Literature DB >> 21287370

A comparison of the image quality of full-time myocardial perfusion SPECT vs wide beam reconstruction half-time and half-dose SPECT.

E Gordon DePuey1, Srinivas Bommireddipalli, John Clark, Anna Leykekhman, Linda B Thompson, Marvin Friedman.   

Abstract

OBJECTIVES: Wide Beam Reconstruction (WBR) (UltraSPECT, Ltd) uses resolution recovery and noise modeling to cope with decreased SPECT count statistics. Because WBR processing reconstructs half the usual SPECT count statistics, we postulate that image quality equivalent to a full-time acquisition can be achieved in either half the time or with half the radiopharmaceutical activity.
METHODS: In 156 consecutive patients (pts) rest and 8-frame gated post-stress myocardial perfusion SPECT was performed following 333-444 and 1184-1480 MBq (9-12 and 32-40 mCi) Tc-99m sestamibi injections, respectively, with full-time (rest = 14 min; stress = 12.3 min) acquisitions processed with OSEM and also separate "half-time" acquisitions processed with WBR. A subsequent group of 160 consecutive pts matched in gender, weight, and chest circumference received "half-dose" rest and stress injections 214.6 ± 22.2 and 647.5 ± 92.5 MBq (5.8 ± 0.6 and 17.5 ± 2.5 mCi) with full-time SPECT acquisitions. Image quality (1 = poor to 5 = excellent) was judged by myocardial count density and uniformity, endocardial edge definition, perfusion defect delineation, right ventricular visualization, and background noise.
RESULTS: Mean image quality for rest, stress, and post-stress gated images were 3.6 ± 0.7, 3.8 ± 0.7, and 3.9 ± 1.0, respectively, for "full-time OSEM; 3.7 ± 0.8, 4.0 ± 0.7, and 4.8 ± 0.4 for "half-time" WBR; and 4.3 ± 0.8, 4.6 ± 0.6, and 4.7 ± 0.6 for "half-dose" WBR. "Half-time" and "half-dose" WBR image quality were both superior to standard full-time OSEM (P's < .001). There was no significant difference between the summed stress and rest scores for "full-time" OSEM vs "half-time" WBR in 82 patients with perfusion defects.
CONCLUSIONS: Both "half-time" and "half-dose" WBR provide myocardial perfusion SPECT quality superior to full-time OSEM, with an associated decrease in scan acquisition time and patient radiation exposure, respectively.

Entities:  

Mesh:

Year:  2011        PMID: 21287370     DOI: 10.1007/s12350-011-9340-9

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  10 in total

1.  Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies.

Authors:  Salvador Borges-Neto; Robert A Pagnanelli; Linda K Shaw; Emily Honeycutt; Shuli C Shwartz; George L Adams; Ralph Edward Coleman
Journal:  J Nucl Cardiol       Date:  2007-07       Impact factor: 5.952

2.  Prospective multicenter evaluation of rapid, gated SPECT myocardial perfusion upright imaging.

Authors:  Jamshid Maddahi; Ricardo Mendez; John J Mahmarian; Gregory Thomas; Hetal Babla; Chuanyong Bai; Samia Arram; Peter Maffetone; Richard Conwell
Journal:  J Nucl Cardiol       Date:  2009-02-27       Impact factor: 5.952

3.  New software methods to cope with reduced counting statistics: shorter SPECT acquisitions and many more possibilities.

Authors:  E Gordon Depuey
Journal:  J Nucl Cardiol       Date:  2009-04-28       Impact factor: 5.952

Review 4.  The year in cardiac imaging.

Authors:  Raymond J Gibbons; Philip A Araoz; Eric E Williamson
Journal:  J Am Coll Cardiol       Date:  2009-01-06       Impact factor: 24.094

5.  Stress-only myocardial perfusion imaging a new paradigm.

Authors:  Ami E Iskandrian
Journal:  J Am Coll Cardiol       Date:  2009-11-13       Impact factor: 24.094

Review 6.  Radiation dose to patients from cardiac diagnostic imaging.

Authors:  Andrew J Einstein; Kevin W Moser; Randall C Thompson; Manuel D Cerqueira; Milena J Henzlova
Journal:  Circulation       Date:  2007-09-11       Impact factor: 29.690

7.  Ordered subset expectation maximization and wide beam reconstruction "half-time" gated myocardial perfusion SPECT functional imaging: a comparison to "full-time" filtered backprojection.

Authors:  E Gordon DePuey; Ramesh Gadiraju; John Clark; Linda Thompson; Frank Anstett; Shuli C Shwartz
Journal:  J Nucl Cardiol       Date:  2008-04-24       Impact factor: 5.952

8.  High-speed myocardial perfusion imaging initial clinical comparison with conventional dual detector anger camera imaging.

