Volume-of-interest (VOI) imaging in C-arm flat-detector CT for high image quality at reduced dose.

Daniel Kolditz1, Yiannis Kyriakou, Willi A Kalender.   

Abstract

PURPOSE: A novel method for flat-detector computed tomography was developed to enable volume-of-interest (VOI) imaging at high resolution, low noise, and reduced dose. For this, a full low-dose overview (OV) scan and a local high-dose scan of a VOI are combined.
METHODS: The first scan yields an overview of the whole object and enables the selection of an arbitrary VOI. The second scan of that VOI assures high image quality within the interesting volume. The combination of the two consecutive scans is based on a forward projection of the reconstructed OV volume that was registered to the VOI. The artificial projection data of the OV scan are combined with the measured VOI data in the raw data domain. Different projection values are matched by an appropriate transformation and weighting. The reconstruction is performed with a standard Feldkamp-type algorithm. In simulations, the combination of OV scan and VOI scan was investigated on a mathematically described phantom. In measurements, spatial resolution and noise were evaluated with image quality phantoms. Modulation transfer functions and noise values were calculated. Measurements of an anthropomorphic head phantom were used to validate the proposed method for realistic applications, e.g., imaging stents. In Monte Carlo simulations, 3D dose distributions were calculated and dose values were assessed quantitatively.
RESULTS: By the proposed combination method, an image is generated which covers the whole object and provides the VOI at high image quality. In the OV image, a resolution of 0.7 lp/mm (line pairs per millimeter) and noise of 63.5 HU were determined. Inside the VOI, resolution was increased to 2.4 lp/mm and noise was decreased to 18.7 HU. For the performed measurements, the cumulative dose was significantly reduced in comparison to conventional scans by up to 93%. The dose of a high-quality scan, for example, was reduced from 97 to less than 7 mGy, while keeping image quality constant within the VOI.
CONCLUSIONS: The proposed VOI application with two scans is an effective way to ensure high image quality within the VOI while simultaneously reducing the cumulative patient dose.

Entities:  

Mesh:

Year:  2010        PMID: 20632582     DOI: 10.1118/1.3427641

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  High resolution dual detector volume-of-interest cone beam breast CT--Demonstration with a bench top system.

Authors:  Youtao Shen; Ying Yi; Yuncheng Zhong; Chao-Jen Lai; Xinming Liu; Zhicheng You; Shuaiping Ge; Tianpeng Wang; Chris C Shaw
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

2.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT.

Authors:  Cynthia H McCollough; Guang Hong Chen; Willi Kalender; Shuai Leng; Ehsan Samei; Katsuyuki Taguchi; Ge Wang; Lifeng Yu; Roderic I Pettigrew
Journal:  Radiology       Date:  2012-06-12       Impact factor: 11.105

3.  Volume-of-interest imaging of the inner ear in a human temporal bone specimen using a robot- driven C-arm flat panel detector CT system.

Authors:  D Kolditz; T Struffert; Y Kyriakou; A Bozzato; A Dörfler; W A Kalender
Journal:  AJNR Am J Neuroradiol       Date:  2011-08-11       Impact factor: 3.825

4.  High-resolution spiral CT of the breast at very low dose: concept and feasibility considerations.

Authors:  Willi A Kalender; Marcel Beister; John M Boone; Daniel Kolditz; Sabrina V Vollmar; Michaela C C Weigel
Journal:  Eur Radiol       Date:  2011-06-09       Impact factor: 5.315

5.  Reconstructing cone-beam CT with spatially varying qualities for adaptive radiotherapy: a proof-of-principle study.

Authors:  Wenting Lu; Hao Yan; Xuejun Gu; Zhen Tian; Ouyang Luo; Liu Yang; Linghong Zhou; Laura Cervino; Jing Wang; Steve Jiang; Xun Jia
Journal:  Phys Med Biol       Date:  2014-09-26       Impact factor: 3.609

6.  Known-component 3D image reconstruction for improved intraoperative imaging in spine surgery: A clinical pilot study.

Authors:  Xiaoxuan Zhang; Ali Uneri; J Webster Stayman; Corinna C Zygourakis; Sheng-Fu L Lo; Nicholas Theodore; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2019-06-30       Impact factor: 4.071

7.  Feasibility of flat panel angiographic CT after intravenous contrast agent application in the postoperative evaluation of patients with clipped aneurysms.

Authors:  M-N Psychogios; D Wachter; A Mohr; P Schramm; A-M Frölich; K Jung; V Rohde; M Knauth
Journal:  AJNR Am J Neuroradiol       Date:  2011-08-18       Impact factor: 3.825

8.  Volume-of-interest Imaging Using Multiple Aperture Devices.

Authors:  W Wang; G J Gang; J H Siewerdsen; J W Stayman
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2019-03-01

9.  Experimental realization of fluence field modulated CT using digital beam attenuation.

Authors:  T P Szczykutowicz; C A Mistretta
Journal:  Phys Med Biol       Date:  2014-02-20       Impact factor: 3.609

10.  How Low Can We Go in Radiation Dose for the Data-Completion Scan on a Research Whole-Body Photon-Counting Computed Tomography System.

Authors:  Zhicong Yu; Shuai Leng; Zhoubo Li; Ahmed F Halaweish; Steffen Kappler; Erik L Ritman; Cynthia H McCollough
Journal:  J Comput Assist Tomogr       Date:  2016 Jul-Aug       Impact factor: 1.826

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.