Literature DB >> 19533264

Wide beam reconstruction "quarter-time" gated myocardial perfusion SPECT functional imaging: a comparison to "full-time" ordered subset expectation maximum.

E Gordon DePuey1, Srinivas Bommireddipalli, John Clark, Linda Thompson, Yossi Srour.   

Abstract

BACKGROUND: Previously we reported that compared to iterative reconstruction with ordered subset expectation maximum (OSEM), wide beam reconstruction (WBR), which incorporates resolution recovery and controls noise during reconstruction without applying a post-processing filter, allows half-time SPECT acquisition with preserved diagnostic quality. We now postulate that with further Poisson noise treatment, quarter-time acquisition is possible.
METHODS: The half-time WBR algorithm was optimized for quarter-time acquisition based upon anthropomorphic cardiac phantom data and a pilot group of 48 patients (pts). Then using the modified algorithm, 209 pts (91 men, 118 women, mean chest circumference = 40 in) were imaged at rest (R) and stress (S) (9/32 mCi (99m)Tc-sestamibi) full-time with OSEM, and again quarter-time with the modified WBR algorithm. The 180 degrees , 64-stop, full-time single-day rest (R) (25 second-per-stop, sps) and 8-frame per cardiac cycle post-stress (S) (20 sps) gated SPECT, and then quarter-time R (6 sps) and post-S (4 sps) gated SPECT were acquired. Blinded observers graded scan quality (1 = poor to 5 = excellent) based on myocardial uniformity, endocardial/epicardial edge definition, and background noise. Perfusion defects were scored using a 17-segment model. Using three commercially available software methods, end-diastolic volume (EDV), end-systolic volume (ESV), and left ventricular ejection fraction (LVEF) were calculated.
RESULTS: For the 209 prospective pts, mean image quality for R full-time OSEM and quarter-time WBR were similar (3.5 +/- 0.9 vs 3.6 +/- 0.7, p NS). For S, quarter-time WBR quality was superior to full-time OSEM (4.3 +/- 0.7 vs 3.9 +/- 0.7) (P = 1.78 x 10(-17)). In 35 pts with chest circumferences >44 inches a longer, 10 sps WBR acquisition improved resting image quality. Of 48 pts with abnormal scans (SSSs > 2 by OSEM) mean summed stress scores, summed rest scores, and summed difference scores were not significantly different with quarter-time WBR vs full-time OSEM (11.2 vs 10.9), (9.1 vs 9.0), (2.0 vs 1.9) (P NS). For the three software methods, there was a good correlation of LVEF, EDV, and ESV determined by WBR vs OSEM (all r > 0.92). ESVs were generally higher with WBR, primarily due to better delineation of the valve plane at end-systole, whereas EDVs were similar. Thus, EFs were significantly lower with WBR.
CONCLUSIONS: For perfusion SPECT quarter-time WBR affords image quality, defect characterization, and functional assessment equivalent to full-time OSEM.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19533264     DOI: 10.1007/s12350-009-9108-7

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  6 in total

1.  Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies.

Authors:  Salvador Borges-Neto; Robert A Pagnanelli; Linda K Shaw; Emily Honeycutt; Shuli C Shwartz; George L Adams; Ralph Edward Coleman
Journal:  J Nucl Cardiol       Date:  2007-07       Impact factor: 5.952

2.  Recent technologic advances in nuclear cardiology.

Authors:  James A Patton; Piotr J Slomka; Guido Germano; Daniel S Berman
Journal:  J Nucl Cardiol       Date:  2007-07       Impact factor: 5.952

3.  Prospective multicenter evaluation of rapid, gated SPECT myocardial perfusion upright imaging.

Authors:  Jamshid Maddahi; Ricardo Mendez; John J Mahmarian; Gregory Thomas; Hetal Babla; Chuanyong Bai; Samia Arram; Peter Maffetone; Richard Conwell
Journal:  J Nucl Cardiol       Date:  2009-02-27       Impact factor: 5.952

4.  Ordered subset expectation maximization and wide beam reconstruction "half-time" gated myocardial perfusion SPECT functional imaging: a comparison to "full-time" filtered backprojection.

Authors:  E Gordon DePuey; Ramesh Gadiraju; John Clark; Linda Thompson; Frank Anstett; Shuli C Shwartz
Journal:  J Nucl Cardiol       Date:  2008-04-24       Impact factor: 5.952

5.  High-speed myocardial perfusion imaging initial clinical comparison with conventional dual detector anger camera imaging.

