Literature DB >> 7643993

Clinical evaluation of a miniature strain-gauge transducer for monitoring intracranial pressure.

S P Gopinath1, C S Robertson, C F Contant, R K Narayan, R G Grossman.   

Abstract

In 25 patients, we evaluated the accuracy of a new miniature strain-gauge transducer developed for the measurement of intracranial pressure (ICP). The ICP in each patient was measured with the intraventricular, miniature strain-gauge transducer, and that value was compared with the ICP measured with a ventriculostomy catheter coupled to an external strain-gauge transducer. From the two monitors, 2218 simultaneous measurements of ICP were obtained. The average ICP measured with the miniature strain-gauge transducer was 15.9 +/- 10.0 mm Hg (range, -3 to 104 mm Hg). The ICP measured with the ventriculostomy-catheter transducer averaged 15.4 +/- 10.1 mm Hg (range, -9 to 104 mm Hg). A highly significant correlation was found over the wide range of pressures observed (n = 2218, r = 0.97, P < 0.001). The average difference between the two measurements of the ICP was 0.5 +/- 2.6 mm Hg, and the differences were equally positive and negative, demonstrating no consistent bias. The two values for the ICP were within 2 mm Hg of each other on 63% of the measurements and within 4 mm Hg of each other on 89% of the measurements. The average zero drift of the miniature strain-gauge transducer, measured at ambient pressure after removal of the catheter, was 0.2 +/- 0.5 mm Hg. The results indicate that this miniature strain-gauge transducer is highly accurate and stable and that it is a reliable alternative to a ventriculostomy for monitoring the ICP.

Entities:  

Mesh:

Year:  1995        PMID: 7643993     DOI: 10.1227/00006123-199506000-00011

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  11 in total

1.  Complications of brain tissue pressure monitoring with a fiberoptic device.

Authors:  A Bekar; S Gören; E Korfali; K Aksoy; S Boyaci
Journal:  Neurosurg Rev       Date:  1998       Impact factor: 3.042

2.  Evaluation of minimally invasive percutaneous CT-controlled ventriculostomy in patients with severe head trauma.

Authors:  M Krötz; U Linsenmaier; K G Kanz; K J Pfeifer; W Mutschler; M Reiser
Journal:  Eur Radiol       Date:  2003-11-06       Impact factor: 5.315

Review 3.  Management of intracranial hypertension.

Authors:  Sunit C Singhi; Lokesh Tiwari
Journal:  Indian J Pediatr       Date:  2009-05-24       Impact factor: 1.967

Review 4.  Management of intracranial hypertension.

Authors:  Leonardo Rangel-Castilla; Leonardo Rangel-Castillo; Shankar Gopinath; Claudia S Robertson
Journal:  Neurol Clin       Date:  2008-05       Impact factor: 3.806

5.  In vivo accuracy of two intraparenchymal intracranial pressure monitors.

Authors:  Thomas Lescot; Vincent Reina; Yannick Le Manach; Filippo Boroli; Dorian Chauvet; Anne-Laure Boch; Louis Puybasset
Journal:  Intensive Care Med       Date:  2011-02-26       Impact factor: 17.440

6.  Intracranial Pressure as an Objective Biomarker of Decompression Adequacy in Large Territory Infarction: A Multicenter Observational Study.

Authors:  Jia Xu Lim; Sherry Jiani Liu; Tien Meng Cheong; Seyed Ehsan Saffari; Julian Xinguang Han; Min Wei Chen
Journal:  Front Surg       Date:  2022-05-06

7.  Feasibility of non-invasive pressure support ventilation in infants with respiratory failure after extubation: a pilot study.

Authors:  Pascal Stucki; Marie-Hélène Perez; Pietro Scalfaro; Quentin de Halleux; François Vermeulen; Jacques Cotting
Journal:  Intensive Care Med       Date:  2009-06-17       Impact factor: 17.440

Review 8.  Advances in Intracranial Pressure Monitoring and Its Significance in Managing Traumatic Brain Injury.

Authors:  Usmah Kawoos; Richard M McCarron; Charles R Auker; Mikulas Chavko
Journal:  Int J Mol Sci       Date:  2015-12-04       Impact factor: 5.923

Review 9.  Management of raised intracranial pressure in children with traumatic brain injury.

Authors:  Vinay Kukreti; Hadi Mohseni-Bod; James Drake
Journal:  J Pediatr Neurosci       Date:  2014 Sep-Dec

Review 10.  Accuracy of intracranial pressure monitoring: systematic review and meta-analysis.

Authors:  Lucia Zacchetti; Sandra Magnoni; Federica Di Corte; Elisa R Zanier; Nino Stocchetti
Journal:  Crit Care       Date:  2015-12-02       Impact factor: 9.097

View more