Juan Xu1, Hiroshi Ishikawa, Gadi Wollstein, Richard A Bilonick, Larry Kagemann, Jamie E Craig, David A Mackey, Alex W Hewitt, Joel S Schuman.
Abstract
PURPOSE: To develop a fully automated algorithm (AP) to perform a volumetric measure of the optic disc using conventional stereoscopic optic nerve head (ONH) photographs, and to compare algorithm-produced parameters with manual photogrammetry (MP), scanning laser ophthalmoscope (SLO) and optical coherence tomography (OCT) measurements.Entities:
Mesh:
Year: 2010 PMID: 20588996 PMCID: PMC2913866 DOI: 10.1364/OE.18.011347
Source DB: PubMed Journal: Opt Express ISSN: 1094-4087 Impact factor: 3.894
Fig. 1: Overview of the process flow
Fig. 2: Parallel stereo image configuration
Fig. 3Optical model of the eye and stereo fundus camera (adapted from Deguchi’s [15]).
Fig. 4: Geometric model of optic nerve head (ONH) with cross sectional profile and projected image. (A) Locating cup margin points in deformable model algorithm. (B) Cross sectional profile of 3D ONH model, with definitions of various ONH parameters, i.e., disc area, rim area, cup area, rim volume and cup volume. (C) The definitions of vertical C/D ratio and C/D area ratio.
Fig. 5Comparison between automatically photogrammetric algorithm and manual disc demarcation. (A) A successful example of disc demarcation. (B) An example of failure in disc demarcation due to prominent peripapillary atrophy.
: Comparison of optic disc measurements (mean ± standard deviation) from automated photogrammetry (AP), manual photogrammetry (MP), Heidelberg retina tomography (HRT) and optical coherence tomography (OCT).
| AP | MP | HRT | OCT | |
|---|---|---|---|---|
| Disc area (mm2) | 2.61 ± 0.58 | 2.53 ± 0.58* | 2.10 ± 0.51* | 2.47 ± 0.63* |
| Rim area (mm2) | 1.59 ± 0.45 | 1.56 ± 0.39 | 1.31 ± 0.34* | 1.46 ± 0.48 |
| Cup area (mm2) | 1.02 ± 0.53 | 0.97 ± 0.49 | 0.79 ± 0.52* | 1.02 ± 0.68 |
| Cup/disc area ratio | 0.38 ± 0.15 | 0.37 ± 0.14 | 0.36 ± 0.18 | 0.39 ± 0.21 |
| Vertical cup/disc ratio | 0.62 ± 0.12 | 0.61 ± 0.12 | 0.57 ± 0.16 | 0.58 ± 0.17 |
| Rim volume (mm3) | 0.23 ± 0.15 | 0.25 ± 0.14 | 0.33 ± 0.15 | 0.34 ± 0.25 |
| Cup volume (mm3) | 0.28 ± 0.28 | 0.26 ± 0.23 | 0.21 ± 0.19 | 0.20 ± 0.19* |
*Comparison with automated photogrammetry was computed only for parameters with constant bias. For all these parameters a statistically significant (p < 0.05) difference was noted.
Imprecision standard deviation of optic disc parameters measured by automated photogrammetry (AP), manual photogrammetry (MP), Heidelberg retina tomography (HRT) and optical coherence tomography (OCT).
| AP | MP | HRT | OCT | |
|---|---|---|---|---|
| Disc area (mm2) | 0.1321 | 0.2463 | 0.2847 | 0.3569 |
| Rim area (mm2) | 0.1572 | 0.0493 | 0.1267 | 0.2029 |
| Cup area (mm2) | 0.2400 | 0.2165 | 0.1754 | 0.1940 |
| Cup/disc area ratio | 0.1070 | 0.0474 | 0.0751 | 0.0905 |
| Vertical cup/disc ratio | 0.1021 | 0.0350 | 0.1062 | 0.0947 |
| Rim volume (mm3) | 0.1164 | 0.1093 | 0.1490 | 0.2724 |
| Cup volume (mm3) | 0.0752 | 0.1433 | 0.1285 | 0.1097 |
Relative imprecision standard deviation ratio (adjusted by bias, 95% confidence interval) of optic disc parameters measured by automated photogrammetry (AP), manual photogrammetry (MP), Heidelberg retina tomography (HRT) and optical coherence tomography (OCT).
|
|
|
| |
|---|---|---|---|
| Disc area (mm2) | 2.0054* [1.4364, 2.7826] | 2.8082* [2.0485, N/A] | 2.9363* [2.1114, N/A] |
| Rim area (mm2) | 0.1164* [0.0153, 0.4121] | 0.4610 [0.0049, N/A] | 0.7878 [0.0003, N/A] |
| Cup area (mm2) | 0.9793 [0.7119, N/A] | 0.7018 [0.5193, 0.7018] | 0.5041 [0.4285, N/A] |
| Cup/disc area ratio | 0.3718* [0.2250, 0.5978] | 0.5119* [0.3173, 0.8039] | 0.4507* [0.2778, 0.7088] |
| Vertical cup/disc ratio | 0.2189* [0.1715, 0.3155] | 0.6027* [0.3542, 0.8765] | 0.4790* [0.5559 1.1572] |
| Rim volume (mm3) | 1.9669 [0.9819, 5.2616] | 15.6618* [2.6423, N/A] | 8.8209* [2.2745, N/A] |
| Cup volume (mm3) | 2.6641* [1.9456, N/A] | 2.4498* [1.7998, 3.4187] | 1.8050* [1.3424, 2.4529] |
*P < 0.05. N/A – value did not converge.
Linear calibration between optic nerve head measurements of automated photogrammetry (AP), manual photogrammetry (MP), HRT and OCT. AP was set as reference and all measurements were accordingly adjusted.
|
|
|
| ||||
|---|---|---|---|---|---|---|
| slope | intercept | slope | intercept | slope | intercept | |
| Disc area (mm2) | 0.9302 | 0.1022 | 0.7678* | 0.0980 | 0.9203 | 0.0674 |
| Rim area (mm2) | 1.5558 | −0.9156 | 0.8221 | 0.0024 | 1.3465 | −0.6810 |
| Cup area (mm2) | 0.9213 | 0.0317 | 1.0415 | −0.2688 | 1.3843* | −0.3920 |
| Cup/disc area ratio | 1.1918 | −0.0813 | 1.3700 | −0.1661 | 1.8753* | −0.3237 |
| Vertical cup/disc ratio | 1.5807* | −0.3777 | 1.8066* | −0.5502 | 2.0120* | −0.6725 |
| Rim volume (mm3) | 0.4776* | 0.1426 | 0.0817* | 0.3128 | 0.2653* | 0.2768 |
| Cup volume (mm3) | 0.7156* | 0.0539 | 0.6979* | 0.0140 | 0.8084* | −0.0325 |
*P < 0.05.
Fig. 6Relative bias comparison between optic nerve head measurements of automated photogrammetry, manual photogrammetry, Heidelberg retina tomography (HRT) and optical coherence tomography (OCT). Automated photogrammetry was used as reference to all other methods