Sinking and fit of abutment of locking taper implant system.

Seung-Jin Moon1, Hee-Jung Kim, Mee-Kyoung Son, Chae-Heon Chung.   

Abstract

STATEMENT OF PROBLEM: Unlike screw-retention type, fixture-abutment retention in Locking taper connection depends on frictional force so it has possibility of abutment to sink.
PURPOSE: In this study, Bicon® Implant System, one of the conical internal connection implant system, was used with applying loading force to the abutments connected to the fixture. Then the amount of sinking was measured.
MATERIAL AND METHODS: 10 Bicon® implant fixtures were used. First, the abutment was connected to the fixture with finger force. Then it was tapped with a mallet for 3 times and loads of 20 kg corresponding to masticatory force using loading application instrument were applied successively. The abutment state, slightly connected to the fixture without pressure was considered as a reference length, and every new abutment length was measured after each load's step was added. The amount of abutment sinking (mm) was gained by subtracting the length of abutment-fixture under each loading condition from reference length.
RESULTS: It was evident, that the amount of abutment sinking in Bicon® Implant System increased as loads were added. When loads of 20 kg were applied more than 5 - 7 times, sinking stopped at 0.45 ± 0.09 mm.
CONCLUSION: Even though locking taper connection type implant shows good adaption to occlusal force, it has potential for abutment sinking as loads are given. When locking taper connection type implant is used, satisfactory loads are recommended for precise abutment location.

Entities:  

Keywords:  Bicon® Implant System; abutment sinking; locking taper connection; masticatory force

Year:  2009        PMID: 21165262      PMCID: PMC2994685          DOI: 10.4047/jap.2009.1.2.97

Source DB:  PubMed          Journal:  J Adv Prosthodont        ISSN: 2005-7806            Impact factor:   1.904


  13 in total

Review 1.  Implants and components: entering the new millennium.

Authors:  P P Binon
Journal:  Int J Oral Maxillofac Implants       Date:  2000 Jan-Feb       Impact factor: 2.804

Review 2.  Clinical complications of osseointegrated implants.

Authors:  C J Goodacre; J Y Kan; K Rungcharassaeng
Journal:  J Prosthet Dent       Date:  1999-05       Impact factor: 3.426

3.  Measurement of stress in mastication. I.

Authors:  D J ANDERSON
Journal:  J Dent Res       Date:  1956-10       Impact factor: 6.116

4.  Axial displacement of abutments into implants and implant replicas, with the tapered cone-screw internal connection, as a function of tightening torque.

Authors:  Bruno Dailey; Laurence Jordan; Olivier Blind; Bruno Tavernier
Journal:  Int J Oral Maxillofac Implants       Date:  2009 Mar-Apr       Impact factor: 2.804

5.  Osseointegrated implants for single tooth replacement: a 1-year report from a multicenter prospective study.

Authors:  T Jemt; W R Laney; D Harris; P J Henry; P H Krogh; G Polizzi; G A Zarb; I Herrmann
Journal:  Int J Oral Maxillofac Implants       Date:  1991       Impact factor: 2.804

6.  Three-dimensional analyses of human bite-force magnitude and moment.

Authors:  T M van Eijden
Journal:  Arch Oral Biol       Date:  1991       Impact factor: 2.633

Review 7.  A multicenter report on osseointegrated oral implants.

Authors:  T Albrektsson
Journal:  J Prosthet Dent       Date:  1988-07       Impact factor: 3.426

8.  In vivo vertical forces on implants.

Authors:  E J Richter
Journal:  Int J Oral Maxillofac Implants       Date:  1995 Jan-Feb       Impact factor: 2.804

9.  Occlusal forces during chewing and swallowing as measured by sound transmission.

Authors:  C H Gibbs; P E Mahan; H C Lundeen; K Brehnan; E K Walsh; W B Holbrook
Journal:  J Prosthet Dent       Date:  1981-10       Impact factor: 3.426

Review 10.  Mechanical complications of dental implants.

Authors:  M S Schwarz
Journal:  Clin Oral Implants Res       Date:  2000       Impact factor: 5.977

View more
  1 in total

1.  Digital evaluation of axial displacement by implant-abutment connection type: An in vitro study.

Authors:  Sung-Jun Kim; KeunBaDa Son; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2018-10-22       Impact factor: 1.904

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.