| Literature DB >> 20927221 |
Biplab Sarkar1, Bhaswar Ghosh, Sukumaran Mahendramohan, Ayan Basu, Jyotirup Goswami, Amitabh Ray.
Abstract
The study was aimed to compare accuracy of monitor unit verification in intensity modulated radiation therapy (IMRT) using 6 MV photons by three different methodologies with different detector phantom combinations. Sixty patients were randomly chosen. Zero degree couch and gantry angle plans were generated in a plastic universal IMRT verification phantom and 30×30×30 cc water phantom and measured using 0.125 cc and 0.6 cc chambers, respectively. Actual gantry and couch angle plans were also measured in water phantom using 0.6 cc chamber. A suitable point of measurement was chosen from the beam profile for each field. When the zero-degree gantry, couch angle plans and actual gantry, couch angle plans were measured by 0.6 cc chamber in water phantom, the percentage mean difference (MD) was 1.35%, 2.94 % and Standard Deviation (SD) was 2.99%, 5.22%, respectively. The plastic phantom measurements with 0.125 cc chamber Semiflex ionisation chamber (SIC) showed an MD=4.21% and SD=2.73 %, but when corrected for chamber-medium response, they showed an improvement, with MD=3.38 % and SD=2.59 %. It was found that measurements with water phantom and 0.6cc chamber at gantry angle zero degree showed better conformity than other measurements of medium-detector combinations. Correction in plastic phantom measurement improved the result only marginally, and actual gantry angle measurement in a flat- water phantom showed higher deviation.Entities:
Keywords: Dose verification; point dose measurement; quality assurance
Year: 2010 PMID: 20927221 PMCID: PMC2936183 DOI: 10.4103/0971-6203.62129
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Measurement conditions
| Gantry angle zero degree | Gantry angle zero degree | Actual planned gantry angles |
| Couch angle zero degree | Couch angle zero degree | Actual planned couch angles |
| All profiles (AB/GT) were generated at fixed isocentric depth | All profiles (AB/GT) were generated at fixed isocentric depth | Only AB profiles were generated at the isocentric depth for all beams |
| Each profile individually analyzed to find a suitable position of measurement for each beam | Each profile individually analyzed to find a suitable position of measurement for each beam | All profiles were summed, and a suitable point of measurement was found |
| Only one shift, either AB (X) or GT (Y), was given, to reach the suitable point of measurement | Only one shift, either AB (X) or GT (Y), was given, to reach the suitable point of measurement | Only one shift in AB direction, all fields individually measured at that point |
| Variation in height was ignored, as it is not possible to measure by either qualitative or quantitative method | Variation in height was ignored, as it is not possible to measure by either qualitative or quantitative method | Variation in height was ignored, as it is not possible to measure by either qualitative or quantitative method |
Figure 1Chamber phantom calibration curve
Figure 2aFrequency plot against percentage deviation (X axis) of measured and calculated values for different square field sizes, ranging between 3×3 cm2 and 25×25 cm2, when measured by 0.6 cc (FIC) chamber in water phantom (WP)
Figure 2bFrequency plot against percentage deviation (X axis) of measured and calculated values for different square field sizes, ranging between 3×3 cm2 and 25×25 cm2, when measured by 0.125 cc (SIC) chamber in water phantom (WP)
Figure 2cFrequency plot against percentage deviation (X axis) of measured and calculated values, ranging between 3×3 cm2 and 25×25 cm2, measured by 0.125 cc (SIC) chamber in plastic phantom (PP)
Figure 3aFrequency plot against percentage difference (X axis) between measured and calculated doses when measured in a water phantom (WP) by a 0.6 cc (FIC) chamber for the plans with actual gantry and couch angles
Figure 3bFrequency plot against percentage difference (X axis) between the calculated and measured dose in water phantom (WP) - gantry angle, couch angle taken to be zero degree
Figure 3cFrequency plot against percentage difference (X axis) between measured and calculated doses when measured in plastic phantom (PP) by a 0.125 cc (SIC) chamber for the plans with all gantry and couch angles taken as zero degree
Figure 3dFrequency plot for plastic phantom measurement as shown in fig 3c when corrected by Chamber phantom calibration curve and Eq-3
Results of measurements with different detector-phantom combinations
| Number of fields | 301 | 301 | 301 | 301 |
| Mean | 2.95 | 1.36 | 4.22 | 3.38 |
| Median | 2.54 | 1.42 | 3.82 | 2.88 |
| Std. deviation | 5.22 | 2.99 | 2.74 | 2.59 |
| Range | 29.81 | 19.73 | 16.69 | 15.08 |
| Minimum | -14.66 | -7.92 | -1.48 | 0.83 |
| Maximum | 15.15 | 11.81 | 15.21 | 14.25 |
| Percent age of measured fields found not in the range (±5%) | 40% | 12.9% | 24% | 17% |