| Literature DB >> 20860445 |
Eric Vandenbussche1, Mohammed Saffarini, Ulrich Hansen, Fabienne Taillieu, Céline Mutschler, Bernard Augereau, Thomas M Gregory.
Abstract
BACKGROUND: Current techniques for measuring in vivo polyethylene wear suffer from a range of problems, resulting in an unacceptable lack of repeatability and/or insufficient accuracy when they are used to measure the low wear rates associated with new, highly crosslinked polyethylene. We describe an improved CT method for measurement of 3D femoral head penetration in PE acetabular cups that has sufficient accuracy and repeatability to allow assessment of the wear potential of modern implants.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20860445 PMCID: PMC3214744 DOI: 10.3109/17453674.2010.519163
Source DB: PubMed Journal: Acta Orthop ISSN: 1745-3674 Impact factor: 3.717
Materials used in this study
| Group | Components | Quantity | Material | Model | Manufacturer | Articular diameter (mm) | Notes |
|---|---|---|---|---|---|---|---|
| A | Femoral Head | 1 | CoCr | Zimmer | 28 (exact) 27.957 | Unused implant with exact diameter determined by coordinate measurement machine | |
| B | Acetabular Cups | 5 | All-PE | Acoplot | Zimmer | 22.2 | All implants were worn artificially by machining the articular surface |
| 5 | MB-PE | Kappa Acora | Zimmer | 22.2 | |||
| C | Acetabular Cups | 2 | All-PE | PSA | Stratek | 26 | All implants were worn naturally in vivo and explanted at revision surgery |
| 2 | All-PE | Charnley | SEM | 32 | |||
| 2 | MB-PE | PSA | Stratek | 26 | |||
| 1 | MB-PE | Screw Ring | Benoist Girard | 28 | |||
| 1 | MB-PE | Spotorno Pressfit | Implex Corp | 28 |
CoCr = cobalt-chromium; MB = metal-backed; PE = polyethylene
Figure 1.3-dimensional image reconstructions of a CT-scanned all-poly acetabular component assembled with its matching femoral stem and 28-mm diameter head. The images were produced using the surface rendering function in OsiriX and illustrate the ‘partial voluming effect’ that increases with thickness of CT section slice: A. 5 mm; B. 2.5 mm; C. 1.25 mm; D. 0.625 mm.
Figure 2.Digitization of points on 3-dimensional image reconstructions of CT- scanned implants. A and B. An all-poly acetabular component assembled with its matching femoral stem and 28-mm diameter head. C and D. A metal-backed PE acetabular component assembled with its matching femoral stem and 32-mm diameter head.
Figure 3.Calculation using 3DReshaper, by the method of least squares, of the center and diameter of a spherical femoral head (panel A) and the center and diameter of a circular rim of a metal-backed acetabular component (panel B).
Accuracy and repeatability of intra- and interobserver CT measurements on unused CoCr femoral heads
| Group | Measurment | Mean | SD | Minimum | Maximum | SEM | 95% CI | Real error | ||
|---|---|---|---|---|---|---|---|---|---|---|
| A | Diameter | Mean intraobserver (mm) | 27.993 | 0.044 | 27.893 | 28.061 | 0.031 | 0.001 | 0.036 | |
| Diameter | Mean interobserver (mm) | 28.007 | 0.022 | 27.984 | 28.061 | 0.016 | 0.001 | 0.050 | ||
| Center | Mean intraobserver | x (mm) | 0.000 | 0.038 | –0.075 | 0.058 | 0.030 | 0.0007 | ||
| y (mm) | 0.000 | 0.021 | –0.027 | 0.032 | 0.018 | 0.0004 | ||||
| z (mm) | 0.000 | 0.096 | –0.183 | 0.164 | 0.065 | 0.0019 | ||||
| Center | Mean interobserver | x (mm) | –0.032 | 0.044 | –0.110 | 0.079 | 0.031 | 0.0008 | ||
| y (mm) | 0.025 | 0.025 | –0.012 | 0.048 | 0.018 | 0.0005 | ||||
| z (mm) | –0.071 | 0.065 | –0.183 | 0.090 | 0.053 | 0.0013 | ||||
SD: standard deviation;
SEM: standard error of the mean;
CI: Confidence interval;
Real error: the difference between the head diameter measured and the real head diameter (27.957mm)
Mean intraobserver measurements from 10 repeatedly blinded CT measurements by the same senior surgeon (EV)
Mean interobserver measurements from 10 repeatedly blinded CT measurements by 10 different surgeons.
Measurments of 3D femoral head penetration in explanted and machined acetabular cups
| Group | Component number | Material | Model | Manufacturer | Articular diameter (mm) | 3D femoral head penetration (mm) | ||
|---|---|---|---|---|---|---|---|---|
| Direct | CT | Real error | ||||||
| B | 1 | All-PE | Acoplot | Zimmer | 22.2 | 0.993 | 1.183 | 0.190 |
| 2 | All-PE | Acoplot | Zimmer | 22.2 | 2.001 | 1.990 | -0.011 | |
| 3 | All-PE | Acoplot | Zimmer | 22.2 | 2.994 | 3.076 | 0.082 | |
| 4 | All-PE | Acoplot | Zimmer | 22.2 | 4.001 | 3.939 | -0.062 | |
| 5 | All-PE | Acoplot | Zimmer | 22.2 | 4.995 | 4.750 | -0.245 | |
| 6 | MB-PE | Kappa Acora | Zimmer | 22.2 | 1.004 | 0.924 | –0.080 | |
| 7 | MB-PE | Kappa Acora | Zimmer | 22.2 | 1.994 | 2.006 | 0.012 | |
| 8 | MB-PE | Kappa Acora | Zimmer | 22.2 | 2.984 | 2.802 | –0.182 | |
| 9 | MB-PE | Kappa Acora | Zimmer | 22.2 | 4.002 | 4.119 | 0.117 | |
| 10 | MB-PE | Kappa Acora | Zimmer | 22.2 | 10.493 | 10.469 | –0.024 | |
| C | 11 | All-PE | PSA | Stratek | 26 | 5.475 | 5.448 | –0.027 |
| 12 | All-PE | PSA | Stratek | 26 | 2.128 | 2.020 | –0.108 | |
| 13 | All-PE | Charnley | SEM | 32 | 3.126 | 3.203 | 0.077 | |
| 14 | All-PE | Charnley | SEM | 32 | 3.229 | 3.181 | –0.048 | |
| 15 | MB-PE | PSA | Stratek | 26 | 3.130 | 3.042 | –0.088 | |
| 16 | MB-PE | PSA | Stratek | 26 | 4.071 | 4.080 | 0.009 | |
| 17 | MB-PE | Screw Ring | Benoist Girard | 28 | 3.943 | 3.920 | –0.023 | |
| 18 | MB-PE | Spotorno Pressfit | Implex Corp | 28 | 2.240 | 2.178 | –0.062 | |
MB: metal-backed; PE: polyethylene