| Literature DB >> 21318463 |
Yolanda Ballestero1, Jesús López-Herce, Javier Urbano, Maria José Solana, Marta Botrán, Jose M Bellón, Angel Carrillo.
Abstract
The objective of this study was to evaluate a new bioreactance method for noninvasive cardiac output (CO) measurement (NICOM) in children. Ten patients between 1 and 144 months of age and with no hemodynamic disturbances were studied. Using bioreactance, heart rate (HR), mean arterial pressure (MAP), and cardiac index (CI) measurements were made every 6-8 h. CI was 2.4 ± 1.03 l/min/1.73 m(2) (range 1-4.9 l/min/1.73 m(2)); There were significant correlations between CI and age (r = 0.50, P = 0.003), weight (r = 0.66, P < 0.001), and MAP (r = 0.369, P = 0.037). Significant differences in CI (P < 0.001) were detected between children weighing <10 kg (1.9 ± 0.73 l/min/1.73 m(2); range 1-3.2), 10-20 kg (2.07 ± 0.7 l/min/1.73 m(2); range 1-3.6), and >20 kg (3.7 ± 0.8 l/min/1.73 m(2); range 2.4-4.9). We conclude that the CI measured by bioreactance in children varies with the age and weight of the patients and is lower than the normal range in a large percentage of measurements. These data suggest that this method is not useful for evaluating CI in small children.Entities:
Mesh:
Year: 2011 PMID: 21318463 DOI: 10.1007/s00246-011-9903-1
Source DB: PubMed Journal: Pediatr Cardiol ISSN: 0172-0643 Impact factor: 1.655