The development of aortic insufficiency in left ventricular assist device-supported patients.

Jennifer Cowger1, Francis D Pagani, Jonathan W Haft, Matthew A Romano, Keith D Aaronson, Theodore J Kolias.   

Abstract

BACKGROUND: Aortic insufficiency (AI) following left ventricular assist device (LVAD) placement can affect device performance. The aim of this study was to examine AI development following LVAD implantation. METHODS AND
RESULTS: Echocardiograms (n=315) from 78 subjects undergoing HeartMate-XVE (n=25 [32%]) or HeartMate-II (n=53 [68%]) implantations from 2004 to 2008 were reviewed. Studies were obtained preoperatively and at 1, 3, 6, 12, 18, and 24 months after surgery. AI was graded on an interval scale (0=none, 0.5=trivial, 1=mild, 1.5=mild-moderate, 2=moderate, 2.5=moderate-severe, 3=severe), and the change in AI at follow-up was analyzed with significance tests. Kaplan-Meier estimates for freedom from moderate or worse AI at follow-up were generated. Mixed-model linear regression was used to identify correlates of AI progression during LVAD support. The median (25th, 75th percentile) duration of LVAD support was 239 (112, 455) days, and preoperative AI grade was 0.0 (0.0, 0.0). At 6 months, 89±4% of subjects (n=49 at risk) were free from moderate or worse AI, but this was reduced to 74±7% (n=29 at risk) and 49±13% (n=13 at risk) by 12 and 18 months, respectively. Correlates (slope±SE) of AI progression included female sex (0.002±0.001; P=0.01), smaller body surface area (-0.003±0.001 per m(2); P=0.0017), and HeartMate-II model type (0.002±0.001; P=0.039). Correlates (β±SE) of progressive AI on postoperative echocardiogram included increasing aortic sinus diameter (0.04±0.01 per mm; P=0.001), an aortic valve that remained closed (0.42±0.06; P<0.001) or only intermittently opened (0.34±0.09; P<0.001), and lower left ventricular diastolic (-0.002±0.0004 per cm(3); P<0.001) and systolic (-0.002±0.0004 per cm(3); P<0.001) volumes.
CONCLUSIONS: AI progresses over time in LVAD-supported patients. As we move toward an era of long-term cardiac support, more studies are needed to determine the clinical significance of these findings.

Entities:  

Mesh:

Year:  2010        PMID: 20739615      PMCID: PMC3089421          DOI: 10.1161/CIRCHEARTFAILURE.109.917765

Source DB:  PubMed          Journal:  Circ Heart Fail        ISSN: 1941-3289            Impact factor:   8.790


  14 in total

1.  Long-term use of a left ventricular assist device for end-stage heart failure.

Authors:  E A Rose; A C Gelijns; A J Moskowitz; D F Heitjan; L W Stevenson; W Dembitsky; J W Long; D D Ascheim; A R Tierney; R G Levitan; J T Watson; P Meier; N S Ronan; P A Shapiro; R M Lazar; L W Miller; L Gupta; O H Frazier; P Desvigne-Nickens; M C Oz; V L Poirier
Journal:  N Engl J Med       Date:  2001-11-15       Impact factor: 91.245

2.  Insufficiency of the native aortic valve and left ventricular assist system inflow valve after support with an implantable left ventricular assist system: signs, symptoms, and concerns.

Authors:  L E Samuels; M P Thomas; E C Holmes; J Narula; J Fitzpatrick; D Wood; B Fyfe; A S Wechsler
Journal:  J Thorac Cardiovasc Surg       Date:  2001-08       Impact factor: 5.209

3.  Acquired commissural fusion of aortic valves in patients with left ventricular assist devices.

Authors:  John H Connelly; Jacki Abrams; Tomas Klima; William K Vaughn; O H Frazier
Journal:  J Heart Lung Transplant       Date:  2003-12       Impact factor: 10.247

4.  Recommendations for chamber quantification: a report from the American Society of Echocardiography's Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology.

Authors:  Roberto M Lang; Michelle Bierig; Richard B Devereux; Frank A Flachskampf; Elyse Foster; Patricia A Pellikka; Michael H Picard; Mary J Roman; James Seward; Jack S Shanewise; Scott D Solomon; Kirk T Spencer; Martin St John Sutton; William J Stewart
Journal:  J Am Soc Echocardiogr       Date:  2005-12       Impact factor: 5.251

Review 5.  What's wrong with Bonferroni adjustments.

Authors:  T V Perneger
Journal:  BMJ       Date:  1998-04-18

6.  Morphologic changes of the aortic wall due to reduced systemic pulse pressure in prolonged non pulsatile left heart bypass.

