Shamurailatpam Dayananda Sharma1, R Prasad, Bina Shetye, V Rangarajan, D Deshpande, S K Shrivastava, K A Dinshaw.
Abstract
Integrated PET/CT has emerged as an integral component of oncology management because of its unique potential of providing both functional and morphological images in a single imaging session. In this work, performance of the 'bismuth germinate (BGO) crystal'-based PET of a newly installed Discovery ST PET/CT was evaluated in 2D and 3D mode for whole-body scanning using National Electrical Manufacturers Association (NEMA) NU 2-2001 protocol and the recommended phantoms. During the entire measurements, the system operates with an energy window of 375-650 keV and 11.7 ns coincidence time window. The set of tests performed were spatial resolution, sensitivity, scatter fraction (SF) and counting rate performance. The average transaxial and axial spatial resolution measured as full width at half maximum (FWHM) of the point spread function at 1 cm (and 10 cm) off-axis was 0.632 (0.691) and 0.491 (0.653) cm in 2D and 0.646 (0.682) and 0.54 (0.601) cm in 3D respectively. The average sensitivity for the two radial positions (R = 0 cm and R = 10 cm) was 2.56 (2.63) cps/kBq in 2D and 11.85 (12.14) cps/kBq in 3D. The average scatter fraction was 19.79% in 2D and 46.19% in 3D. The peak noise equivalent counting rate (NECR) evaluated with single random subtraction was 89.41 kcps at 49 kBq/cc in 2D and 60 kcps at 12 kBq/cc in 3D acquisition mode. The NECR with delayed random subtraction was 61.47 kcps at 40.67 kBq/cc in 2D and 45.57 kcps at 16.45 kBq/cc in 3D. The performance of the PET scanner was satisfactory within the manufacturer-specified limits. The test result of PET shows excellent system sensitivity with relatively uniform resolution throughout the FOV, making this scanner highly suitable for whole-body studies.Entities:
Keywords: National Electrical Manufacturers Association NU2 - 2001; PET/CT; performance measurement
Year: 2007 PMID: 21224924 PMCID: PMC3014099 DOI: 10.4103/0971-6203.37479
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Arrangement of detector modules in D-ST PET rings
Figure 2Experimental setup for spatial resolution measurement. Three 18F point sources contained in capillary tubes are placed at X = 0 cm, Y = 1 cm; X = 0 cm, Y = 10 cm; and X = 10 cm, Y = 0 cm in the center of the axial FOV of the scanner
Figure 3Experimental setup for sensitivity measurement. A line source surrounded by aluminum tubes of known thickness is suspended at the center of the scanner FOV
Figure 4Experimental setup for scatter fraction (SF) and count rate (CR) performance test. Cylindrical polyethylene phantom (20 cm diameter and 70 cm length) with the line source inserted posteriorly was positioned at the center of scanner FOV
Figure 5Axial and transverse profiles for a point source at 1 cm off center measured in 3D following NEMA NU 2-2001 protocol
Transaxial and axial resolution of a point source in 2D and 3D kept at different radial positions
| 0.632 | 0.646 | 0.491 | 0.540 | 1.271 | 1.237 | 1.078 | 1.232 | |
| 0.691 | 0.682 | 0.653 | 0.601 | 1.338 | 1.289 | 1.343 | 1.399 | |
Transaxial - average of radial and tangential; FWHM - full width at half maximum; FWTM - full width at tenth maximum
Figure 6Sensitivity across axial FOV measured in 2D (Figure 6a) and 3D (Figure 6b) according to NEMA NU 2-2001 protocol
Sensitivity in 2D and 3D acquisition modes
| 2.565 | 11.854 | |
| 2.627 | 12.137 |
Figure 7Scatter fraction in 3D mode across axial FOV of scanner measured according to NEMA NU 2-2001 protocol
Figure 8Counting rate performance of PET scanner at various levels of activity within phantom. Data acquired in 3D mode following NEMA NU 2-2001 protocol. Trues - solid; Randoms - dotted; Scatter - dashed; NECR (1R) - dash dot; NECR (2R) - dash dot dot
Counting rate performance in 2D and 3D acquisition modes
| Peak true rate | 309.29 @ 122.66 | 342.51@ 49.61 |
| Peak random rate | 1303.58 @ 134.84 | 2443.66 @ 79.64 |
| Peak scatter rate | 76.32 @ 122.66 | 293.86 @ 49.61 |
| Peak NEC ( | 89.41 @ 49 | 60 @ 12 |
| Peak NEC ( | 61.47 @ 40.67 | 45.57 @ 16.45 |