A maximum NEC criterion for Compton collimation to accurately identify true coincidences in PET.

Garry Chinn1, Craig S Levin.   

Abstract

In this work, we propose a new method to increase the accuracy of identifying true coincidence events for positron emission tomography (PET). This approach requires 3-D detectors with the ability to position each photon interaction in multi-interaction photon events. When multiple interactions occur in the detector, the incident direction of the photon can be estimated using the Compton scatter kinematics (Compton Collimation). If the difference between the estimated incident direction of the photon relative to a second, coincident photon lies within a certain angular range around colinearity, the line of response between the two photons is identified as a true coincidence and used for image reconstruction. We present an algorithm for choosing the incident photon direction window threshold that maximizes the noise equivalent counts of the PET system. For simulated data, the direction window removed 56%-67% of random coincidences while retaining > 94% of true coincidences from image reconstruction as well as accurately extracted 70% of true coincidences from multiple coincidences.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21317079      PMCID: PMC3667991          DOI: 10.1109/TMI.2011.2113379

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  15 in total

1.  Application of annihilation coincidence detection to transaxial reconstruction tomography.

Authors:  M E Phelps; E J Hoffman; N A Mullani; M M Ter-Pogossian
Journal:  J Nucl Med       Date:  1975-03       Impact factor: 10.057

2.  Study of the performance of a novel 1 mm resolution dual-panel PET camera design dedicated to breast cancer imaging using Monte Carlo simulation.

Authors:  Jin Zhang; Peter D Olcott; Garry Chinn; Angela M K Foudray; Craig S Levine
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

3.  Model-based scatter correction for fully 3D PET.

Authors:  J M Ollinger
Journal:  Phys Med Biol       Date:  1996-01       Impact factor: 3.609

4.  Quantitation in positron emission computed tomography: 7. A technique to reduce noise in accidental coincidence measurements and coincidence efficiency calibration.

Authors:  M E Casey; E J Hoffman
Journal:  J Comput Assist Tomogr       Date:  1986 Sep-Oct       Impact factor: 1.826

5.  An electronically collimated gamma camera for single photon emission computed tomography. Part II: Image reconstruction and preliminary experimental measurements.

Authors:  M Singh; D Doria
Journal:  Med Phys       Date:  1983 Jul-Aug       Impact factor: 4.071

6.  An electronically collimated gamma camera for single photon emission computed tomography. Part I: Theoretical considerations and design criteria.

Authors:  M Singh
Journal:  Med Phys       Date:  1983 Jul-Aug       Impact factor: 4.071

7.  Design study of a high-resolution breast-dedicated PET system built from cadmium zinc telluride detectors.

Authors:  Hao Peng; Craig S Levin
Journal:  Phys Med Biol       Date:  2010-04-19       Impact factor: 3.609

8.  Study of a high-resolution, 3D positioning cadmium zinc telluride detector for PET.

Authors:  Y Gu; J L Matteson; R T Skelton; A C Deal; E A Stephan; F Duttweiler; T M Gasaway; C S Levin
Journal:  Phys Med Biol       Date:  2011-02-18       Impact factor: 3.609

9.  Bayesian reconstruction of photon interaction sequences for high-resolution PET detectors.

Authors:  Guillem Pratx; Craig S Levin
Journal:  Phys Med Biol       Date:  2009-08-04       Impact factor: 3.609

10.  Impact of high energy resolution detectors on the performance of a PET system dedicated to breast cancer imaging.

Authors:  Craig S Levin; Angela M K Foudray; Frezghi Habte
Journal:  Phys Med       Date:  2006       Impact factor: 2.685

View more
  4 in total

Review 1.  Promising new photon detection concepts for high-resolution clinical and preclinical PET.

Authors:  Craig S Levin
Journal:  J Nucl Med       Date:  2012-02       Impact factor: 10.057

2.  Effect of CZT system characteristics on Compton scatter event recovery.

Authors:  Sheng Yang; Mohan Li; Michael Reed; James Hugg; Henry Chen; Shiva Abbaszadeh
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2019-05-06

3.  Study of electrode pattern design for a CZT-based PET detector.

Authors:  Y Gu; C S Levin
Journal:  Phys Med Biol       Date:  2014-05-01       Impact factor: 3.609

4.  Positioning true coincidences that undergo inter-and intra-crystal scatter for a sub-mm resolution cadmium zinc telluride-based PET system.

Authors:  Shiva Abbaszadeh; Garry Chinn; Craig S Levin
Journal:  Phys Med Biol       Date:  2018-01-09       Impact factor: 3.609

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.