Wear of ultra-high molecular weight polyethylene (UHMWPE) in total knee prostheses: a review of key influences.

T M McGloughlin1, A G Kavanagh.   

Abstract

The formation and development of wear is now widely accepted as one of the major concerns in the long-term survivorship of contemporary knee prostheses in vivo. This review examines the role of surface topography, third-body debris, load, contact mechanics and material quality in the wear process. Some of the kinematic and physiological issues that need to be modelled in the development of wear testing regimes for evaluation of material combinations and geometrical combinations in total knee implant designs are considered. Wear testing procedures and some of the results from wear tests are discussed and the need to consider the impact of rolling and sliding in the study of wear in total knee components is highlighted. The dominant wear mechanisms that occur in vivo are identified and the role of these mechanisms is currently being examined experimentally at the University of Limerick wear testing machine.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10997056     DOI: 10.1243/0954411001535390

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  8 in total

1.  Surface damage versus tibial polyethylene insert conformity: a retrieval study.

Authors:  Markus A Wimmer; Michel P Laurent; Jeannie D Haman; Joshua J Jacobs; Jorge O Galante
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

2.  In vivo oxidation contributes to delamination but not pitting in polyethylene components for total knee arthroplasty.

Authors:  Francisco J Medel; Steven M Kurtz; Javad Parvizi; Gregg R Klein; Matthew J Kraay; Clare M Rimnac
Journal:  J Arthroplasty       Date:  2010-09-28       Impact factor: 4.757

3.  What Is the Incidence of Cobalt-Chromium Damage Modes on the Bearing Surface of Contemporary Femoral Component Designs for Total Knee Arthroplasty?

Authors:  Christina M Arnholt; Daniel W MacDonald; Gregg R Klein; Harold E Cates; Clare M Rimnac; Steven M Kurtz; Sevi Kocagoz; Antonia F Chen
Journal:  J Arthroplasty       Date:  2018-05-17       Impact factor: 4.757

4.  Projections and Epidemiology of Revision Hip and Knee Arthroplasty in the United States to 2030.

Authors:  Andrew M Schwartz; Kevin X Farley; George N Guild; Thomas L Bradbury
Journal:  J Arthroplasty       Date:  2020-02-19       Impact factor: 4.757

5.  An 8- to 10-year review of the Rotaglide total knee replacement.

Authors:  M D Horwitz; S Awan; M B Chatoo; D J Stott; D P Powles
Journal:  Int Orthop       Date:  2007-12-13       Impact factor: 3.075

6.  Analysis of Carbon Fiber Reinforced PEEK Hinge Mechanism Articulation Components in a Rotating Hinge Knee Design: A Comparison of In Vitro and Retrieval Findings.

Authors:  Ronja A Schierjott; Alexander Giurea; Hans-Joachim Neuhaus; Jens Schwiesau; Andreas M Pfaff; Sandra Utzschneider; Gianluca Tozzi; Thomas M Grupp
Journal:  Biomed Res Int       Date:  2016-12-22       Impact factor: 3.411

7.  Surface wear in a custom manufactured temporomandibular joint prosthesis.

Authors:  Nikolas De Meurechy; Merve Kübra Aktan; Bart Boeckmans; Stijn Huys; Denis R Verwilghen; Annabel Braem; Maurice Y Mommaerts
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2022-01-28       Impact factor: 3.405

8.  Posterior cruciate ligament balancing in total knee arthroplasty: a numerical study with a dynamic force controlled knee model.

Authors:  Arnd Steinbrück; Matthias Woiczinski; Patrick Weber; Peter Ernst Müller; Volkmar Jansson; Christian Schröder
Journal:  Biomed Eng Online       Date:  2014-07-02       Impact factor: 2.819