Estimates of total analytical error in consumer and hospital glucose meters contributed by hematocrit, maltose, and ascorbate.

Martha E Lyon1, Jeffrey A DuBois, Gordon H Fick, Andrew W Lyon.   

Abstract

BACKGROUND: Patients and physicians expect accurate whole blood glucose monitoring even when patients are anemic, are undergoing peritoneal dialysis, or have slightly elevated ascorbate levels. The objective of this study was to estimate analytical error in two consumer and two hospital glucose meters contributed by variations in hematocrit, maltose, ascorbate, and imprecision.
METHODS: The influence of hematocrit (20-60%), maltose, and ascorbate were tested alone and in combination with each glucose meter and with a reference plasma glucose method at three concentrations of glucose. Precision was determined by consecutive analysis (n=20) at three levels of glucose. Multivariate regression analysis was used to estimate the bias associated with the interferences, alone and in combination. Total analytical error was estimated as |% bias|+1.96 (% imprecision).
RESULTS: Three meters demonstrated hematocrit bias that was dependent upon glucose concentration. Maltose had profound concentration-dependent positive bias on the consumer meters, and the extent of maltose bias was dependent on hematocrit. Ascorbate produced small but statistically significant biases on three meters. Coincident low hematocrit, presence of maltose, and presence of ascorbate increased the observed bias and was summarized by estimation of total analytical error. Among the four glucose meter devices assessed, estimates of total analytical error in glucose measurement ranged from 6 to 68% under the conditions tested.
CONCLUSIONS: The susceptibility of glucose meters to clinically significant analytical biases is highly device-dependent, and low hematocrit exacerbated the observed analytical error.
© 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21129345      PMCID: PMC3005060          DOI: 10.1177/193229681000400624

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  18 in total

Review 1.  Assessing the quality of glucose monitor studies: a critical evaluation of published reports.

Authors:  John Mahoney; John Ellison
Journal:  Clin Chem       Date:  2007-05-03       Impact factor: 8.327

2.  Managing quality vs. measuring uncertainty in the medical laboratory.

Authors:  James O Westgard
Journal:  Clin Chem Lab Med       Date:  2010       Impact factor: 3.694

3.  False elevation of blood glucose levels measured by GDH-PQQ-based glucometers occurs during all daily dwells in peritoneal dialysis patients using icodextrin.

Authors:  Chiu-Yeh Tsai; Su-Chu Lee; Chi-Chih Hung; Jia-Jung Lee; Mei-Chuan Kuo; Shang-Jyh Hwang; Hung-Chun Chen
Journal:  Perit Dial Int       Date:  2010-02-26       Impact factor: 1.756

4.  System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197.

Authors:  Guido Freckmann; Annette Baumstark; Nina Jendrike; Eva Zschornack; Serge Kocher; Jacques Tshiananga; Frank Heister; Cornelia Haug
Journal:  Diabetes Technol Ther       Date:  2010-03       Impact factor: 6.118

5.  The food and drug administration is now preparing to establish tighter performance requirements for blood glucose monitors.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2010-05-01

Review 6.  Analytical interferences in point-of-care testing glucometers by icodextrin and its metabolites: an overview.

Authors:  Katelijne M J Floré; Joris R Delanghe
Journal:  Perit Dial Int       Date:  2009 Jul-Aug       Impact factor: 1.756

7.  Assessing glucose monitor performance--a standardized approach.

Authors:  John J Mahoney; John M Ellison
Journal:  Diabetes Technol Ther       Date:  2007-12       Impact factor: 6.118

8.  Interference studies with two hospital-grade and two home-grade glucose meters.

Authors:  Martha E Lyon; Leland B Baskin; Sandy Braakman; Steven Presti; Jeffrey Dubois; Terry Shirey
Journal:  Diabetes Technol Ther       Date:  2009-10       Impact factor: 6.118

9.  Differences in glucose values obtained from point-of-care glucose meters and laboratory analysis in critically ill patients.

Authors:  Anjannette Cook; Delyla Laughlin; Margery Moore; Doreen North; Kathleen Wilkins; Gay Wong; Allyson Wallace-Scroggs; Lisa Halvorsen
Journal:  Am J Crit Care       Date:  2009-01       Impact factor: 2.228

Review 10.  A review of standards and statistics used to describe blood glucose monitor performance.

Authors:  Jan S Krouwer; George S Cembrowski
Journal:  J Diabetes Sci Technol       Date:  2010-01-01
View more
  10 in total

Review 1.  Blood glucose measurements in critically ill patients.

Authors:  Tom Van Herpe; Dieter Mesotten
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

2.  Analysis of an electrochemical blood glucose monitoring system with hematocrit compensation: improved accuracy by design.

Authors:  Martha E Lyon; Andrew W Lyon
Journal:  J Diabetes Sci Technol       Date:  2012-05-01

3.  Impact of High-Dose Intravenous Vitamin C for Treatment of Sepsis on Point-of-Care Blood Glucose Readings.

Authors:  Anna Peyton Howell; Jenna L Parrett; Daniel R Malcom
Journal:  J Diabetes Sci Technol       Date:  2019-11-25

4.  Analysis of the performance of the CONTOUR® TS Blood Glucose Monitoring System: when regulatory performance criteria are met, should we have confidence to use a medical device with all patients?

Authors:  Martha E Lyon; Andrew W Lyon
Journal:  J Diabetes Sci Technol       Date:  2011-01-01

Review 5.  Is there a relationship between mean blood glucose and glycated hemoglobin?

Authors:  Konstantinos Makris; Loukia Spanou
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

Review 6.  Glycemic targets and approaches to management of the patient with critical illness.

Authors:  Dieter Mesotten; Greet Van den Berghe
Journal:  Curr Diab Rep       Date:  2012-02       Impact factor: 4.810

7.  Correlation Between Capillary and Venous Blood Glucose in the Lactose Tolerance Test.

Authors:  José Luis Domínguez Jiménez; Antonio Fernández Suárez
Journal:  Dig Dis Sci       Date:  2015-08-30       Impact factor: 3.199

8.  N-Acetylcysteine Interference with a Glucose Dehydrogenase Linked Glucose Meter.

Authors:  Martha E Lyon; Andrew W Lyon
Journal:  J Diabetes Sci Technol       Date:  2021-03-14

9.  Evaluation of an Electrochemical Point-of-Care Meter for Measuring Glucose Concentration in Blood from Periparturient Dairy Cattle.

Authors:  A A Megahed; M W H Hiew; J R Townsend; J B Messick; P D Constable
Journal:  J Vet Intern Med       Date:  2015-10-20       Impact factor: 3.333

10.  Spurious elevation of glucose concentration during administration of high dose of ascorbic acid in a patient with type 2 diabetes on hemodialysis.

Authors:  Soo Kyoung Kim; Jong Ryeal Hahm; Ho-Su Kim; Sungsu Kim; Tae Sik Jung; Jung Hwa Jung; Hye Ryun Lee; Deok Ryong Kim
Journal:  Yonsei Med J       Date:  2013-09       Impact factor: 2.759

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.