An inter-laboratory comparison study of image quality of PET scanners using the NEMA NU 2-2001 procedure for assessment of image quality.

Helmar Bergmann1, Georg Dobrozemsky, Gregory Minear, Rudolf Nicoletti, Martin Samal.   

Abstract

An inter-laboratory comparison study was conducted to assess the image quality of PET scanners in Austria. The survey included both dedicated PET scanners (D-PET, n = 8) and coincidence cameras (GC-PET, n = 7). Measurement of image quality was based on the NEMA (National Electrical Manufacturers Association) NU 2-2001 protocol and the IEC (International Electrotechnical Commission) body phantom. The latter contains six fillable spheres ranging in diameter from 37 mm down to 10 mm and a 'lung' insert. The two largest lesions L1-2 simulate cold lesions, the four smaller ones (L3-6) are filled with 18F and activity concentration ratios relative to background of 8:1 and 4:1, respectively. Acquisition and reconstruction in the study employed the participating institutes' standard oncological processing protocol. Calculation of contrast of the spheres was performed with a fully automated procedure. Contrast quality indices (CQIs) reflecting global performance were obtained by summing individual contrast values. Other image quality parameters calculated according to the NEMA protocol were background variability and relative error for correction of attenuation and scatter. Contrast values obtained were 61 +/- 16 and 37 +/- 14 for L1 (per cent contrast +/- SD for D-PET and GC-PET, respectively), 57 +/- 16 and 29 +/- 16 for L2, 46 +/- 10 and 26 +/- 6.3 for L3, 37 +/- 10 and 15 +/- 4.3 for L4, 26 +/- 11.5 and 6.1 +/- 2.5 for L5, 14 +/- 7.1 and 2.6 +/- 2.6 for L6, with D-PET systems consistently being superior to GC-PET systems. CQIs permitted ranking of the scanners, also demonstrating a clear distinction between D-PET and GC-PET systems. Background variability was largest for GC-PET systems; the relative error of attenuation and scatter correction was significantly correlated with image quality for D-PET systems only. The study demonstrated considerable differences in image quality not only between GC-PET and D-PET systems but also between individual D-PET systems with possible consequences for clinical interpretation of images and measurement of quantitative indices such as the standardized uptake value. The study provided valuable feedback to the participants as well as baseline data for improving interchangeability of PET images and of quantitative indices between different laboratories.

Entities:  

Mesh:

Year:  2005        PMID: 15876661     DOI: 10.1088/0031-9155/50/10/001

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  10 in total

1.  Data acquisition and analysis: the strength of methodology in nuclear medicine and molecular imaging.

Authors:  Peter Knoll; Horst Köhn; Siroos Mirzaei
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-06       Impact factor: 9.236

Review 2.  Image-based biomarkers in clinical practice.

Authors:  John E Bayouth; Thomas L Casavant; Michael M Graham; Milan Sonka; Manickam Muruganandham; John M Buatti
Journal:  Semin Radiat Oncol       Date:  2011-04       Impact factor: 5.934

3.  Acceptance testing for nuclear medicine instrumentation.

Authors:  Ellinor Busemann Sokole; Anna Płachcínska; Alan Britten
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-03       Impact factor: 9.236

4.  Routine quality control recommendations for nuclear medicine instrumentation.

Authors:  Ellinor Busemann Sokole; Anna Płachcínska; Alan Britten; Maria Lyra Georgosopoulou; Wendy Tindale; Rigobert Klett
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-03       Impact factor: 9.236

5.  Multi-centre variability of ISCEV standard ERGs in two normal adults.

Authors:  R Hamilton; A Al Abdlseaed; J Healey; M M Neveu; L Brown; D Keating; V A McBain; D Sculfor; D A Thompson
Journal:  Doc Ophthalmol       Date:  2015-02-27       Impact factor: 2.379

6.  Automated model-based quantitative analysis of phantoms with spherical inserts in FDG PET scans.

Authors:  Ethan J Ulrich; John J Sunderland; Brian J Smith; Imran Mohiuddin; Jessica Parkhurst; Kristin A Plichta; John M Buatti; Reinhard R Beichel
Journal:  Med Phys       Date:  2017-11-23       Impact factor: 4.071

7.  Artificial Neural Network-Based System for PET Volume Segmentation.

Authors:  Mhd Saeed Sharif; Maysam Abbod; Abbes Amira; Habib Zaidi
Journal:  Int J Biomed Imaging       Date:  2010-09-26

8.  Quality assurance of positron emission tomography/computed tomography for radiation therapy.

Authors:  Lei Xing
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008       Impact factor: 7.038

9.  Whole-body PET acceptance test in 2D and 3D using NEMA NU 2-2001 protocol.

Authors:  Shamurailatpam Dayananda Sharma; R Prasad; Bina Shetye; V Rangarajan; D Deshpande; S K Shrivastava; K A Dinshaw
Journal:  J Med Phys       Date:  2007-10

10.  Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard.

Authors:  Ivo Rausch; Jacobo Cal-González; David Dapra; Hans Jürgen Gallowitsch; Peter Lind; Thomas Beyer; Gregory Minear
Journal:  EJNMMI Phys       Date:  2015-10-26
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.