Authors:  Tali Sharir; Simona Ben-Haim; Konstantine Merzon; Vitali Prochorov; Dalia Dickman; Shlomo Ben-Haim; Daniel S Berman
Journal:  JACC Cardiovasc Imaging       Date:  2008-03

6.  Multicenter investigation comparing a highly efficient half-time stress-only attenuation correction approach against standard rest-stress Tc-99m SPECT imaging.

Authors:  Timothy M Bateman; Gary V Heller; A Iain McGhie; Staci A Courter; Robert A Golub; James A Case; S James Cullom
Journal:  J Nucl Cardiol       Date:  2009-06-23       Impact factor: 5.952

  6 in total
  37 in total

1.  Patient-centered imaging.

Authors:  E Gordon Depuey; John J Mahmarian; Todd D Miller; Andrew J Einstein; Christopher L Hansen; Thomas A Holly; Edward J Miller; Donna M Polk; L Samuel Wann
Journal:  J Nucl Cardiol       Date:  2012-04       Impact factor: 5.952

2.  Advances in SPECT camera software and hardware: currently available and new on the horizon.

Authors:  E Gordon DePuey
Journal:  J Nucl Cardiol       Date:  2012-06       Impact factor: 5.952

3.  A qualitative and quantitative assessment of the impact of three processing algorithms with halving of study count statistics in myocardial perfusion imaging: filtered backprojection, maximal likelihood expectation maximisation and ordered subset expectation maximisation with resolution recovery.

Authors:  B N Modi; J L E Brown; G Kumar; R M Driver; A D Kelion; A M Peters; J C Fowler
Journal:  J Nucl Cardiol       Date:  2012-06-30       Impact factor: 5.952

4.  Recommendations for reducing radiation exposure in myocardial perfusion imaging.

Authors:  Manuel D Cerqueira; Kevin C Allman; Edward P Ficaro; Christopher L Hansen; Kenneth J Nichols; Randall C Thompson; William A Van Decker; Marko Yakovlevitch
Journal:  J Nucl Cardiol       Date:  2010-08       Impact factor: 5.952

5.  Single photon-emission computed tomography.

Authors:  Thomas A Holly; Brian G Abbott; Mouaz Al-Mallah; Dennis A Calnon; Mylan C Cohen; Frank P DiFilippo; Edward P Ficaro; Michael R Freeman; Robert C Hendel; Diwakar Jain; Scott M Leonard; Kenneth J Nichols; Donna M Polk; Prem Soman
Journal:  J Nucl Cardiol       Date:  2010-10       Impact factor: 5.952

6.  Very low-activity stress/high-activity rest, single-day myocardial perfusion SPECT with a conventional sodium iodide camera and wide beam reconstruction processing.

Authors:  E Gordon DePuey; Pashmina Ata; Rick Wray; Marvin Friedman
Journal:  J Nucl Cardiol       Date:  2012-07-10       Impact factor: 5.952

7.  Sources of variability of gated myocardial perfusion SPECT quantitative parameters.

Authors:  E Gordon DePuey
Journal:  J Nucl Cardiol       Date:  2015-09-04       Impact factor: 5.952

8.  Combined supine and prone imaging acquisition in cardiac SPECT: A turn for the better.

Authors:  Mylan C Cohen
Journal:  J Nucl Cardiol       Date:  2016-01-07       Impact factor: 5.952

9.  Wide-beam reconstruction half-time SPECT improves diagnostic certainty and preserves normalcy and accuracy: a quantitative perfusion analysis.

Authors:  Regina S Druz; Lawrence M Phillips; Michelle Chugkowski; Loukas Boutis; Bruce Rutkin; Stanley Katz
Journal:  J Nucl Cardiol       Date:  2010-12-23       Impact factor: 5.952

10.  Myocardial perfusion imaging with a solid-state camera: simulation of a very low dose imaging protocol.

Authors:  Ryo Nakazato; Daniel S Berman; Sean W Hayes; Mathews Fish; Richard Padgett; Yuan Xu; Mark Lemley; Rafael Baavour; Nathaniel Roth; Piotr J Slomka
Journal:  J Nucl Med       Date:  2013-01-15       Impact factor: 10.057

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.