Authors:  Tali Sharir; Simona Ben-Haim; Konstantine Merzon; Vitali Prochorov; Dalia Dickman; Shlomo Ben-Haim; Daniel S Berman
Journal:  JACC Cardiovasc Imaging       Date:  2008-03

9.  Wide beam reconstruction "quarter-time" gated myocardial perfusion SPECT functional imaging: a comparison to "full-time" ordered subset expectation maximum.

Authors:  E Gordon DePuey; Srinivas Bommireddipalli; John Clark; Linda Thompson; Yossi Srour
Journal:  J Nucl Cardiol       Date:  2009-06-17       Impact factor: 5.952

10.  A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction.

Authors:  Carmelo V Venero; Gary V Heller; Timothy M Bateman; A Iain McGhie; Alan W Ahlberg; Deborah Katten; Staci A Courter; Robert J Golub; James A Case; S James Cullom
Journal:  J Nucl Cardiol       Date:  2009-07-07       Impact factor: 5.952

  10 in total
  33 in total

1.  Reduced stress dose with rapid acquisition CZT SPECT MPI in a non-obese clinical population: comparison to coronary angiography.

Authors:  W Lane Duvall; Joseph M Sweeny; Lori B Croft; Eric Ginsberg; Krista A Guma; Milena J Henzlova
Journal:  J Nucl Cardiol       Date:  2011-12-07       Impact factor: 5.952

2.  Patient-centered imaging.

Authors:  E Gordon Depuey; John J Mahmarian; Todd D Miller; Andrew J Einstein; Christopher L Hansen; Thomas A Holly; Edward J Miller; Donna M Polk; L Samuel Wann
Journal:  J Nucl Cardiol       Date:  2012-04       Impact factor: 5.952

3.  Advances in SPECT camera software and hardware: currently available and new on the horizon.

Authors:  E Gordon DePuey
Journal:  J Nucl Cardiol       Date:  2012-06       Impact factor: 5.952

4.  A qualitative and quantitative assessment of the impact of three processing algorithms with halving of study count statistics in myocardial perfusion imaging: filtered backprojection, maximal likelihood expectation maximisation and ordered subset expectation maximisation with resolution recovery.

Authors:  B N Modi; J L E Brown; G Kumar; R M Driver; A D Kelion; A M Peters; J C Fowler
Journal:  J Nucl Cardiol       Date:  2012-06-30       Impact factor: 5.952

5.  Very low-activity stress/high-activity rest, single-day myocardial perfusion SPECT with a conventional sodium iodide camera and wide beam reconstruction processing.

Authors:  E Gordon DePuey; Pashmina Ata; Rick Wray; Marvin Friedman
Journal:  J Nucl Cardiol       Date:  2012-07-10       Impact factor: 5.952

Review 6.  Recent advances in cardiac SPECT instrumentation and system design.

Authors:  Mark F Smith
Journal:  Curr Cardiol Rep       Date:  2013-08       Impact factor: 2.931

7.  High diagnostic accuracy of low-dose gated-SPECT with solid-state ultrafast detectors: preliminary clinical results.

Authors:  Alessia Gimelli; Matteo Bottai; Dario Genovesi; Assuero Giorgetti; Fabio Di Martino; Paolo Marzullo
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-09-02       Impact factor: 9.236

8.  Opportunities for improvement on current nuclear cardiology practices and radiation exposure in Latin America: Findings from the 65-country IAEA Nuclear Cardiology Protocols cross-sectional Study (INCAPS).

Authors:  João V Vitola; Fernando Mut; Erick Alexánderson; Thomas N B Pascual; Mathew Mercuri; Ganesan Karthikeyan; Nathan Better; Madan M Rehani; Ravi Kashyap; Maurizio Dondi; Diana Paez; Andrew J Einstein
Journal:  J Nucl Cardiol       Date:  2016-02-22       Impact factor: 5.952

9.  Myocardial perfusion imaging with a solid-state camera: simulation of a very low dose imaging protocol.

Authors:  Ryo Nakazato; Daniel S Berman; Sean W Hayes; Mathews Fish; Richard Padgett; Yuan Xu; Mark Lemley; Rafael Baavour; Nathaniel Roth; Piotr J Slomka
Journal:  J Nucl Med       Date:  2013-01-15       Impact factor: 10.057

10.  Optimizing radiation dose and imaging time with conventional myocardial perfusion SPECT: Technical aspects.

Authors:  Piotr Slomka; Guido Germano
Journal:  J Nucl Cardiol       Date:  2016-03-08       Impact factor: 5.952

View more