Comparison of PACS and hard-copy 51-inch radiographs for measuring leg length and deformity.

Saurabh Khakharia1, Daniel Bigman, Austin T Fragomen, Helene Pavlov, S Robert Rozbruch.   

Abstract

BACKGROUND: The picture archiving and communication system (PACS) eventually will replace the use of standard hard-copy radiographs. It is unknown whether measurements of limb length discrepancy (LLD) and deformity on PACS compare in accuracy and reproducibility with those from hard-copy radiographs. QUESTIONS/PURPOSES: We compared the reproducibility and reliability of LLD and deformity measurements for each of these two media.
METHODS: We retrospectively reviewed 51-inch standing lower extremity images obtained for LLD or deformity analysis from 40 patients to compare the measurements and their reliability on hard-copy film with those performed on soft-copy PACS. Two observers independently performed measurements twice using each system at 1-week intervals to minimize interobserver or intraobserver bias. Intraclass correlation coefficients (ICCs) were determined to test intraobserver and interrater reliability of Rater 1 and Rater 2.
RESULTS: Interrater reliability of measurements made on hard copy ranged from 0.69 to 0.99 and PACS-derived measurements ranged from 0.66 to 0.98. Intraobserver reliability for Rater 1 for measurements made on hard copy ranged from 0.853 to 0.999 and PACS-derived measurements ranged from 0.80 to 0.996. Intraobserver reliability for Rater 2 for measurements made on hard copy ranged from 0.931 to 0.999 and PACS-derived measurements ranged from 0.962 to 0.999.
CONCLUSIONS: Each system yielded comparable reliability for measurements, therefore, transition to PACS to perform measurements in patients with LLD or deformity can be made with confidence. LEVEL OF EVIDENCE: Level III, diagnostic study. See the Guidelines for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Year:  2010        PMID: 20625949      PMCID: PMC3008877          DOI: 10.1007/s11999-010-1460-z

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  12 in total

1.  PACS in sonography: accuracy of interpretation using film compared with monitor display. Picture archiving and communication systems.

Authors:  B S Hertzberg; M A Kliewer; E K Paulson; D H Sheafor; K S Freed; J D Bowie; B A Carroll; D H DeLong; R C Nelson
Journal:  AJR Am J Roentgenol       Date:  1999-11       Impact factor: 3.959

2.  Accuracy of interpretation of CT scans: comparing PACS monitor displays and hard-copy images.

Authors:  Bruce I Reiner; Eliot L Siegel; Frank J Hooper
Journal:  AJR Am J Roentgenol       Date:  2002-12       Impact factor: 3.959

3.  A comparison of conventional film, CR hard copy and PACS soft copy images of the chest: analyses of ROC curves and inter-observer agreement.

Authors:  Gwyneth C Weatherburn; Deborah Ridout; Nicola H Strickland; Peter Robins; Christine M Glastonbury; Walter Curati; Chris Harvey; Clair Shadbolt
Journal:  Eur J Radiol       Date:  2003-09       Impact factor: 3.528

4.  Alignment and articular orientation of lower limbs: manual vs computer-aided measurements on digital radiograms.

Authors:  Umberto Rozzanigo; Andrea Pizzoli; Chiara Minari; Roberto Caudana
Journal:  Radiol Med       Date:  2005-03       Impact factor: 3.469

5.  Reporting requirements for skeletal digital radiography: comparison of soft-copy and hard-copy presentation.

Authors:  P J O'Connor; A G Davies; R C Fowler; D J Lintott; R F Bury; G J Parkin; D Martinez; A Saifuddin; A R Cowen
Journal:  Radiology       Date:  1998-04       Impact factor: 11.105

6.  Computed radiography for characterisation of the weight-bearing knee.

Authors:  J Sanfridsson; G Svahn; K Jonsson; L Ryd
Journal:  Acta Radiol       Date:  1997-07       Impact factor: 1.990

7.  Computed radiographic imaging: use in low-dose leg length radiography.

Authors:  M S Kogutt
Journal:  AJR Am J Roentgenol       Date:  1987-06       Impact factor: 3.959

8.  Orthopaedic measurements with computed radiography. Methodological development, accuracy, and radiation dose with special reference to the weight-bearing lower extremity and the dislocating patella.

