4Martines-Mañas 2000 [29]Fiberoptic561–12−24/+35
5Poca 2002 [19]Fiberoptic1261–11−12/+7
6Stendel 2003-1 [17]Fiberoptic501–320/+12
7Gelabert-González 2006 [9]Fiberoptic6241–5−17/+21
8Gopinath 1995 [23]Microstrain gauge252–7−2/+2
9Stendel 2003-2 [17]Microstrain gauge983–28−2/+3
10Koskinen 2005 [8]Microstrain gauge1281–16−5/+4
11Citerio 2008 [30]Microstrain gauge891–10−4/+8
12Al-Tamimi 2009 [20]Microstrain gauge883; 6b NA
13Lang 2003 [31]Microstrain gauge843–28−2/+2
Laboratory studies
14Czosnyka 1996-1 [32]Fiberoptic13−0.8/+0.8
15Czosnyka 1996-2 [32]Fiberoptic13−0.4/+0.4
16Piper 2001 [33]Fiberoptic341–12−13/+22
17Sundbӓrg 1987 [12]Microstrain gauge13−2/+2
18Czosnyka 1996-3 [32]Microstrain gauge13−0.8/+0.8
19Morgalla 1999 [10]Microstrain gauge710−4/+3
20Citerio 2004 [11]Microstrain gauge1050/+2

Studies reporting drift were classified according to the setting (clinical or laboratory). Some papers reported more than one study; one paper is cited three times. Since different probes were analyzed in some studies, a single reference may appear more than once

aIndicates how many readings of drift were used to calculate the range, mean and/or median (see text for more details)

bThe study by Al-Tamimi et al. included data from two centers and reported the median observation time for both

ICP Intracranial pressure, NA Not available

Main features of the studies reporting drift Studies reporting drift were classified according to the setting (clinical or laboratory). Some papers reported more than one study; one paper is cited three times. Since different probes were analyzed in some studies, a single reference may appear more than once aIndicates how many readings of drift were used to calculate the range, mean and/or median (see text for more details) bThe study by Al-Tamimi et al. included data from two centers and reported the median observation time for both ICP Intracranial pressure, NA Not available No differences were detected between fiberoptic probes and microstrain-gauge probes (mean fiberoptic 0.59 ± 1.8; mean microstrain-gauge 0.95 ± 0.23; p = 0.8), or between data from clinical and laboratory studies (clinical mean 0.81 ± 0.83; laboratory mean 0.48 ± 0.63; p = 0.82). Eleven papers addressed the degree of drift as a function of the duration of use; ten found no correlation while one [17] described a weak positive correlation (Spearman’s correlation coefficient 0.342; p = 0.001).

