Deriving the modulation transfer function of CT from extremely noisy edge profiles.

Issei Mori1, Yoshio Machida.   

Abstract

The point spread function (PSF) method is currently the one predominantly used to determine the modulation transfer function (MTF) of an X-ray CT system. However, the image examined with the PSF method must have a very high contrast-to-noise ratio (CNR); it must also be reconstructed with a fine pixel pitch using a zooming reconstruction. Therefore, the PSF method is often inappropriate for describing the MTF of clinical operating conditions when image linearity is not guaranteed. The edge spread function (ESF) method requires no zooming reconstruction, but its susceptibility to image noise is no better than that of the PSF method. We describe a technique for rendering the ESF method robust to image noise. We smooth out the noisy ESF through multiple stages of filtering. Invariably, the line spread function (LSF) obtained from the smoothed ESF is blurred, and the MTF obtained from the LSF is incorrect. However, because the filtering that has been applied is known, much of the LSF blurring can be corrected. An estimate of the true LSF is obtainable from the blurred LSF, assuming that the true LSF is not very different from either a Gaussian or a composite of multiple Gaussians. For an image reconstructed with a kernel for soft-tissue imaging, the MTF obtained by our method is sufficiently consistent with the theoretical MTF, even when the CNR is as low as 2.

Mesh:

Year:  2008        PMID: 20821125     DOI: 10.1007/s12194-008-0039-9

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  14 in total

1.  Determination of the presampled MTF in computed tomography.

Authors:  J M Boone
Journal:  Med Phys       Date:  2001-03       Impact factor: 4.071

2.  Accuracy of a simple method for deriving the presampled modulation transfer function of a digital radiographic system from an edge image.

Authors:  Egbert Buhr; Susanne Günther-Kohfahl; Ulrich Neitzel
Journal:  Med Phys       Date:  2003-09       Impact factor: 4.071

3.  New method of evaluating edge-preserving adaptive filters for computed tomography (CT): digital phantom method.

Authors:  Miwa Okumura; Takamasa Ota; Shinsuke Tsukagoshi; Kazuhiro Katada
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2006-07-20

4.  A simple method for determining the modulation transfer function in digital radiography.

Authors:  H Fujita; D Y Tsai; T Itoh; K Doi; J Morishita; K Ueda; A Ohtsuka
Journal:  IEEE Trans Med Imaging       Date:  1992       Impact factor: 10.048

5.  The effect of dose reduction and feasibility of edge-preserving noise reduction on the detection of liver lesions using MSCT.

Authors:  Johannes Wessling; Rainer Esseling; Rainer Raupach; Stefanie Fockenberg; Nani Osada; Joachim Gerss; Walter Heindel; Roman Fischbach
Journal:  Eur Radiol       Date:  2007-01-12       Impact factor: 5.315

Review 6.  Transfer function analysis of radiographic imaging systems.

Authors:  C E Metz; K Doi
Journal:  Phys Med Biol       Date:  1979-11       Impact factor: 3.609

7.  The line spread function and modulation transfer function of a computed tomographic scanner.

Authors:  P F Judy
Journal:  Med Phys       Date:  1976 Jul-Aug       Impact factor: 4.071

8.  A simplified approach for modulation transfer function determinations in computed tomography.

Authors:  E L Nickoloff; R Riley
Journal:  Med Phys       Date:  1985 Jul-Aug       Impact factor: 4.071

9.  Theoretical resolution of computed tomography systems.

Authors:  G H Glover; R L Eisner
Journal:  J Comput Assist Tomogr       Date:  1979-02       Impact factor: 1.826

10.  Computed tomography with fan beam geometry.

Authors:  T M Peters; R M Lewitt
Journal:  J Comput Assist Tomogr       Date:  1977-10       Impact factor: 1.826

View more
  7 in total

1.  Aliased noise in X-ray CT images and band-limiting processing as a preventive measure.

Authors:  Kazuhiro Sato; Miho Shidahara; Mitsunori Goto; Isao Yanagawa; Noriyasu Homma; Issei Mori
Journal:  Radiol Phys Technol       Date:  2015-01-11

2.  An evaluation of spatial resolution of a prototype proton CT scanner.

Authors:  Tia E Plautz; V Bashkirov; V Giacometti; R F Hurley; R P Johnson; P Piersimoni; H F-W Sadrozinski; R W Schulte; A Zatserklyaniy
Journal:  Med Phys       Date:  2016-12       Impact factor: 4.071

3.  Modulation transfer function measurement of CT images by use of a circular edge method with a logistic curve-fitting technique.

Authors:  Tomomi Takenaga; Shigehiko Katsuragawa; Makoto Goto; Masahiro Hatemura; Yoshikazu Uchiyama; Junji Shiraishi
Journal:  Radiol Phys Technol       Date:  2014-08-21

4.  Spatial resolution measurement for iterative reconstruction by use of image-averaging techniques in computed tomography.

Authors:  Atsushi Urikura; Katsuhiro Ichikawa; Takanori Hara; Eiji Nishimaru; Yoshihiro Nakaya
Journal:  Radiol Phys Technol       Date:  2014-06-01

5.  Tilted-wire method for measuring resolution properties of CT images under extremely low-contrast and high-noise conditions.

Authors:  Chiaki Tominaga; Hiroki Azumi; Mitsunori Goto; Masaaki Taura; Noriyasu Homma; Issei Mori
Journal:  Radiol Phys Technol       Date:  2018-02-23

6.  Impact of ROI Size on the Accuracy of Noise Measurement in CT on Computational and ACR Phantoms.

Authors:  Choirul Anam; Pandji Triadyaksa; Ariij Naufal; Zaenal Arifin; Zaenul Muhlisin; Evi Setiawati; Wahyu Setia Budi
Journal:  J Biomed Phys Eng       Date:  2022-08-01

7.  Helium ion beam imaging for image guided ion radiotherapy.

Authors:  M Martišíková; T Gehrke; S Berke; G Aricò; O Jäkel
Journal:  Radiat Oncol       Date:  2018-06-14       Impact factor: 3.481

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.