Authors:  T Nishimura; E Tatsumi; S Takaichi; Y Taenaka; Y Wakisaka; T Nakatani; T Masuzawa; Y Takewa; M Nakamura; S Endo; Y S Sohn; H Takano
Journal:  ASAIO J       Date:  1997 Sep-Oct       Impact factor: 2.872

7.  Surgical management of valvular disease in patients requiring left ventricular assist device support.

Authors:  V Rao; J P Slater; N M Edwards; Y Naka; M C Oz
Journal:  Ann Thorac Surg       Date:  2001-05       Impact factor: 4.330

8.  Partial aortic valve fusion induced by left ventricular assist device.

Authors:  A G Rose; S J Park; A J Bank; L W Miller
Journal:  Ann Thorac Surg       Date:  2000-10       Impact factor: 4.330

9.  Effects of left ventricular assist device support and outflow graft location upon aortic blood flow.

Authors:  Kenneth N Litwak; Steven C Koenig; Hiroyuki Tsukui; Shin'ichiro Kihara; Zhongjun Wu; George M Pantalos
Journal:  ASAIO J       Date:  2004 Sep-Oct       Impact factor: 2.872

10.  Fusion of aortic valve commissures in patients supported by a continuous axial flow left ventricular assist device.

Authors:  James O Mudd; Jonathan D Cuda; Marc Halushka; Karl A Soderlund; John V Conte; Stuart D Russell
Journal:  J Heart Lung Transplant       Date:  2008-10-26       Impact factor: 10.247

View more
  55 in total

Review 1.  Echocardiographic assessment for ventricular assist device placement.

Authors:  Antolin S Flores; Michael Essandoh; Gregory C Yerington; Amar M Bhatt; Manoj H Iyer; William Perez; Victor R Davila; Ravi S Tripathi; Katja Turner; Galina Dimitrova; Michael J Andritsos
Journal:  J Thorac Dis       Date:  2015-12       Impact factor: 2.895

Review 2.  Left ventricular assist device implantation strategies and outcomes.

Authors:  LaVone A Smith; Leora T Yarboro; Jamie L W Kennedy
Journal:  J Thorac Dis       Date:  2015-12       Impact factor: 2.895

3.  Analysis of aortic valve commissural fusion after support with continuous-flow left ventricular assist device.

Authors:  Jerson R Martina; Marguerite E I Schipper; Nicolaas de Jonge; Faiz Ramjankhan; Roel A de Weger; Jaap R Lahpor; Aryan Vink
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-06-24

Review 4.  Management of aortic insufficiency in the continuous flow left ventricular assist device population.

Authors:  Jonathan Holtz; Jeffrey Teuteberg
Journal:  Curr Heart Fail Rep       Date:  2014-03

Review 5.  [Ventricular long-term support with implantable continuous flow pumps: on the way to a gold standard in the therapy of terminal heart failure].

Authors:  T Krabatsch; E Potapov; S Soltani; M Dandel; V Falk; C Knosalla
Journal:  Herz       Date:  2015-04       Impact factor: 1.443

6.  HeartMate 3-a "Step" in the right direction.

Authors:  Juglans Alvarez; Vivek Rao
Journal:  J Thorac Dis       Date:  2017-05       Impact factor: 2.895

7.  The angle of the outflow graft to the aorta can affect recirculation due to aortic insufficiency under left ventricular assist device support.

Authors:  Kei Iizuka; Tomohiro Nishinaka; Daichi Akiyama; Hirohito Sumikura; Toshihide Mizuno; Tomonori Tsukiya; Yoshiaki Takewa; Kenji Yamazaki; Eisuke Tatsumi
Journal:  J Artif Organs       Date:  2018-07-23       Impact factor: 1.731

8.  Assessment of myocardial viability and left ventricular function in patients supported by a left ventricular assist device.

Authors:  Deepak K Gupta; Hicham Skali; Jose Rivero; Patricia Campbell; Leslie Griffin; Colleen Smith; Courtney Foster; Brian Claggett; Robert J Glynn; Gregory Couper; Michael M Givertz; Mandeep R Mehra; Marcelo Di Carli; Scott D Solomon; Marc A Pfeffer
Journal:  J Heart Lung Transplant       Date:  2014-01-27       Impact factor: 10.247

9.  Preservation of native aortic valve flow and full hemodynamic support with the TORVAD using a computational model of the cardiovascular system.

Authors:  Jeffrey R Gohean; Mitchell J George; Kay-Won Chang; Erik R Larson; Thomas D Pate; Mark Kurusz; Raul G Longoria; Richard W Smalling
Journal:  ASAIO J       Date:  2015 May-Jun       Impact factor: 2.872

10.  Living Without a Pulse: The Vascular Implications of Continuous-Flow Left Ventricular Assist Devices.

Authors:  Suneet N Purohit; William K Cornwell; Jay D Pal; JoAnn Lindenfeld; Amrut V Ambardekar
Journal:  Circ Heart Fail       Date:  2018-06       Impact factor: 8.790

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.