Discussion

This review found that when two ICP measurements were taken simultaneously using different sensors the averages were close, but with a large standard deviation. The 10 studies in Group 1 compared various sensors to ventricular catheters. Ventricular catheters are connected through a fluid-filled system (VFC) to an external transducer, which can be zeroed at will, offering a reliable reference standard. The calculation of the average ICP difference in Group 1 included both positive and negative values, which may have reduced the final picture (the absolute difference). While the majority of studies indicated only small differences between the intraparenchymal probe and the VFC, two reported widely differing ICP levels. Schickner et al. [4] studied 10 patients with refractory ICP. They used a Camino fiberoptic probe, and a ventricular catheter was inserted to drain CSF. The Camino readings, on average, were 9 mmHg higher than the VFC, with single episodes of 40 mmHg difference. Eide et al. [18] retrospectively studied patients after subarachnoid or intraparenchymal bleeding. In five cases a Codman catheter was compared with simultaneous readings by VFC; differences were greater than 10 mmHg in three cases, and one single case had large differences in all measurements, with negative ICP. The authors suggested that significant human errors in zeroing, or sensor damage, may have occurred. Group 2 comprised five studies comparing parenchymal probes without a ventricular catheter as a standard reference. Often these papers were published in order to prove the existence of ICP gradients, and it is therefore not surprising that they found a striking mean difference [2]. Additionally, in this group all differences were positive, so there were no negative values to offset the final average. This may explain why the mean differences were slightly larger, but still only around 1–2 mmHg. To sum up all the studies, we used both a fixed model and a random model. Although all the studies involved a comparison of simultaneous ICP measurements, the populations were not homogeneous. They reflected a cross-section of clinical cases, with differences in the manufacturer and type of sensor used, the sensor placement, and the type and extent of brain injury. Since all these factors can potentially affect ICP measurements, it cannot be assumed that all the studies were investigating an unknown but constant difference between paired ICP readings. We therefore feel that the assumptions for the random effects model are more appropriate. Nonetheless, the results with both models were closely comparable. According to the random model, the combined mean difference between probes was 1.5 mmHg, with a 95 % CI of 0.7–2.3 mmHg. The simple pooled estimate of the standard deviation (unweighted) across the 15 data sets was 4.4 mmHg. Based on this estimate, a future study (similar to the average composition of the 15 studies examined) comparing two simultaneous ICP readings would be expected to have 95 % of the observed differences in the interval 1.5 ± (2 × 4.4) mmHg. Using the random standard error from the meta-analysis, which includes between-study variability, this interval would be slightly higher at 1.5 ± (2 × 5.7) mmHg. According to the random effects model, 70 % of readings could therefore vary in the range of ±6 mmHg, and 95 % of readings in the range of ±11.4 mmHg. Discrepancies can be “true”, when there are pressure differences in the various areas of the brain, or due to inaccurate readings. True differences, when intracranial gradients are caused by expanding masses or unilateral hemorrhages, call for clinical judgment. In clinical routine, ICP is monitored with a single catheter in the overwhelming majority of patients. In case of unilateral masses with large shift, pressure gradients ( with higher pressure on the lesion side) are possible. In this situation the ICP data have to be interpreted considering where the catheter is placed. Inaccurate readings due to technical problems are more worrying since they are not due to actual intracranial gradients (and cannot be suspected in the computed tomography scan) but to erroneous measurement. Judging from the results of the meta-analysis, true discrepancies are relatively rare, but may be significant, up to ±11.4 mmHg. Clinicians should always carefully analyze the monitored data and also look at trends, possibly in the context of multimodal monitoring. Fluid-coupled catheters can be re-zeroed as often as necessary while other sensors, in which zeroing cannot be repeated after insertion, may become less precise over time. Zero drift has been reported for several catheters under clinical and laboratory conditions [8–11, 19]. We found the mean drift was limited (on average less than 1 mmHg); however, single catheters varied widely. In general the pressures were within the limits specified by the manufacturers. There seem to be no significant differences between the various probes in clinical and experimental studies. All papers but one [20] showed no clear correlation between drift and length of ICP monitoring. We can therefore assume that drift is not related to the duration of monitoring, as previously indicated [7]. Our review, pooling different studies often with different designs, has limitations. The bulk of evidence on this topic is also fairly old: the median year of publication of studies on accuracy was 1998, and the range 1992–2012; for studies on drift the median year of publication was 2001 (range 1987–2009). The standards for publication have improved with time, with the result that recent studies provided more comprehensive information compared to the old ones.

Conclusions

The clinical implications of our findings may be relevant both for future comparisons and for daily care. ICP monitoring is widely used to guide diagnostic and therapeutic decisions, so it is essential to know how reliable are the current techniques. The latest update of the Brain Trauma Foundation guidelines [7] concluded that parenchymal transducer devices measure ICP similarly to VFC but have the potential for differences. Our findings confirm that the average error is clinically negligible, but the interval of error is wide.

Key messages

A systematic literature review and meta-analysis indicates that the average error between simultaneous ICP measures is small (in the order of 1.5–1.6 mmHg), but in 30 % of readings it could exceed ±6 mmHg When there are intracranial gradients due to expanding masses or unilateral hemorrhages, there are likely to be larger differences between separate ICP measures. In the absence of intracranial gradients, the risk of inaccurate measurement seems rare. These findings are important both for future comparisons studies and for daily care.
  32 in total

1.  Intracranial pressure and clinical status: assessment of two intracranial pressure transducers.

Authors:  K Banister; I R Chambers; M S Siddique; H M Femandes; A D Mendelow
Journal:  Physiol Meas       Date:  2000-11       Impact factor: 2.833

2.  Clinical comparison of the Spiegelberg parenchymal transducer and ventricular fluid pressure.

Authors:  I R Chambers; M S Siddique; K Banister; A D Mendelow
Journal:  J Neurol Neurosurg Psychiatry       Date:  2001-09       Impact factor: 10.154

3.  Fiberoptic intraparenchymal brain pressure monitoring with the Camino V420 monitor: reflections on our experience in 163 severely head-injured patients.

Authors:  Maria-Antonia Poca; Juan Sahuquillo; Mercedes Arribas; Marcelino Báguena; Sonia Amorós; Enrique Rubio
Journal:  J Neurotrauma       Date:  2002-04       Impact factor: 5.269

4.  Clinical evaluation of intraparenchymal Spiegelberg pressure sensor.

Authors:  Josef-Michael Lang; Jürgen Beck; Michael Zimmermann; Volker Seifert; Andreas Raabe
Journal:  Neurosurgery       Date:  2003-06       Impact factor: 4.654

5.  Bench test assessment of the new Raumedic Neurovent-P ICP sensor: a technical report by the BrainIT group.

Authors:  G Citerio; I Piper; M Cormio; D Galli; S Cazzaniga; P Enblad; P Nilsson; C Contant; I Chambers
Journal:  Acta Neurochir (Wien)       Date:  2004-08-30       Impact factor: 2.216

6.  Intracranial pressure monitoring: fiberoptic monitor compared with the ventricular catheter.

Authors:  D J Schickner; R F Young
Journal:  Surg Neurol       Date:  1992-04

7.  A clinical evaluation of the Camino subdural screw and ventricular monitoring kits.

Authors:  I R Chambers; A D Mendelow; E J Sinar; P Modha
Journal:  Neurosurgery       Date:  1990-03       Impact factor: 4.654

8.  The Camino intracranial pressure sensor: is it optimal technology? An internal audit with a review of current intracranial pressure monitoring technologies.