Authors:  J Sanfridsson
Journal:  Acta Radiol Suppl       Date:  2001

9.  Fracture interpretation using electronic presentation: a comparison.

Authors:  L W Bancroft; T H Berquist; R L Morin; J H Pietan; J M Knudsen; H J Williams
Journal:  J Digit Imaging       Date:  2000-02       Impact factor: 4.056

10.  Quantification of axial alignment of the lower extremity on conventional and digital total leg radiographs.

Authors:  J Sailer; M Scharitzer; P Peloschek; A Giurea; H Imhof; S Grampp
Journal:  Eur Radiol       Date:  2004-08-05       Impact factor: 5.315

View more
  14 in total

1.  Femoral lengthening with lengthening over a nail has fewer complications than intramedullary skeletal kinetic distraction.

Authors:  Shahab Mahboubian; Matthew Seah; Austin T Fragomen; S Robert Rozbruch
Journal:  Clin Orthop Relat Res       Date:  2011-12-06       Impact factor: 4.176

2.  [Spectrum of indications for intramedullary or external fixators for axis correction and limb lengthening].

Authors:  F Schiedel; R Rödl
Journal:  Orthopade       Date:  2013-12       Impact factor: 1.087

3.  Evaluation of lower limb axial alignment using digital radiography stitched films in pre-operative planning for total knee replacement.

Authors:  Michael J Neil; Jereme B Atupan; Juan Paulo L Panti; Robert A J Massera; Stewart Howard
Journal:  J Orthop       Date:  2016-06-29

4.  Distal femoral osteotomy: is internal fixation better than external?

Authors:  K T Matthew Seah; Raheel Shafi; Austin T Fragomen; S Robert Rozbruch
Journal:  Clin Orthop Relat Res       Date:  2011-01-06       Impact factor: 4.176

5.  Bone Ninja application is reliable alternative to PACS in measuring preoperative and postoperative alignment parameters of total knee arthroplasty.

Authors:  Nuthan Jagadeesh; Ankur Kariya; Srinath Pammi; Vishwanath M Shivalingappa; Chethana Ramakrishna
Journal:  J Orthop       Date:  2022-08-12

6.  Brain MRI measurements at a term-equivalent age and their relationship to neurodevelopmental outcomes.

Authors:  H W Park; H-K Yoon; S B Han; B S Lee; I Y Sung; K S Kim; E A Kim
Journal:  AJNR Am J Neuroradiol       Date:  2013-08-29       Impact factor: 3.825

7.  A Comparison of Femoral Lengthening Methods Favors the Magnetic Internal Lengthening Nail When Compared with Lengthening Over a Nail.

Authors:  Austin T Fragomen; Anton M Kurtz; Jonathan R Barclay; Joseph Nguyen; S Robert Rozbruch
Journal:  HSS J       Date:  2018-01-05

8.  Proximal tibial osteotomy for genu varum: Radiological evaluation of deformity correction with a plate vs external fixator.

Authors:  S Ali Ghasemi; David T Zhang; Austin Fragomen; S Robert Rozbruch
Journal:  World J Orthop       Date:  2021-03-18

9.  Biomechanical and Functional Improvements Gained by Proximal Tibia Osteotomy Correction of Genu Varum in Patients with Knee Pain.

Authors:  Rachael J Da Cunha; Andrew P Kraszewski; Howard J Hillstrom; Austin T Fragomen; S Robert Rozbruch
Journal:  HSS J       Date:  2019-03-19

10.  Comparison of PACS and Bone Ninja mobile application for assessment of lower extremity limb length discrepancy and alignment.

Authors:  Amanda T Whitaker; Martin G Gesheff; Julio J Jauregui; John E Herzenberg
Journal:  J Child Orthop       Date:  2016-07-22       Impact factor: 1.548

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.