| Literature DB >> 10076972 |
A Boyer1, L Xing, C M Ma, B Curran, R Hill, A Kania, A Bleier.
Abstract
A treatment planning system to compute intensity modulated radiotherapy (IMRT) treatments using inverse planning was investigated. The system was designed to optimize the intensity patterns required to treat a specified target volume with specified normal structure constraints. A beam model that uses the convolution of pencil beams was used to compute the dose distributions. A multileaf collimator leaf-setting sequence intended to produce the intensity pattern was computed along with the monitor units required to deliver each of a number of fixed-gantry modulated fields. Computer calculations are commonly verified using an independent manual procedure. It is difficult to calculate treatment delivery monitor units for this variant of IMRT using manual methods. Since manual calculations are not feasible, it is important both to understand and to verify the calculation of treatment monitor units by the planning system algorithm. A formal analysis was made of the dose calculation model and the monitor unit calculation embedded in the algorithm. Experimental verification of the dose delivered by plans computed with the methodology demonstrated an agreement of better than 4% between the dose model and measurements.Mesh:
Substances:
Year: 1999 PMID: 10076972 DOI: 10.1118/1.598502
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071