Authors:  I Piper; A Barnes; D Smith; L Dunn
Journal:  Neurosurgery       Date:  2001-11       Impact factor: 4.654

9.  A comparison of intraparenchymatous and intraventricular pressure recording in clinical practice.

Authors:  G Sundbärg; C H Nordström; K Messeter; S Söderström
Journal:  J Neurosurg       Date:  1987-12       Impact factor: 5.115

10.  Clinical evaluation of a new intracranial pressure monitoring device.

Authors:  R Stendel; J Heidenreich; A Schilling; R Akhavan-Sigari; R Kurth; T Picht; T Pietilä; O Suess; C Kern; J Meisel; M Brock
Journal:  Acta Neurochir (Wien)       Date:  2003-03       Impact factor: 2.216

View more
  17 in total

1.  Differentiate the Source and Site of Intracranial Pressure Measurements Using More Precise Nomenclature.

Authors:  DaiWai M Olson; Stefany Ortega Peréz; Jonathan Ramsay; Chethan P Venkatasubba Rao; Jose I Suarez; Molly McNett; Venkatesh Aiyagari
Journal:  Neurocrit Care       Date:  2019-04       Impact factor: 3.210

2.  Intracranial pressure thresholds in severe traumatic brain injury: Con : The injured brain is not aware of ICP thresholds!

Authors:  Raimund Helbok; G Meyfroidt; R Beer
Journal:  Intensive Care Med       Date:  2018-07-05       Impact factor: 17.440

Review 3.  Evidence-Based Management of External Ventricular Drains.

Authors:  David Y Chung; DaiWai M Olson; Sayona John; Wazim Mohamed; Monisha A Kumar; Bradford B Thompson; Guy A Rordorf
Journal:  Curr Neurol Neurosci Rep       Date:  2019-11-26       Impact factor: 5.081

Review 4.  Brainstem Monitoring in the Neurocritical Care Unit: A Rationale for Real-Time, Automated Neurophysiological Monitoring.

Authors:  James L Stone; Julian E Bailes; Ahmed N Hassan; Brian Sindelar; Vimal Patel; John Fino
Journal:  Neurocrit Care       Date:  2017-02       Impact factor: 3.210

Review 5.  Imaging Guidance for Therapeutic Delivery: The Dawn of Neuroenergetics.

Authors:  Vilakshan Alambyan; Jonathan Pace; Persen Sukpornchairak; Xin Yu; Hamza Alnimir; Ryan Tatton; Gautham Chitturu; Anisha Yarlagadda; Ciro Ramos-Estebanez
Journal:  Neurotherapeutics       Date:  2020-04       Impact factor: 7.620

6.  Fluidic Considerations of Measuring Intracranial Pressure Using an Open External Ventricular Drain.

Authors:  Peter G Beidler; Alexander Novokhodko; Laura M Prolo; Samuel Browd; Barry R Lutz
Journal:  Cureus       Date:  2021-05-29

Review 7.  Impact of Intracranial Pressure Monitoring on Prognosis of Patients With Severe Traumatic Brain Injury: A PRISMA Systematic Review and Meta-Analysis.

Authors:  Jinsong Han; Shumao Yang; Chunyu Zhang; Ming Zhao; Anmin Li
Journal:  Medicine (Baltimore)       Date:  2016-02       Impact factor: 1.889

8.  Transcranial Doppler Monitoring of Intracranial Pressure Plateau Waves.

Authors:  Danilo Cardim; Bernhard Schmidt; Chiara Robba; Joseph Donnelly; Corina Puppo; Marek Czosnyka; Peter Smielewski
Journal:  Neurocrit Care       Date:  2017-06       Impact factor: 3.210

Review 9.  Review: pathophysiology of intracranial hypertension and noninvasive intracranial pressure monitoring.

Authors:  Nicolas Canac; Kian Jalaleddini; Samuel G Thorpe; Corey M Thibeault; Robert B Hamilton
Journal:  Fluids Barriers CNS       Date:  2020-06-23

Review 10.  Measuring intracranial pressure by invasive, less invasive or non-invasive means: limitations and avenues for improvement.

Authors:  Karen Brastad Evensen; Per Kristian Eide
Journal:  Fluids Barriers CNS       Date:  2020-05-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.